Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

The overall risk of clinically significant adverse events related to EMI in recipients of CIEDs is very low. Therefore, no special precautions are needed when household appliances are used. Environmental and industrial sources of EMI are relatively safe when the exposure time is limited and distance from the CIEDs is maximized. The risk of EMI-induced events is highest within the hospital environment. Physician awareness of the possible interactions and methods to minimize them is warranted.

[1]  G. Papadimitriou,et al.  Cardiac Rhythm Management Devices and Electroconvulsive Therapy: A Critical Review Apropos of a Depressed Patient With a Pacemaker , 2011, The journal of ECT.

[2]  A. Gillis,et al.  Pacemakers, implantable cardioverter/defibrillators, and extracorporeal shockwave lithotripsy: evidence-based guidelines for the modern era. , 2008, Journal of endourology.

[3]  John Gosbee,et al.  ACR guidance document for safe MR practices: 2007. , 2007, AJR. American journal of roentgenology.

[4]  Mehul B. Patel,et al.  Do Media Players Cause Interference with Pacemakers? , 2009, Clinical cardiology.

[5]  J. Mugica,et al.  Study of Interactions Between Permanent Pacemakers and Electronic Antitheft Surveillance Systems , 2000, Pacing and clinical electrophysiology : PACE.

[6]  D P Zipes,et al.  Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators. , 1999, Circulation.

[7]  B. Wilkoff,et al.  Effects of Extracorporeal Shock Wave Lithotripsy on Cardiac Pacemakers and its Safety in Patients with Implanted Cardiac Pacemakers , 1988, Pacing and clinical electrophysiology : PACE.

[8]  A. Henning,et al.  Radiation‐Induced Effects in Multiprogrammable Pacemakers and Implantable Defibrillators , 1991, Pacing and clinical electrophysiology : PACE.

[9]  H. Halperin,et al.  Effects of Surgical and Endoscopic Electrocautery on Modern‐Day Permanent Pacemaker and Implantable Cardioverter‐Defibrillator Systems , 2008, Pacing and clinical electrophysiology : PACE.

[10]  R. Sarin,et al.  Radiation therapy with implanted cardiac pacemaker devices: a clinical and dosimetric analysis of patients and proposed precautions. , 2011, Clinical oncology (Royal College of Radiologists (Great Britain)).

[11]  T. Guarnieri,et al.  Increased Pacing Threshold After an Automatic Defibrillator Shock in Dogs: Effects of Class I and Class II Antiarrhythmic Drugs , 1988, Pacing and clinical electrophysiology : PACE.

[12]  T. Hashimoto,et al.  Influence of secondary neutrons induced by proton radiotherapy for cancer patients with implantable cardioverter defibrillators , 2012, Radiation oncology.

[13]  M. Kavaliers,et al.  A survey of magnetic fields in the dental operatory. , 1994, Journal.

[14]  R R Price,et al.  The AAPM/RSNA physics tutorial for residents. MR imaging safety considerations. Radiological Society of North America. , 1999, Radiographics : a review publication of the Radiological Society of North America, Inc.

[15]  C. Kalkman,et al.  Monitor‐induced tachycardia in a patient with a rate‐responsive pacemaker , 2006, Anaesthesia.

[16]  G. Trevi,et al.  Implantable Cardioverter Defibrillators and Cellular Telephones: Is There Any Interference? , 1999, Pacing and clinical electrophysiology : PACE.

[17]  R. Ozdemir,et al.  Effects of Mobile Telephones on the Function of Implantable Cardioverter Defibrillators , 2005, Annals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc.

[18]  M. Chester,et al.  The simultaneous use of a biventricular implantable cardioverter defibrillator (ICD) and transcutaneous electrical nerve stimulation (TENS) unit: implications for device interaction. , 2003, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[19]  V. Vandewalle,et al.  Use of an Implantable Cardioverter Defibrillator in a Patient with Two Implanted Neurostimulators for Severe Parkinson's Disease , 2000, Pacing and clinical electrophysiology : PACE.

[20]  F. Romeo,et al.  Safety and efficacy of a new magnetic resonance imaging-compatible pacing system: early results of a prospective comparison with conventional dual-chamber implant outcomes. , 2010, Heart rhythm.

[21]  O. Bertel,et al.  Hazards of an alternative medicine device in a patient with a pacemaker. , 2004, The New England journal of medicine.

[22]  S. Ryf,et al.  Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators. , 2007, Heart rhythm.

[23]  R. Schuster,et al.  Life-Threatening Pacemaker Dysfunction Associated With Therapeutic Radiation: A Case Report , 2009, Angiology.

