Commissioning dose computation models for spot scanning proton beams in water for a commercially available treatment planning system.

PURPOSE To present our method and experience in commissioning dose models in water for spot scanning proton therapy in a commercial treatment planning system (TPS). METHODS The input data required by the TPS included in-air transverse profiles and integral depth doses (IDDs). All input data were obtained from Monte Carlo (MC) simulations that had been validated by measurements. MC-generated IDDs were converted to units of Gy mm(2)/MU using the measured IDDs at a depth of 2 cm employing the largest commercially available parallel-plate ionization chamber. The sensitive area of the chamber was insufficient to fully encompass the entire lateral dose deposited at depth by a pencil beam (spot). To correct for the detector size, correction factors as a function of proton energy were defined and determined using MC. The fluence of individual spots was initially modeled as a single Gaussian (SG) function and later as a double Gaussian (DG) function. The DG fluence model was introduced to account for the spot fluence due to contributions of large angle scattering from the devices within the scanning nozzle, especially from the spot profile monitor. To validate the DG fluence model, we compared calculations and measurements, including doses at the center of spread out Bragg peaks (SOBPs) as a function of nominal field size, range, and SOBP width, lateral dose profiles, and depth doses for different widths of SOBP. Dose models were validated extensively with patient treatment field-specific measurements. RESULTS We demonstrated that the DG fluence model is necessary for predicting the field size dependence of dose distributions. With this model, the calculated doses at the center of SOBPs as a function of nominal field size, range, and SOBP width, lateral dose profiles and depth doses for rectangular target volumes agreed well with respective measured values. With the DG fluence model for our scanning proton beam line, we successfully treated more than 500 patients from March 2010 through June 2012 with acceptable agreement between TPS calculated and measured dose distributions. However, the current dose model still has limitations in predicting field size dependence of doses at some intermediate depths of proton beams with high energies. CONCLUSIONS We have commissioned a DG fluence model for clinical use. It is demonstrated that the DG fluence model is significantly more accurate than the SG fluence model. However, some deficiencies in modeling the low-dose envelope in the current dose algorithm still exist. Further improvements to the current dose algorithm are needed. The method presented here should be useful for commissioning pencil beam dose algorithms in new versions of TPS in the future.

[1]  Xiaodong Zhang,et al.  Beyond Gaussians: a study of single-spot modeling for scanning proton dose calculation , 2012, Physics in medicine and biology.

[2]  Bijan Arjomandy,et al.  Patient-specific quality assurance for prostate cancer patients receiving spot scanning proton therapy using single-field uniform dose. , 2011, International journal of radiation oncology, biology, physics.

[3]  Baden,et al.  Laboratory Report (LR) to the paper Foundation of an analytical proton beamlet model for inclusion in a general proton dose calculation system [arXiv:1009.0832] , 2010, 1009.2187.

[4]  R. Mohan,et al.  An MCNPX Monte Carlo model of a discrete spot scanning proton beam therapy nozzle. , 2010, Medical physics.

[5]  E. Matsinos,et al.  Theoretical methods for the calculation of Bragg curves and 3D distributions of proton beams , 2010, 1008.3645.

[6]  Radhe Mohan,et al.  Experimental characterization of the low-dose envelope of spot scanning proton beams , 2010, Physics in medicine and biology.

[7]  D Robertson,et al.  Intensity modulated proton therapy treatment planning using single-field optimization: the impact of monitor unit constraints on plan quality. , 2010, Medical physics.

[8]  Radhe Mohan,et al.  Monte Carlo investigation of the low-dose envelope from scanned proton pencil beams , 2010, Physics in medicine and biology.

[9]  X. Zhu,et al.  Computation of doses for large-angle Coulomb scattering of proton pencil beams , 2009, Physics in medicine and biology.

[10]  Uwe Titt,et al.  Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas M.D. Anderson Cancer Center, Proton Therapy Center, Houston. , 2009, Medical physics.

[11]  Radhe Mohan,et al.  The M. D. Anderson proton therapy system. , 2009, Medical physics.

[12]  Erik Grusell,et al.  Reference dosimetry in a scanned pulsed proton beam using ionisation chambers and a Faraday cup , 2008, Physics in medicine and biology.

[13]  B. Schaffner Proton dose calculation based on in-air fluence measurements , 2008, Physics in medicine and biology.

[14]  Alfred R. Smith,et al.  Initial beam size study for passive scatter proton therapy. II. Changes in delivered depth dose profiles. , 2007, Medical physics.

[15]  Matthias Fippel,et al.  A pencil beam algorithm for intensity modulated proton therapy derived from Monte Carlo simulations , 2005, Physics in medicine and biology.

[16]  D. Low,et al.  A technique for the quantitative evaluation of dose distributions. , 1998, Medical physics.

[17]  D. Schardt,et al.  Magnetic scanning system for heavy ion therapy , 1993 .

[18]  Hans A. Bethe,et al.  Moliere's theory of multiple scattering , 1953 .

[19]  B. Jones,et al.  Proton and Charged Particle Radiotherapy , 2010 .

[20]  P. Andreo,et al.  Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water , 2001 .

[21]  E. Pedroni,et al.  Dose calculation models for proton treatment planning using a dynamic beam delivery system: an attempt to include density heterogeneity effects in the analytical dose calculation. , 1999, Physics in medicine and biology.

[22]  E. Pedroni,et al.  The 200-MeV proton therapy project at the Paul Scherrer Institute: conceptual design and practical realization. , 1995, Medical physics.

[23]  Icru Prescribing, recording, and reporting photon beam therapy , 1993 .

[24]  E. Pedroni,et al.  Dose calculation and optimization for 3D conformal voxel scanning , 1992, Radiation and environmental biophysics.