New method of estimating wavelength-dependent optical path length ratios for oxy- and deoxyhemoglobin measurement using near-infrared spectroscopy.

In near-infrared spectroscopy (NIRS), concentration changes in oxy- and deoxyhemoglobin are calculated using an attenuation change of the measurement light and by solving a linear equation based on the modified Lambert-Beer law. While solving this equation, we need to know the wavelength-dependent mean optical path lengths of the measurement lights. However, it is very difficult to know these values by a continuous-wave-type (CW-type) system. We propose a new method of estimating wavelength-dependent optical path length ratios of the measurement lights based on the data obtained by a triple wavelength CW-type NIRS instrument. The proposed method does not give a path length itself, but it gives a path length ratio. Thus, it is possible to obtain the accurate hemoglobin concentration changes without cross talk, although the method cannot contribute to the quantification of the absolute magnitude of hemoglobin changes. The method is based on the principle that two possible estimations of hemoglobin concentration changes calculated using a triple-wavelength measurement system should be identical. The method was applied to the experimental data of human subjects' foreheads. The estimated path length ratios were very similar to literature values obtained by using picosecond laser pulses and a streak camera detector [M. Essenpreis et al., Appl. Opt. 32(4), 418-425 (1993)].

[1]  David A. Boas,et al.  Factors affecting the accuracy of near-infrared spectroscopy concentration calculations for focal changes in oxygenation parameters , 2003, NeuroImage.

[2]  D. Delpy,et al.  Performance comparison of several published tissue near-infrared spectroscopy algorithms. , 1995, Analytical biochemistry.

[3]  D. Boas,et al.  Hemodynamic evoked response of the sensorimotor cortex measured noninvasively with near-infrared optical imaging. , 2003, Psychophysiology.

[4]  D. Delpy,et al.  Optical pathlength measurements on adult head, calf and forearm and the head of the newborn infant using phase resolved optical spectroscopy. , 1995, Physics in medicine and biology.

[5]  J. Hebden Advances in optical imaging of the newborn infant brain. , 2003, Psychophysiology.

[6]  A. Maki,et al.  Experimental prediction of the wavelength-dependent path-length factor for optical intrinsic signal analysis. , 2007, Applied optics.

[7]  Elizabeth M C Hillman,et al.  Optical brain imaging in vivo: techniques and applications from animal to man. , 2007, Journal of biomedical optics.

[8]  A. Villringer,et al.  Cross talk in the Lambert-Beer calculation for near-infrared wavelengths estimated by Monte Carlo simulations. , 2002, Journal of biomedical optics.

[9]  Yoko Hoshi,et al.  Functional near-infrared spectroscopy: current status and future prospects. , 2007, Journal of biomedical optics.

[10]  A. Villringer,et al.  Determination of the wavelength dependence of the differential pathlength factor from near-infrared pulse signals , 1998, Physics in medicine and biology.

[11]  A. Villringer,et al.  Beyond the Visible—Imaging the Human Brain with Light , 2003, Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism.

[12]  Y. Hoshi Functional near-infrared optical imaging: utility and limitations in human brain mapping. , 2003, Psychophysiology.

[13]  Huijuan Zhao,et al.  Maps of optical differential pathlength factor of human adult forehead, somatosensory motor and occipital regions at multi-wavelengths in NIR. , 2002, Physics in medicine and biology.

[14]  S. Arridge,et al.  Spectral Dependence of Temporal Point Spread Functions in Human Tissues , 2022 .