[24]  C. Pandozi,et al.  Transthoracic DC Shock May Represent a Serious Hazard in Pacemaker Dependent Patients , 1995, Pacing and clinical electrophysiology : PACE.

[25]  B. Allaria,et al.  Technical and Clinical Problems in Patients with Simultaneous Implantation of a Cardiac Pacemaker and Spinal Cord Stimulator , 1993, Pacing and clinical electrophysiology : PACE.

[26]  C. Chiu,et al.  A Prospective Pediatric Clinical Trial of Digital Music Players: Do They Interfere with Pacemakers? , 2009, Journal of cardiovascular electrophysiology.

[27]  H. Halperin,et al.  Quantitative Assessment of Artifacts on Cardiac Magnetic Resonance Imaging of Patients With Pacemakers and Implantable Cardioverter-Defibrillators , 2011, Circulation. Cardiovascular imaging.

[28]  J. Conti Cardiac Pacing, Defibrillation and Resynchronization: A Clinical Approach , 2009 .

[29]  M. Sweesy,et al.  Electromagnetic interference in cardiac rhythm management devices. , 2004, AACN clinical issues.

[30]  A. Last Radiotherapy in patients with cardiac pacemakers. , 1998, The British journal of radiology.

[31]  Frank G Shellock,et al.  Regarding the value reported for the term "spatial gradient magnetic field" and how this information is applied to labeling of medical implants and devices. , 2011, AJR. American journal of roentgenology.

[32]  J. Langberg,et al.  The Effects of Extracorporeal Shock Wave Lithotripsy on Pacemaker Function , 1987, Pacing and clinical electrophysiology : PACE.

[33]  C. Hurkmans,et al.  Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators. , 2005, International journal of radiation oncology, biology, physics.

[34]  J. Gimbel Magnetic Resonance Imaging of Implantable Cardiac Rhythm Devices at 3.0 Tesla , 2008, Pacing and clinical electrophysiology : PACE.

[35]  P. van der Kemp,et al.  Electromagnetic interference in pacemakers in single-engine fixed-wing aircraft: a European perspective. , 2002, Aviation, space, and environmental medicine.

[36]  C. Waller,et al.  Adverse effects of direct current cardioversion on cardiac pacemakers and electrodes Is external cardioversion contraindicated in patients with permanent pacing systems? , 2004, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[37]  Isabel Deisenhofer,et al.  Safety of Screening Procedures With Hand-Held Metal Detectors Among Patients With Implanted Cardiac Rhythm Devices , 2011, Annals of Internal Medicine.

[38]  S. Barold,et al.  Reprogramming of Implanted Pacemaker Following External Defibrillation , 1978, Pacing and clinical electrophysiology : PACE.

[39]  J. Rosenow,et al.  Simultaneous use of bilateral subthalamic nucleus stimulators and an implantable cardiac defibrillator. Case report. , 2003, Journal of neurosurgery.

[40]  K. Ellenbogen,et al.  Acute Effects of Radiofrequency Ablation of Atrial Arrhythmias on Implanted Permanent Pacing Systems , 1996, Pacing and clinical electrophysiology : PACE.

[41]  Eorge,et al.  INTERFERENCE WITH CARDIAC PACEMAKERS BY CELLULAR TELEPHONES , 2000 .

[42]  Richard G Trohman,et al.  Interference in Implanted Cardiac Devices, Part I , 2002, Pacing and clinical electrophysiology : PACE.

[43]  S. Barold,et al.  Adverse acute and chronic effects of electrical defibrillation and cardioversion on implanted unipolar cardiac pacing systems. , 1983, Journal of the American College of Cardiology.

[44]  Joshua W. Guag,et al.  In vitro tests reveal sample radiofrequency identification readers inducing clinically significant electromagnetic interference to implantable pacemakers and implantable cardioverter-defibrillators. , 2010, Heart rhythm.

[45]  M. Kroll,et al.  TASER conducted electrical weapons and implanted pacemakers and defibrillators , 2009, 2009 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[46]  N. Chan,et al.  Inappropriate implantable cardioverter-defibrillator shock due to external alternating current leak: report of two cases. , 2005, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[47]  Hirosuke Kobayashi,et al.  X‐ray Radiation Causes Electromagnetic Interference in Implantable Cardiac Pacemakers , 2010, Pacing and clinical electrophysiology : PACE.

[48]  M. Keuchel,et al.  Capsule Endoscopy in Patients with Cardiac Pacemakers and Implantable Cardioverter Defibrillators: (Re)evaluation of the Current State in Germany, Austria, and Switzerland 2010 , 2011, Gastroenterology research and practice.

[49]  F. Venditti,et al.  Effect of Extracorporeal Shock Wave Lithotripsy on Implantable Cardioverter Defibrillator , 1993, Pacing and clinical electrophysiology : PACE.

