Self-Calibration of the Phase Angle Errors of RVDs at Frequencies Up to 100 kHz

A set of coaxial resistive voltage dividers (RVDs) with serial–parallel connection has been designed. A new method has also been proposed to calibrate phase angle errors of RVDs at frequencies up to 100 kHz. The method is mainly based on a special proportional relationship with the resistance and capacitance between two RVDs with identical structure, and the assumption is that the stray capacitance distribution is identical for both RVDs. The absolute phase angle error of an RVD can be derived from the phase difference between the two RVDs. The RVD with ratio of 100:1 consisting of 100-<inline-formula> <tex-math notation="LaTeX">$\Omega $ </tex-math></inline-formula> resistive elements has been built and self-calibrated. After the self-calibration of phase angle errors of the RVDs, a buffer has been introduced to the output of the RVD to reduce external loading error for the practical application. Finally, the standard uncertainty in phase angle error of the 100:1 RVD with buffer consisting of 100-<inline-formula> <tex-math notation="LaTeX">$\Omega $ </tex-math></inline-formula> resistive elements is within <inline-formula> <tex-math notation="LaTeX">$30~\mu $ </tex-math></inline-formula>rad at 100 kHz.

[1]  V. Tarasso,et al.  Towards an electric power standard for frequencies up to 1 MHz , 2012, 2012 Conference on Precision electromagnetic Measurements.

[2]  I. Budovsky,et al.  Voltage divider characterisation at frequencies up to 200 kHz , 2000, Conference on Precision Electromagnetic Measurements. Conference Digest. CPEM 2000 (Cat. No.00CH37031).

[3]  Karl-Erik Rydler,et al.  Precision measurement system for characterisation of phase displacement of voltage dividers up to 1 MHz , 2010, CPEM 2010.

[4]  U. Pogliano,et al.  Characterization of resistive dividers for a wideband power analyzer , 2014, 29th Conference on Precision Electromagnetic Measurements (CPEM 2014).

[5]  L. Lingziang,et al.  A new precision AC resistance divider , 1988 .

[6]  Lu Zuliang,et al.  Harmonic Power Standard at NIM and Its Compensation Algorithm , 2010, IEEE Transactions on Instrumentation and Measurement.

[7]  I. Budovsky,et al.  Development of a precision resistive voltage divider for frequencies up to 100 kHz , 2010, CPEM 2010.

[8]  S. Svensson,et al.  Voltage dividers with low phase angle errors for a wideband power measuring system , 2002, Conference Digest Conference on Precision Electromagnetic Measurements.

[9]  Ilya Budovsky Standard of Electrical Power at Frequencies Up to 200 kHz , 2009, IEEE Transactions on Instrumentation and Measurement.