PMN-PT single crystal, high-frequency ultrasonic needle transducers for pulsed-wave Doppler application

High-frequency needle ultrasound transducers with an aperture size of 0.4 mm were fabricated using lead magnesium niobate-lead titanate (PMN-33%PT) as the active piezoelectric material. The active element was bonded to a conductive silver particle matching layer and a conductive epoxy backing through direct contact curing. An outer matching layer of parylene was formed by vapor deposition. The active element was housed within a polyimide tube and a 20-gauge needle housing. The magnitude and phase of the electrical impedance of the transducer were 47 Omega and -38deg, respectively. The measured center frequency and -6 dB fractional bandwidth of the PMN-PT needle transducer were 44 MHz and 45%, respectively. The two-way insertion loss was approximately 15 dB. In vivo high-frequency, pulsed-wave Doppler patterns of blood flow in the posterior portion and in vitro ultrasonic backscatter microscope (UBM) images of the rabbit eye were obtained with the 44-MHz needle transducer

[1]  K. Shung,et al.  Single crystal PZN/PT-polymer composites for ultrasound transducer applications , 2000, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[2]  Thomas R. Shrout,et al.  Characteristics of relaxor-based piezoelectric single crystals for ultrasonic transducers , 1996, 1996 IEEE Ultrasonics Symposium. Proceedings.

[3]  Alan Robert Selfridge The design and fabrication of ultrasonic transducers and transducer arrays , 1982 .

[4]  L From,et al.  A 40-100 MHz B-scan ultrasound backscatter microscope for skin imaging. , 1995, Ultrasound in medicine & biology.

[5]  J. Weiss,et al.  Injection of tissue plasminogen activator into a branch retinal vein in eyes with central retinal vein occlusion. , 2001, Ophthalmology.

[6]  A. Arciniegas Treatment of the occlusion of the central retinal vein by section of the posterior ring. , 1984, Annals of ophthalmology.

[7]  K. Shung,et al.  Design of efficient, broadband single-element (20-80 MHz) ultrasonic transducers for medical imaging applications , 2003, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[8]  Jie Chen,et al.  Review: commercialization of piezoelectric single crystals for medical imaging applications , 2005, IEEE Ultrasonics Symposium, 2005..

[9]  D. A. Christopher,et al.  A high-frequency pulsed-wave Doppler ultrasound system for the detection and imaging of blood flow in the microcirculation. , 1997, Ultrasound in medicine & biology.

[10]  G. Michelson,et al.  Increased vascular resistance for venous outflow in central retinal vein occlusion. , 1997, Ophthalmology.

[11]  M. J. Zipparo,et al.  Single crystal piezoelectrics: a revolutionary development for transducers , 2000, 2000 IEEE Ultrasonics Symposium. Proceedings. An International Symposium (Cat. No.00CH37121).

[12]  S. Saitoh,et al.  A 3.7 MHz phased array probe using 0.91Pb(Zn(1/3)Nb(2/3))O(3)-0.09PbTiO(3 ). , 1999, IEEE transactions on ultrasonics, ferroelectrics, and frequency control.

[13]  E. Opremcak,et al.  RADIAL OPTIC NEUROTOMY WITH ADJUNCTIVE INTRAOCULAR TRIAMCINOLONE FOR CENTRAL RETINAL VEIN OCCLUSION: 63 Consecutive Cases , 2006, Retina.

[14]  G. Hutchins,et al.  CENTRAL RETINAL VEIN OCCLUSION: A Prospective Histopathologic Study of 29 Eyes in 28 Cases , 2005, Retina.

[15]  F. Foster,et al.  Fabrication of high frequency spherically shaped ceramic transducers , 1994, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.