Virtual bench testing of new generation coronary stents.

AIMS The aim of this study was to compare the stent strut apposition and stent induced vessel wall stresses of currently used coronary stent designs. This may help to better understand their clinical performance and provide the insights necessary for further optimisation. METHODS AND RESULTS We compared the stent strut apposition of six different stent designs when deployed (at 12 atm) in an idealised stenosed vessel using a novel approach based on finite element simulations. Additional insights into the mechanical behaviour of the investigated stents were obtained by virtually quantifying the stent induced vessel wall stresses. For the investigated stenosed vessel model, the percentage of malapposed struts (distance to wall>10 µm) ranged between 9% (Integrity stent) and 43% (Promus Element stent). The largest strut-artery distances were observed at the plaque shoulders. The 95 percentile of the axial stress within the intima ranges from 32 (Promus Element stent) to 83 kPa (Liberté stent). Stress peaks were mainly located at the inner curvatures of the vessel model and at the stent ends. CONCLUSIONS For the investigated case, considerable differences were observed between the studied stent platforms in terms of stent strut apposition and stent induced vessel wall stresses. These differences in mechanical behaviour may help to explain clinical observations.

[1]  N. Weissman,et al.  Natural history of intravascular ultrasound-detected edge dissections from coronary stent deployment. , 2000, American heart journal.

[2]  G. Stone,et al.  Randomized Comparison of Everolimus-Eluting and Paclitaxel-Eluting Stents: Two-Year Clinical Follow-Up From the Clinical Evaluation of the Xience V Everolimus Eluting Coronary Stent System in the Treatment of Patients With De Novo Native Coronary Artery Lesions (SPIRIT) III Trial , 2009, Circulation.

[3]  P. Serruys,et al.  Incomplete Stent Apposition After Implantation of Paclitaxel-Eluting Stents or Bare Metal Stents: Insights From the Randomized TAXUS II Trial , 2005, Circulation.

[4]  Pascal Verdonck,et al.  A Novel Simulation Strategy for Stent Insertion and Deployment in Curved Coronary Bifurcations: Comparison of Three Drug-Eluting Stents , 2009, Annals of Biomedical Engineering.

[5]  Jeffrey W Moses,et al.  Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery. , 2003, The New England journal of medicine.

[6]  Gerhard Sommer,et al.  Determination of layer-specific mechanical properties of human coronary arteries with nonatherosclerotic intimal thickening and related constitutive modeling. , 2005, American journal of physiology. Heart and circulatory physiology.

[7]  K. Nasu,et al.  Incomplete stent apposition after coronary stent implantation: myth or reality? , 2009, Journal of interventional cardiology.

[8]  StéphaneCook,et al.  Incomplete Stent Apposition and Very Late Stent Thrombosis After Drug-Eluting Stent Implantation , 2007 .

[9]  G. Stone,et al.  Long-term impact of routinely detected early and late incomplete stent apposition: an integrated intravascular ultrasound analysis of the TAXUS IV, V, and VI and TAXUS ATLAS workhorse, long lesion, and direct stent studies. , 2010, JACC. Cardiovascular interventions.

[10]  S. Marso,et al.  Incomplete stent apposition: should we appose or oppose? , 2010, JACC: Cardiovascular Interventions.

[11]  Peter Barlis,et al.  Intravascular optical coherence tomography: optimisation of image acquisition and quantitative assessment of stent strut apposition. , 2007, EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology.

[12]  Late incomplete stent apposition after sirolimus-eluting stent implantation: a serial intravascular ultrasound analysis. , 2005, Journal of the American College of Cardiology.

[13]  Seung‐Jung Park,et al.  Late Stent Malapposition After Drug-Eluting Stent Implantation: An Intravascular Ultrasound Analysis With Long-Term Follow-Up , 2006, Circulation.

[14]  G. Stone,et al.  Outcome after acute incomplete sirolimus-eluting stent apposition as assessed by serial intravascular ultrasound. , 2006, The American journal of cardiology.

[15]  P. Serruys,et al.  The fate of incomplete stent apposition with drug-eluting stents: an optical coherence tomography-based natural history study. , 2010, European heart journal.

[16]  P. Fitzgerald,et al.  Impact of late incomplete stent apposition after sirolimus-eluting stent implantation on 4-year clinical events: intravascular ultrasound analysis from the multicentre, randomised, RAVEL, E-SIRIUS and SIRIUS trials , 2007, Heart.

[17]  W. Carroll,et al.  A platinum-chromium steel for cardiovascular stents. , 2010, Biomaterials.

[18]  Caitríona Lally,et al.  Determination of the influence of stent strut thickness using the finite element method: implications for vascular injury and in-stent restenosis , 2009, Medical & Biological Engineering & Computing.

[19]  T. Stijnen,et al.  Late stent malapposition risk is higher after drug-eluting stent compared with bare-metal stent implantation and associates with late stent thrombosis. , 2010, European heart journal.

[20]  P. Boekstegers,et al.  Influence of stent design on 1‐year outcome after coronary stent placement: A randomized comparison of five stent types in 1,147 unselected patients , 2000, Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions.