Validation of a Body-Conducted Sound Sensor for Respiratory Sound Monitoring and a Comparison with Several Sensors

The ideal respiratory sound sensor exhibits high sensitivity, wide-band frequency characteristics, and excellent anti-noise properties. We investigated the body-conducted sound sensor (BCS) and verified its usefulness in respiratory sound monitoring through comparison with an air-coupled microphone (ACM) and acceleration sensor (B & K: 8001). We conducted four experiments for comparison: (1) estimation by equivalent circuit model of sensors and measurement by a sensitivity evaluation system; (2) measurement of tissue-borne sensitivity-to-air-noise sensitivity ratio (SRTA); (3) respiratory sound measurement through a simulator; and (4) actual respiratory sound measurement using human subjects. For (1), the simulation and measured values of all the sensors showed good agreement; BCS demonstrated sensitivity ~10 dB higher than ACM and higher sensitivity in the high-frequency segments compared with 8001. In (2), BCS showed high SRTA in the 600–1000 and 1200–2000-Hz frequency segments. In (3), BCS detected wheezes in the high-frequency segments of the respiratory sound. Finally, in (4), the sensors showed similar characteristics and features in the high-frequency segments as the simulators, where typical breathing sound detection was possible. BCS displayed a higher sensitivity and anti-noise property in high-frequency segments compared with the other sensors and is a useful respiratory sound sensor.

[1]  H. Pasterkamp,et al.  Lung sound spectra at standardized air flow in normal infants, children, and adults. , 1996, American journal of respiratory and critical care medicine.

[2]  Kiyohiro Shikano,et al.  Non-Audible Murmur (NAM) Recognition , 2006, IEICE Trans. Inf. Syst..

[3]  B. Debaene,et al.  Accuracy of respiratory rate monitoring using a non-invasive acoustic method after general anaesthesia. , 2012, British journal of anaesthesia.

[4]  M Nissan,et al.  Spectral characteristics of chest wall breath sounds in normal subjects. , 1995, Thorax.

[5]  H S Hira,et al.  LUNG SOUNDS , 1978, The Journal of the Association of Physicians of India.

[6]  H. Pasterkamp,et al.  Measurement of respiratory acoustical signals. Comparison of sensors. , 1993, Chest.

[7]  P. Doerschuk,et al.  Sonic phase delay from trachea to chest wall: spatial and inhaled gas dependence , 2006, Medical and Biological Engineering and Computing.

[8]  D. S. Morillo,et al.  Monitoring and Analysis of Cardio Respiratory and Snoring Signals by using an Accelerometer , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[9]  T. Hirahara High-sensitivity body-conducted sound sensor , 2010 .

[10]  George R. Wodicka,et al.  Air-Borne and Tissue-Borne Sensitivities of Bioacoustic Sensors Used on the Skin Surface , 2009, IEEE Transactions on Biomedical Engineering.

[11]  T. Shimohata,et al.  Natural course and potential prognostic factors for sleep-disordered breathing in multiple system atrophy. , 2017, Sleep medicine.

[12]  George R. Wodicka,et al.  Design, construction, and evaluation of a bioacoustic transducer testing (BATT) system for respiratory sounds , 2006, IEEE Transactions on Biomedical Engineering.

[13]  Mikihiro Nakagiri,et al.  Statistical Voice Conversion Techniques for Body-Conducted Unvoiced Speech Enhancement , 2012, IEEE Transactions on Audio, Speech, and Language Processing.

[14]  S. Kraman,et al.  Does the vesicular lung sound come only from the lungs? , 1983, The American review of respiratory disease.

[15]  Wei Zhang,et al.  A Unified Framework for Street-View Panorama Stitching , 2016, Sensors.

[16]  Makoto Otani,et al.  Frequency characteristics of several non-audible murmur (NAM) microphones , 2009 .

[17]  A. Dittmar,et al.  The relationship between normal lung sounds, age, and gender. , 2000, American journal of respiratory and critical care medicine.

[18]  Hans Pasterkamp,et al.  Comparison of lung sound transducers using a bioacoustic transducer testing system. , 2006, Journal of applied physiology.

[19]  H. Nakano,et al.  Breath Sound Intensity during Tidal Breathing in COPD Patients. , 2015, Internal medicine.

[20]  Andrey Vyshedskiy,et al.  Automated lung sound analysis in patients with pneumonia. , 2004, Respiratory care.

[21]  A. Vyshedskiy,et al.  Sound transmission in the lung as a function of lung volume. , 2002, Journal of applied physiology.

[22]  Toshimitsu Shinohara,et al.  Validation of a new system of tracheal sound analysis for the diagnosis of sleep apnea-hypopnea syndrome. , 2004, Sleep.

[23]  H. Pasterkamp,et al.  Effect of gas density on respiratory sounds. , 1996, American journal of respiratory and critical care medicine.

[24]  Cheng-Ta Yang,et al.  Design of Wearable Breathing Sound Monitoring System for Real-Time Wheeze Detection , 2017, Sensors.

[25]  Zahra Moussavi,et al.  Automatic and Unsupervised Snore Sound Extraction From Respiratory Sound Signals , 2011, IEEE Transactions on Biomedical Engineering.

[26]  Jeffrey J Ward,et al.  Technology for Enhancing Chest Auscultation in Clinical Simulation , 2011, Respiratory Care.

[27]  T. Shimoda,et al.  Lung sound analysis can be an index of the control of bronchial asthma. , 2017, Allergology international : official journal of the Japanese Society of Allergology.

[28]  J Vanderschoot,et al.  Abnormal lung sounds in patients with asthma during episodes with normal lung function. , 1994, Chest.

[29]  Lu Yu,et al.  Using the Entropy of Tracheal Sounds to Detect Apnea during Sedation in Healthy Nonobese Volunteers , 2013, Anesthesiology.

[30]  Hiroshi Ishiguro,et al.  Using Body-Conducted Acoustic Sensors for Human-Robot Communication in Noisy Environments , 2013 .

[31]  H. Pasterkamp,et al.  Chest surface mapping of lung sounds during methacholine challenge , 1997, Pediatric pulmonology.

[32]  M. Toda,et al.  Contact-Type Vibration Sensors Using Curved Clamped PVDF Film , 2006, IEEE Sensors Journal.

[33]  Tomoki Toda,et al.  Silent-speech enhancement using body-conducted vocal-tract resonance signals , 2010, Speech Commun..

[34]  Zahra Moussavi,et al.  A robust method for estimating respiratory flow using tracheal sounds entropy , 2006, IEEE Transactions on Biomedical Engineering.