Dosimetric characterization of an image-guided stereotactic small animal irradiator

Small animal irradiation provides an important tool used by preclinical studies to assess and optimize new treatment strategies such as stereotactic ablative radiotherapy. Characterization of radiation beams that are clinically and geometrically scaled for the small animal model is uniquely challenging for orthovoltage energies and minute field sizes. The irradiator employs a commercial x-ray device (XRAD 320, Precision x-ray, Inc.) with a custom collimation system to produce 1-10 mm diameter beams and a 50 mm reference beam. Absolute calibrations were performed using the AAPM TG-61 methodology. Beam's half-value layer (HVL) and timer error were measured with an ionization chamber. Percent depth dose (PDD), output factors (OFs) and off-axis ratios were measured using radiochromic film, a diode and a pinpoint ionization chamber at 19.76 and 24.76 cm source-to-surface distance (SSD). PDD measurements were also compared with Monte Carlo (MC) simulations. In-air and in-water absolute calibrations for the reference 50 mm diameter collimator at 19.76 cm SSD were measured as 20.96 and 20.79 Gy min(-1), respectively, agreeing within 0.8%. The HVL at 250 kVp and 15 mAs was measured to be 0.45 mm Cu. The reference field PDD MC simulation results agree with measured data within 3.5%. PDD data demonstrate typical increased penetration with increasing field size and SSD. For collimators larger than 5 mm in diameter, OFs measured using film, an ion chamber and a diode were within 3% agreement.

[1]  Lech Papiez,et al.  Excessive toxicity when treating central tumors in a phase II study of stereotactic body radiation therapy for medically inoperable early-stage lung cancer. , 2006, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[2]  Peter Kazanzides,et al.  Image-guided small animal radiation research platform: calibration of treatment beam alignment , 2009, Physics in medicine and biology.

[3]  Philip M Evans,et al.  Calculation of x-ray spectra emerging from an x-ray tube. Part I. electron penetration characteristics in x-ray targets. , 2007, Medical physics.

[4]  W. Franklin,et al.  Multi-institutional phase I/II trial of stereotactic body radiation therapy for liver metastases. , 2009, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[5]  Christopher H Contag,et al.  Design and evaluation of a variable aperture collimator for conformal radiotherapy of small animals using a microCT scanner. , 2007, Medical physics.

[6]  James C L Chow,et al.  Treatment planning for a small animal using Monte Carlo simulation. , 2007, Medical physics.

[7]  I. Lax,et al.  The role of radiotherapy in treatment of stage I non-small cell lung cancer. , 2003, Lung cancer.

[8]  David A. Jaffray,et al.  TH-C-BRC-06: Implementation of An Image-Guided System for Conformal Small Animal Irradiation: Characterization, QA, and Workflow , 2009 .

[9]  Timothy D Solberg,et al.  An Orthotopic Lung Tumor Model for Image-Guided Microirradiation in Rats , 2010, Radiation research.

[10]  Jaeho Cho,et al.  High dose-per-fraction irradiation of limited lung volumes using an image-guided, highly focused irradiator: simulating stereotactic body radiotherapy regimens in a small-animal model. , 2010, International journal of radiation oncology, biology, physics.

[11]  S Mutic,et al.  MicroRT-small animal conformal irradiator. , 2007, Medical physics.

[12]  Shankar Vallabhajosula,et al.  (18)F-labeled positron emission tomographic radiopharmaceuticals in oncology: an overview of radiochemistry and mechanisms of tumor localization. , 2007, Seminars in nuclear medicine.

[13]  Paul J Keall,et al.  Investigation of the effects of treatment planning variables in small animal radiotherapy dose distributions. , 2010, Medical physics.

[14]  Sasa Mutic,et al.  Progress toward a microradiation therapy small animal conformal irradiator. , 2006, Medical physics.

[15]  James C L Chow,et al.  Treatment planning for a small animal using Monte Carlo simulation. , 2007, Medical physics.

[16]  David A. Jaffray,et al.  SU‐GG‐J‐70: Development of An Image‐Guided Conformal Small Animal Irradiation Platform , 2008 .

[17]  S M Seltzer,et al.  AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology. , 2001, Medical physics.

[18]  Joseph O Deasy,et al.  Feasibility of small animal cranial irradiation with the microRT system. , 2008, Medical physics.

