1 {\Omega}-10 k{\Omega} high precision transportable setup to calibrate multifunction electrical instruments

A temperature controlled 1 {\Omega}-10 k{\Omega} standard Resistors transportable setup was developed at National Institute of Metrological Research, (INRIM) for the calibration and adjustment of multifunction electrical instruments. The two Standards consist respectively of two 10 {\Omega} and 100 k{\Omega} parallel connected resistors nets inserted in a temperature controlled aluminium box. Novelty of the realization is the oil insertion of the 1 {\Omega} net with its internal connectors lowering the thermo-electromotive forces (emfs) effects. Short and mid-term stabilities of the setup Standards resulted on the order and in some cases better than other top level 1 {\Omega} and 10 k{\Omega} commercial Standards. The transport effect turning off the setup temperature control did not cause appreciable measurement deviations on the two Standards. The Standards uncertainties meet those requested by DMMs and MFCs manufacturers to calibrate and adjust these instruments. A test to adjust a multifunction calibrator gave satisfactory results.

[1]  Francesca Pennecchi,et al.  ON THE IN-USE UNCERTAINTY OF AN INSTRUMENT , 2004 .

[2]  E. Gasparotto,et al.  Analysis of a national comparison in the field of electrical low dc resistance , 2014 .

[3]  N. Kusters,et al.  A Direct-Current-Comparator Ratio Bridge for Four-Terminal Resistance Measurements , 1966 .

[4]  Milena Astrua,et al.  A temperature variable high accuracy 10 kΩ resistor , 2009 .

[5]  A. C. Grossenbacher Development of a precision one ohm resistance standard , 1998, 1998 Conference on Precision Electromagnetic Measurements Digest (Cat. No.98CH36254).

[6]  T. J. Witt,et al.  Report on the 1990 International Comparison of 1 Ω and 10 kΩ Resistance Standards at the BIPM , 1992 .

[7]  B.J. Pritchard,et al.  Fabrication of reference standard 1 ohm resistors from Evanohm S alloy , 1990, Conference on Precision Electromagnetic Measurements.

[8]  Giancarlo Marullo-Reedtz,et al.  Traceability of the 10 k/spl Omega/ standard at IEN , 2000, Conference on Precision Electromagnetic Measurements. Conference Digest. CPEM 2000 (Cat. No.00CH37031).

[9]  Umberto Pogliano,et al.  Stability evaluation of high-precision multifunction instruments for traceability transfer , 2000, IEEE Trans. Instrum. Meas..

[10]  J. L. Thomas A new design of precision resistance standard , 1930 .

[11]  P. P. Capra,et al.  Traceability of the 10 kΩ standard at IEN , 2000 .

[12]  Flavio Galliana,et al.  Inter-laboratories comparison at 100 GΩ and 1 TΩ level to evaluate the traceability transfer from INRIM in the field of high dc resistance , 2012 .

[13]  T. J. Witt,et al.  An accurate 10 k Omega resistance standard , 1990, Conference on Precision Electromagnetic Measurements.

[14]  J. L. Thomas,et al.  Stability of double-walled manganin resistors , 1946 .

[15]  N. Fletcher,et al.  A transportable thermoregulated enclosure for standard resistors , 2012, 2012 Conference on Precision electromagnetic Measurements.

[16]  R. Poepel The Josephson Effect and Voltage Standards , 1992 .