Position-history and spin-history artifacts in fMRI time series

What is the impact of the spin history and position history on signal intensity after the alignment of acquired volumes? This question arises in many fMRI studies. We will focus on spin-history artefacts generated by the position-history of the scanned object. In fMRI an object is driven to steady state by applying a few dummy scans at the start of each measurement. A change in object position will disrupt the tissue's steady state magnetization. The disruption will propagate to the next few acquired volumes until a steady state is reached again. The variables which are involved in changing the longitudinal magnetization are: the shape and the position of the slice profiles, the times at which RF pulses occurred, the equilibrium magnetization map and the T1 map. Knowledge of these variables enables the prediction of those situations and the locations where the spin-history may compromise the fMRI analysis. In this paper we present a simulation of spin-history artefacts. The simulation shows that these effects, following a displacement, are similar to the transient period at the beginning of the measurement. Introducing gaps between the acquired slices increases these artefacts.

[1]  Karl J. Friston,et al.  Analysis of fMRI Time-Series Revisited , 1995, NeuroImage.

[2]  J. Pauly,et al.  Characterization and reduction of the transient response in steady‐state MR imaging , 2001, Magnetic resonance in medicine.

[3]  S J Riederer,et al.  A prospective approach to correct for inter‐image head rotation in FMRI , 1998, Magnetic resonance in medicine.

[4]  S Thesen,et al.  Prospective acquisition correction for head motion with image‐based tracking for real‐time fMRI , 2000, Magnetic resonance in medicine.

[5]  C. Jack,et al.  Prospective multiaxial motion correction for fMRI , 2000, Magnetic resonance in medicine.

[6]  Todd B. Parrish,et al.  Impact of signal‐to‐noise on functional MRI , 2000 .

[7]  M. D’Esposito,et al.  Empirical Analyses of BOLD fMRI Statistics , 1997, NeuroImage.

[8]  M. D’Esposito,et al.  Empirical analyses of BOLD fMRI statistics. I. Spatially unsmoothed data collected under null-hypothesis conditions. , 1997, NeuroImage.

[9]  V L Morgan,et al.  Comparison of functional MRI image realignment tools using a computer‐generated phantom , 2001, Magnetic resonance in medicine.

[10]  Karl J. Friston,et al.  Movement‐Related effects in fMRI time‐series , 1996, Magnetic resonance in medicine.

[11]  R. Turner,et al.  Characterization and Correction of Interpolation Effects in the Realignment of fMRI Time Series , 2000, NeuroImage.

[12]  Karl J. Friston,et al.  Analysis of functional MRI time‐series , 1994, Human Brain Mapping.

[13]  C R Meyer,et al.  Motion correction in fMRI via registration of individual slices into an anatomical volume , 1999, Magnetic resonance in medicine.

[14]  J. Hajnal,et al.  Artifacts due to stimulus correlated motion in functional imaging of the brain , 1994, Magnetic resonance in medicine.