The Differentiation of Loose and Secure Femoral Implants in Total Hip Replacement Using a Vibrational Technique: An Anatomical and Pilot Clinical Study

Anatomical study: The common pathway to failure in total hip replacements is loosening of the acetabular and femoral components. The reliable diagnosis of this loosening is difficult. A non-invasive technique has been developed which analyses a vibration signal transmitted through the femur. This can distinguish between a femoral component which is loose and one which is firmly fixed. Charnley femoral components were cemented securely into ten cadaver femurs and the transmitted vibration signal recorded. The prostheses were then loosened first at the cement-prosthesis and then at the cement-bone interface. The tests were then repeated. Consistent and distinct differences between output vibration signals were observed between the firmly implanted and the loose prostheses. Pilot clinical study: Seven patients admitted for recision surgery because of clinical and radiological evidence of femoral implant loosening were tested using this technique. A further four patients with secure femoral components were tested approximately two weeks after total hip replacement. The differences in output signal, which had been observed between firmly implanted and loose prostheses during the anatomical study, were again demonstrated.

[1]  T. Hirayama,et al.  Assessment of fracture healing by vibration. , 1979, Acta orthopaedica Scandinavica.

[2]  R. J. Collier,et al.  Non-invasive method of measuring resonant frequency of a human tibia in vivo. Part 1. , 1987, Journal of biomedical engineering.

[3]  S. Kattapuram,et al.  Prosthetic hip replacements: plain films and arthrography for component loosening. , 1982, AJR. American journal of roentgenology.

[4]  L. S. Matthews,et al.  Sonic diagnosis of bone fracture healing--a preliminary study. , 1976, Journal of biomechanics.

[5]  K. Porter,et al.  The Charnley total hip arthroplasty in patients under age 40. , 1985, Clinical orthopaedics and related research.

[6]  W. Harris,et al.  Failed total hip replacement: assessment by plain radiographs, arthrograms, and aspiration of the hip joint. , 1984, The Journal of bone and joint surgery. American volume.

[7]  H. Horoszowski,et al.  Sequential use of technetium 99m MDP and gallium 67 citrate imaging in the evaluation of painful total hip replacement. , 1980, The British journal of radiology.

[8]  H H Doemland,et al.  Assessment of fracture healing by spectral analysis. , 1986, Journal of medical engineering & technology.

[9]  R. Mollan,et al.  Vibration arthrography as a diagnostic aid in diseases of the knee. A preliminary report. , 1987, The Journal of bone and joint surgery. British volume.