[50]  E. Giltay,et al.  Electroconvulsive Therapy (ECT) in a Patient with a Dual-chamber Sensing, VDDR Pacemaker , 2005, The journal of ECT.

[51]  Fred Morady,et al.  Adverse Interaction Between a Left Ventricular Assist Device and an Implantable Cardioverter Defibrillator , 2007, Journal of cardiovascular electrophysiology.

[52]  Giovanni Calcagnini,et al.  Interference between mobile phones and pacemakers: a look inside. , 2007, Annali dell'Istituto superiore di sanita.

[53]  S. Dubner,et al.  Electromagnetic Interference from Wireless Video‐Capsule Endoscopy on Implantable Cardioverter‐Defibrillators , 2007, Pacing and clinical electrophysiology : PACE.

[54]  Peter Hunold,et al.  Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment. , 2011, Heart rhythm.

[55]  S. Priori,et al.  Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices. , 2008, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[56]  L. Haegeli,et al.  Effect of a Taser shot to the chest of a patient with an implantable defibrillator. , 2006, Heart rhythm.

[57]  F. Eschwège,et al.  Influence of high-energy photon beam irradiation on pacemaker operation. , 2002, Physics in medicine and biology.

[58]  M. Strüber,et al.  Implantable defibrillator with left ventricular assist device compatibility. , 2009, Interactive cardiovascular and thoracic surgery.

[59]  M. Chung,et al.  The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors: facilities and patient management this document was developed as a joint project with , 2011, Heart rhythm.

[60]  R. Fimmers,et al.  Safety of brain 3-T MR imaging with transmit-receive head coil in patients with cardiac pacemakers: pilot prospective study with 51 examinations. , 2008, Radiology.

[61]  F. Morady,et al.  Simultaneous use of implantable cardioverter-defibrillators and left ventricular assist devices in patients with severe heart failure. , 2010, The American journal of cardiology.

[62]  Hugo A Katus,et al.  Radiation therapy-induced electrical reset of an implantable cardioverter defibrillator device located outside the irradiation field. , 2004, Journal of electrocardiology.

[63]  S. Dubner,et al.  Electromagnetic interference with implantable cardiac pacemakers by video capsule. , 2005, Gastrointestinal endoscopy.

[64]  Terry O Woods,et al.  MR labeling information for implants and devices: explanation of terminology. , 2009, Radiology.

[65]  J Rod Gimbel,et al.  Unexpected pacing inhibition upon exposure to the 3T static magnetic field prior to imaging acquisition: what is the mechanism? , 2011, Heart rhythm.

[66]  M. McIvor,et al.  Study of Pacemaker and Implantable Cardioverter Defibrillator Triggering by Electronic Article Surveillance Devices (SPICED TEAS) , 1998, Pacing and clinical electrophysiology : PACE.

[67]  August A. de Rotte,et al.  Electromagnetic interference in pacemakers in single-engine fixed-wing aircraft: a European perspective. , 2002 .

[68]  M. Chung,et al.  The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors: facilities and patient management: executive summary this document was developed as a , 2011, Heart rhythm.

[69]  N. Goldschlager,et al.  Environmental Effects on Cardiac Pacing Systems , 2008 .

[70]  J. Leighton,et al.  Safety of capsule endoscopy in patients with pacemakers. , 2004, Gastrointestinal endoscopy.

[71]  Mehmet Ileri,et al.  The effects of mobile phones on pacemaker function. , 2005, International journal of cardiology.

[72]  A. Walmsley,et al.  Ultrasonic dental scaler: associated hazards. , 2003, Journal of clinical periodontology.

[73]  J. Nicolás,et al.  Washing Machine Associated 50 Hz Detected As Ventricular Fibrillation by An Implanted Cardioverter Defibrillator , 2001, Pacing and clinical electrophysiology : PACE.

[74]  A. Badreldin,et al.  Third‐Generation Mobile Phones (UMTS) Do Not Interfere with Permanent Implanted Pacemakers , 2010, Pacing and clinical electrophysiology : PACE.

[75]  Martine Souques,et al.  Clinical study of interference with cardiac pacemakers by a magnetic field at power line frequencies. , 2005, Journal of the American College of Cardiology.

[76]  B Maisch,et al.  Interactions between Pacemakers and Security Systems , 1998, Pacing and clinical electrophysiology : PACE.

[77]  M. Ilan,et al.  The safety of dental mini-magnets in patients with permanent cardiac pacemakers. , 1995, The Journal of prosthetic dentistry.

[78]  Maria A. Stuchly,et al.  Pacemaker interference by magnetic fields at power line frequencies , 2002, IEEE Transactions on Biomedical Engineering.