[19]  T D Solberg,et al.  An x-ray image guidance system for small animal stereotactic irradiation , 2010, Physics in medicine and biology.

[20]  J. Itami,et al.  Clinical outcomes of single‐fraction stereotactic radiation therapy of lung tumors , 2006, Cancer.

[21]  Erik Tryggestad,et al.  The small-animal radiation research platform (SARRP): dosimetry of a focused lens system , 2007, Physics in medicine and biology.

[22]  Effects of read-out light sources and ambient light on radiochromic film. , 1998, Physics in medicine and biology.

[23]  Rahil,et al.  Development of a MicroCT-Based Image-Guided Conformal Radiotherapy System for Small Animals , 2010 .

[24]  Andrea Bezjak,et al.  Stereotactic body radiation therapy for inoperable early stage lung cancer. , 2010, JAMA.

[25]  Eduardo G. Moros,et al.  TU‐C‐BRD‐03: An Integrated Robotic‐Based Irradiation System for Small Animal Research , 2009 .

[26]  M. Hiraoka,et al.  Stereotactic body radiation therapy for early-stage non-small-cell lung cancer: the Japanese experience , 2004, International Journal of Clinical Oncology.

[27]  Peter Kazanzides,et al.  Small Animal Radiation Research Platform: Imaging, Mechanics, Control and Calibration , 2007, MICCAI.

[28]  I. Lax,et al.  Stereotactic high dose fraction radiation therapy of extracranial tumors using an accelerator. Clinical experience of the first thirty-one patients. , 1995, Acta oncologica.

[29]  Paul J Keall,et al.  Kilovoltage beam Monte Carlo dose calculations in submillimeter voxels for small animal radiotherapy. , 2009, Medical physics.

[30]  M. Brundage,et al.  Involved-field radiotherapy alone for early-stage non-small-cell lung cancer. , 2000, International journal of radiation oncology, biology, physics.

[31]  Stephanie Frost,et al.  Extracranial stereotactic radioablation: results of a phase I study in medically inoperable stage I non-small cell lung cancer. , 2003, Chest.

[32]  Ping Li,et al.  Promising clinical outcome of stereotactic body radiation therapy for patients with inoperable Stage I/II non-small-cell lung cancer. , 2006, International journal of radiation oncology, biology, physics.

[33]  L. Gaspar,et al.  Multi-institutional phase I/II trial of stereotactic body radiation therapy for lung metastases. , 2009, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[34]  O Nalcioglu,et al.  Can dynamic contrast-enhanced MRI (DCE-MRI) predict tumor recurrence and lymph node status in patients with breast cancer? , 2007, Annals of oncology : official journal of the European Society for Medical Oncology.

[35]  F. Lohr,et al.  Stereotactic single-dose radiation therapy of liver tumors: results of a phase I/II trial. , 2001, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[36]  G. Poludniowski Calculation of x-ray spectra emerging from an x-ray tube. Part II. X-ray production and filtration in x-ray targets. , 2007, Medical physics.

[37]  Peter Kazanzides,et al.  High-resolution, small animal radiation research platform with x-ray tomographic guidance capabilities. , 2008, International journal of radiation oncology, biology, physics.

[38]  C Ross Schmidtlein,et al.  Precise radiochromic film dosimetry using a flat-bed document scanner. , 2005, Medical physics.

[39]  J. Debus,et al.  Stereotactic radiation therapy of liver metastases: update of the initial phase-I/II trial. , 2004, Frontiers of radiation therapy and oncology.

[40]  E Grabbe,et al.  In vivo imaging in experimental preclinical tumor research–A review , 2007, Cytometry. Part A : the journal of the International Society for Analytical Cytology.

[41]  D A Jaffray,et al.  Characterization of image quality and image-guidance performance of a preclinical microirradiator. , 2011, Medical physics.

[42]  Lei Xing,et al.  Stereotactic body radiation therapy in multiple organ sites. , 2007, Journal of clinical oncology : official journal of the American Society of Clinical Oncology.

[43]  J. Wong,et al.  Computed tomography-guided frameless stereotactic radiotherapy for stage I non-small cell lung cancer: a 5-year experience. , 2001, International journal of radiation oncology, biology, physics.