Observations at the articular surface of hip prostheses: an analytical electron microscopy study on wear and corrosion.

We used scanning electron microscopy in combination with X-ray microanalysis to evaluate Co-, Cr-, and Mo-based human femoral hip prostheses. In total, 23 retrieved implants and four new implants were included in this study. Scanning electron microscopy of the polished surface of all arthroplasties showed, in addition to the polishing marks, small round and angular holes or pits. Other types of surface irregularities were interpreted as wear or corrosion of the metal compound. In all cases studied, corrosion propagated from holes at the surface of the polished prosthesis heads, in some cases also along phase boundaries. X-ray microanalysis of the intact prosthetic surface showed a relative composition of the elements Co, Cr, and Mo, which was in agreement with the manufacturer's information (63:33:4%). However, X-ray microanalysis spot analysis of the surface holes showed deviation in the relative composition of the elements Co, Cr, and Mo and also the presence of Si, Ti, and Al. Furthermore, Ti and Al could be traced back at an artificially made fracture plane of a new prosthesis. Therefore, Ti and Al have to be present during the manufacturing process. These impurities in the metal prosthesis alloy may create a galvanic element with the Co, Cr, Mo alloy of the implant. If this is the case, such a galvanic element in combination with the electrolyte environment formed by body fluids, can induce galvanic corrosion with release of metal particles.

[1]  Joshua J. Jacobs,et al.  Corrosion of metal orthopaedic implants. , 1998, The Journal of bone and joint surgery. American volume.

[2]  M. Ritter,et al.  Radiolucency at the bone-cement interface in total knee replacement. The effects of bone-surface preparation and cement technique. , 1994, The Journal of bone and joint surgery. American volume.

[3]  J O Galante,et al.  Composition and morphology of wear debris in failed uncemented total hip replacement. , 1994, The Journal of bone and joint surgery. British volume.

[4]  C. Montero-Ocampo,et al.  Effect of carbon content on the resistance to localized corrosion of as-cast cobalt-based alloys in an aqueous chloride solution. , 1995, Journal of biomedical materials research.

[5]  W H Harris,et al.  Extensive localized bone resorption in the femur following total hip replacement. , 1976, The Journal of bone and joint surgery. American volume.

[6]  S. Brown,et al.  Release of hexavalent chromium from corrosion of stainless steel and cobalt-chromium alloys. , 1995, Journal of biomedical materials research.

[7]  H. Koerten,et al.  Inflammatory response and degradation of three types of calcium phosphate ceramic in a non-osseous environment. , 1994, Journal of biomedical materials research.

[8]  J O Galante,et al.  Endosteal erosion in association with stable uncemented femoral components. , 1990, The Journal of bone and joint surgery. American volume.

[9]  R. Levy,et al.  Effects of metallic ions and diphosphonates on inhibition of pericardial bioprosthetic tissue calcification and associated alkaline phosphatase activity. , 1993, Biomaterials.

[10]  A. Ferguson,et al.  The ionization of metal implants in living tissues. , 1960, The Journal of bone and joint surgery. American volume.

[11]  J. Galante,et al.  Metallosis associated with a stable titanium-alloy femoral component in total hip replacement. A case report. , 1990, The Journal of bone and joint surgery. American volume.

[12]  W. Harris,et al.  Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space. , 1992, The Journal of bone and joint surgery. American volume.

[13]  L. Dorr,et al.  Metal ion concentrations in retrieved polyethylene total hip inserts and implications for artifactually high readings in tissue. , 1993, Journal of biomedical materials research.