[79]  M. Stanton,et al.  Electromagnetic interference from welding and motors on implantable cardioverter-defibrillators as tested in the electrically hostile work site. , 1996, Journal of the American College of Cardiology.

[80]  Claus Schmitt,et al.  Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators? , 2003, Journal of the American College of Cardiology.

[81]  W Boivin,et al.  CHARACTERIZATION OF THE MAGNETIC FIELDS AROUND WALK-THROUGH AND HAND-HELD METAL DETECTORS , 2003, Health physics.

[82]  J. Venselaar The effects of ionizing radiation on eight cardiac pacemakers and the influence of electromagnetic interference from two linear accelerators. , 1985, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[83]  J. Stockman,et al.  A Prospective Evaluation of a Protocol for Magnetic Resonance Imaging of Patients With Implanted Cardiac Devices , 2013 .

[84]  A. Madrid,et al.  Dysfunction of Implantable Defibrillators Caused by Slot Machines , 1997, Pacing and clinical electrophysiology : PACE.

[85]  G. Louridas,et al.  Inappropriate ICD therapy due to electrical interference: external alternating current leakage. , 2004, The Journal of invasive cardiology.

[86]  J Rod Gimbel,et al.  Electronic article surveillance systems and interactions with implantable cardiac devices: risk of adverse interactions in public and commercial spaces. , 2007, Mayo Clinic proceedings.

[87]  P. Arnsbo,et al.  Management of Spinal Cord Stimulators in Patients with Cardiac Pacemakers , 1990, Pacing and clinical electrophysiology : PACE.

[88]  C. Mannheimer,et al.  Risk of Interference from Transcutaneous Electrical Nerve Stimulation on the Sensing Function of Implantable Defibrillators , 2008, Pacing and clinical electrophysiology : PACE.

[89]  R. Schlegel,et al.  Effects of an increased air gap on the in vitro interaction of wireless phones with cardiac pacemakers. , 2000, Bioelectromagnetics.

[90]  M. Chung,et al.  Safety and efficacy of radiofrequency energy catheter ablation of atrial fibrillation in patients with pacemakers and implantable cardiac defibrillators. , 2005, Heart rhythm.

[91]  D. Bluemke,et al.  Safety of Magnetic Resonance Imaging in Patients With Cardiovascular Devices: An American Heart Association Scientific Statement From the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention , 2007, Circulation.

[92]  M. J. Solan,et al.  Treatment of patients with cardiac pacemakers and implantable cardioverter-defibrillators during radiotherapy. , 2004, International journal of radiation oncology, biology, physics.

[93]  A. Manolis,et al.  In-vivo testing of digital cellular telephones in patients with implantable cardioverter-defibrillators. , 2001, European heart journal.

[94]  M. Messier,et al.  Effects of Radiofrequency Catheter Ablation on Patients with Permanent Pacemakers , 1997, Journal of Interventional Cardiac Electrophysiology.

[95]  M. Santini,et al.  Evaluation of Electromagnetic Interference of GSM Mobile Phones with Pacemakers Featuring Remote Monitoring Functions , 2006, Pacing and clinical electrophysiology : PACE.

[96]  H. Mond,et al.  The 11th World Survey of Cardiac Pacing and Implantable Cardioverter‐Defibrillators: Calendar Year 2009–A World Society of Arrhythmia's Project , 2011, Pacing and clinical electrophysiology : PACE.

[97]  C. Schmitt,et al.  Incidence of Electromagnetic Interference in Implantable Cardioverter Defibrillators , 2001, Pacing and clinical electrophysiology : PACE.

[98]  L. Saxon,et al.  Taser‐Induced Rapid Ventricular Myocardial Capture Demonstrated by Pacemaker Intracardiac Electrograms , 2007, Journal of cardiovascular electrophysiology.

[99]  P. Andréll,et al.  Interference of transcutaneous electrical nerve stimulation with permanent ventricular stimulation: a new clinical problem? , 2009, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[100]  J. Langberg,et al.  The Effect of Radiofrequency Catheter Ablation on Permanent Pacemakers: An Experimental Study , 1990, Pacing and clinical electrophysiology : PACE.

[101]  P. Ellinor,et al.  Intermittent, erratic behaviour of an implantable cardioverter defibrillator secondary to a hidden magnetic source of interference. , 2011, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[102]  M. Sontag,et al.  Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. American Association of Physicists in Medicine. , 1994, Medical physics.

[103]  G A Ewy,et al.  Myocardial Necrosis from Direct Current Countershock: Effect of Paddle Electrode Size and Time Interval Between Discharges , 1974, Circulation.