Are dialysis devices usable as ozone gas exchangers?

A study aimed to compare the efficiency of the ozone transfer of four hydrophilic dialysis filters, and one hydrophobic gas-exchange device (GED) has been performed. Obviously, the former should be specifically used only for dialysis. Unfortunately, some clinicians incautiously use them as GEDs. It has been shown that: (i) dialysis filters present a wide range of gas-exchange yield (from 0 up to 70%), often related to variability according to the treatment time; (ii) by scanning microscopy, it has been noticed that hollow fibers are somewhat altered by ozone; and (iii) because their constitutive materials may not be ozone-resistant, they may release toxic compounds harmful for the patients. On the contrary, the appropriate GED is ozone-transfer efficient, is ozone-resistant, and is suitable for blood autotransfusion and ozonation.

[1]  N. Di Paolo,et al.  Oxygenation-ozonation of blood during extracorporeal circulation: in vitro efficiency of a new gas exchange device. , 2007, Artificial organs.

[2]  M. Djafer,et al.  A novel membrane reactor for ozone water treatment , 2007 .

[3]  F. Kung,et al.  The effect of covalently bonded conjugated linoleic acid on the reduction of oxidative stress and blood coagulation for polysulfone hemodialyzer membrane. , 2006, International journal of biological macromolecules.

[4]  V. Bocci,et al.  Oxygenation-Ozonization of Blood During Extracorporeal Circulation (EBOO). Part III: A New Medical Approach , 2004 .

[5]  Andreas Pierratos,et al.  New approaches to hemodialysis. , 2004, Annual review of medicine.

[6]  E. Suzuki,et al.  Determination of Compounds Inhibiting Bacterial Growth in Sterilized Medical Devices , 2003 .

[7]  Jian Shen,et al.  Ozone-induced grafting phosphorylcholine polymer onto silicone film grafting 2-methacryloyloxyethyl phosphorylcholine onto silicone film to improve hemocompatibility , 2003 .

[8]  J. Curtis,et al.  The practical application of ozone in dialysis. , 2002, Nephrology news & issues.

[9]  W-C. Lin,et al.  The Grafting of Chitosan Oligomer to Polysulfone Membrane via Ozone-Treatment and its Effect on Anti-Bacterial Activity , 2002 .

[10]  M. Margaglione,et al.  Effects of ethylene oxide and steam sterilization on dialysis-induced cytokine release by cuprophan membrane. , 2002, Artificial organs.

[11]  N. Di Paolo,et al.  Ozonation of Blood during Extracorporeal Circulation II. Comparative Analysis of Several Oxygenator-Ozonators and Selection of One Type , 2001, The International journal of artificial organs.

[12]  N. Di Paolo,et al.  Ozonation of Blood during Extracorporeal Circulation. I. Rationale, Methodology and Preliminary Studies , 1999, The International journal of artificial organs.

[13]  R. Layman-Amato Disinfection of an RO system: clearing the issues , 1995 .

[14]  É. Kiss,et al.  Disinfection of Regenerated Dialyzers with Ozone , 1992, The International journal of artificial organs.

[15]  A. Brunel [Ozone therapy]. , 1972, Anuario Brasileiro De Odontologia.