NC-stat sensory nerve conduction studies in the median and ulnar nerves of symptomatic patients

OBJECTIVE This study evaluated validity and reliability of automated median and ulnar sensory nerve conduction study (NCS) measurements by the NC-stat. METHODS Median and ulnar distal sensory latencies (DSL) and amplitudes (SNAP) were measured in sixty subjects with the NC-stat and by a neurologist (reference) using traditional instrumentation. The median-ulnar DSL differences (MUD) was calculated. Validity was quantified by the Pearson correlation. Reliability was evaluated by the intraclass correlation coefficient (ICC), Bland-Altman analysis, and inter-rater agreement of MUD abnormalities. RESULTS As a result of differences in electrode placement, NC-stat and reference mean values had systematic differences. The correlation ranged from 0.70 (ulnar DSL) to 0.91 (median DSL). The ICC ranged from 0.69 (ulnar DSL) to 0.91 (median DSL). In Bland-Altman analysis of DSLs, NC-stat measurements had a bias of 0.56 ms (median) and 0.31 ms (ulnar) and precision of 0.31 and 0.30 ms. Inter-rater agreement for MUD abnormalities was 93.8% (raw) and 0.80 (Kappa). CONCLUSIONS NC-stat validity and reliability metrics were similar to traditional NCS. Use of the NC-stat would require applicable reference ranges. SIGNIFICANCE NC-stat median and ulnar NCS are valid and reliable. This device may be useful for increasing availability of NCS when clinically appropriate.

[1]  G. Dawson,et al.  The relative excitability and conduction velocity of sensory and motor nerve fibres in man , 1956, The Journal of physiology.

[2]  An Automated Electrodiagnostic Technique for Detection of Carpal Tunnel Syndrome , 2000 .

[3]  G. Deuschl,et al.  Recommendations for the practice of clinical neurophysiology: guidelines of the International Federation of Clinical Neurophysiology. , 1999, Electroencephalography and clinical neurophysiology. Supplement.

[4]  J Kimura,et al.  Facts, fallacies, and fancies of nerve conduction studies: Twenty‐first annual Edward H. Lambert lecture , 1997, Muscle & nerve.

[5]  J. Glass,et al.  Inter‐ and intraexaminer reliability of nerve conduction measurements in patients with diabetic neuropathy , 1994, Neurology.

[6]  B. Perkins,et al.  Diabetic neuropathy: a review emphasizing diagnostic methods , 2003, Clinical Neurophysiology.

[7]  D Cros,et al.  Median neuropathy at the wrist: diagnostic utility of clinical findings and an automated electrodiagnostic device. , 2000, Journal of occupational and environmental medicine.

[8]  D. Altman,et al.  Comparing methods of measurement: why plotting difference against standard method is misleading , 1995, The Lancet.

[9]  L. Melton,et al.  The Rochester Diabetic Neuropathy Study , 1991, Neurology.

[10]  Morris A. Fisher Comparison of automated and manual F-wave latency measurements , 2005, Clinical Neurophysiology.

[11]  Royce C. Lewis,et al.  Treatment of carpal tunnel syndrome by members of the American Society for Surgery of the Hand: results of a questionnaire. , 1987, The Journal of hand surgery.

[12]  C. Krarup,et al.  The carpal tunnel syndrome: diagnostic utility of the history and physical examination findings. , 1990, Annals of internal medicine.

[13]  R. Werner,et al.  Prolonged median sensory latency as a predictor of future carpal tunnel syndrome , 2001, Muscle & nerve.

[14]  Y. So,et al.  Literature review of the usefulness of nerve conduction studies and electromyography for the evaluation of patients with carpal tunnel syndrome , 1993, Muscle & nerve.

[15]  Shai N Gozani,et al.  Median and ulnar nerve conduction measurements in patients with symptoms of diabetic peripheral neuropathy using the NC-stat system. , 2004, Diabetes technology & therapeutics.

[16]  P. Valensi,et al.  Reproducibility of Parameters for Assessment of Diabetic Neuropathy , 1993, Diabetic medicine : a journal of the British Diabetic Association.

[17]  Xuan Kong,et al.  Detection of Lumbosacral Nerve Root Compression With a Novel Composite Nerve Conduction Measurement , 2002, Spine.

[18]  Bernard R. Rosner,et al.  Fundamentals of Biostatistics. , 1992 .

[19]  S. Evers,et al.  Reproducibility of different nerve conduction velocity measurements in healthy test subjects and patients suffering from diabetic polyneuropathy. , 1997, Electromyography and clinical neurophysiology.

[20]  E. Stålberg,et al.  Effect of Heating on Nerve Conduction in Carpal Tunnel Syndrome , 2004, Journal of clinical neurophysiology : official publication of the American Electroencephalographic Society.

[21]  Robert Tibshirani,et al.  An Introduction to the Bootstrap , 1994 .

[22]  G. Gronseth,et al.  Distal symmetrical polyneuropathy: Definition for clinical research , 2005, Muscle & nerve.

[23]  V. Bril,et al.  Electrophysiological monitoring in clinical trials , 1998 .

[24]  E F Shaw Wilgis,et al.  Timing of improvement after carpal tunnel release. , 2004, Journal of surgical orthopaedic advances.

[25]  J. Albers,et al.  Reliability of nerve conduction studies among active workers , 1999, Muscle & nerve.

[26]  Erik Stålberg,et al.  Recommendations for the practice of Clinical Neurophysiology: Guidelines of the International Federation of clinical physiology (EEG suppl 52) , 1999 .

[27]  A. Wilbourn Sensory Nerve Conduction Studies , 1994, Journal of clinical neurophysiology : official publication of the American Electroencephalographic Society.

[28]  J. Glass,et al.  Inter‐ and intra‐examiner reliability of nerve conduction measurements in normal subjects , 1991, Annals of neurology.

[29]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[30]  Shai N Gozani,et al.  Time course and predictors of median nerve conduction after carpal tunnel release. , 2004, The Journal of hand surgery.

[31]  S. Rutkove Effects of temperature on neuromuscular electrophysiology , 2001, Muscle & nerve.