A 3-D PW ultrasonic Doppler flowmeter: theory and experimental characterization

A complete 3-D ultrasonic pulsed Doppler system has been developed to measure quantitatively the velocity vector field of a fluid flow independently of the probe position. The probe consists of four 2.5 MHz piezocomposite ultrasonic transducers (one central transmitter and three receivers separated by 120/spl deg/) to measure the velocity projections along three different directions. The Doppler shift of the three channels is calculated by analog phase and quadrature demodulation, then digitally processed to extract the mean velocity from the complex spectrum. The accuracy of the 3-D Doppler technique has been tested on a moving string phantom providing an error of about 4% for both amplitude and direction with an acquisition window of 100 ms.

[1]  O. Bonnefous Measurement of the complete (3D) velocity vector of blood flows , 1988, IEEE 1988 Ultrasonics Symposium Proceedings..

[2]  W. M. Gardiner,et al.  Three-dimensional Doppler velocimetry of flow jets , 1988, IEEE Transactions on Biomedical Engineering.

[3]  V. Newhouse,et al.  Ultrasound Doppler Probing of Flows Transverse with Respect to Beam Axis , 1987, IEEE Transactions on Biomedical Engineering.

[4]  V.L. Newhouse,et al.  Time Dilation and Inversion Properties and the Output Spectrum of Pulsed Doppler Flowmeters , 1983, IEEE Transactions on Sonics and Ultrasonics.

[5]  Andrew R. Walker,et al.  Evaluating doppler devices using a moving string test target , 1982, Journal of clinical ultrasound : JCU.

[6]  Katherine W. Ferrara,et al.  Improved color flow mapping using the wideband maximum likelihood estimator , 1990, IEEE Symposium on Ultrasonics.

[7]  3-D velocity field measurement using multiple ultrasonic plane detections and high-order correlation analysis , 1990, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[8]  I. Hein,et al.  Triple-beam lens transducers for three-dimensional ultrasonic fluid flow estimation , 1995, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[9]  K. Beach,et al.  Vector Doppler: accurate measurement of blood velocity in two dimensions. , 1992, Ultrasound in medicine & biology.

[10]  W. O’Brien,et al.  A flexible blood flow phantom capable of independently producing constant and pulsatile flow with a predictable spatial flow profile for ultrasound flow measurement validations , 1992, IEEE Transactions on Biomedical Engineering.

[11]  G. Trahey,et al.  Angle Independent Ultrasonic Detection of Blood Flow , 1987, IEEE Transactions on Biomedical Engineering.

[12]  R. Cobbold,et al.  A nonstationary signal simulation model for continuous wave and pulsed Doppler ultrasound , 1989, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[13]  C. K. Yuen,et al.  Theory and Application of Digital Signal Processing , 1978, IEEE Transactions on Systems, Man, and Cybernetics.

[14]  D. Cathignol,et al.  Three-dimensional vector flow estimation using two transducers and spectral width , 1994, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.