Partial CO2 Rebreathing Cardiac Output – Operating Principles of the NICO™ System

The partial rebreathing method of cardiac output estimation is reviewed with a particular focus on its application for continuous monitoring, rebreathing and implementations and from both a historical and technical perspective. The assumptions of the method are discussed as well as the various implementations. The NICO monitor and rebreathing valve are described from a functional view. The clinical data including (a) comparisons between bolus thermodilution and continuous thermodilution in patients in the OR setting, (b) comparisons to continuous thermodilution with both the Baxter and Abbott continuous cardiac output devices and (c) comparison between different means of shunt correction are presented. Compared to conventional cardiac output methods, the partial CO2 rebreathing technique is non-invasive, can easily be automated and can provide real-time and continuous cardiac output monitoring. Taking advantage of modern sophisticated sensor and signal processing technology and integrating multiple monitored physiological variables the NICO monitor is the first commercially available cardiac output system making use of the partial rebreathing of CO2.

[1]  M. Ragaller,et al.  Partial carbon dioxide rebreathing: a reliable technique for noninvasive measurement of nonshunted pulmonary capillary blood flow. , 1997, Critical care medicine.

[2]  John F. Nunn,et al.  Applied Respiratory Physiology , 1977 .

[3]  R. Roy,et al.  Noninvasive measurement of cardiac output using partial CO/sub 2/ rebreathing , 1988, IEEE Transactions on Biomedical Engineering.

[4]  N. Ty Smith The Society for Technology in Anesthesia , 2005, Journal of Clinical Monitoring.

[5]  C. Stoutenbeek,et al.  Non-invasive pulmonary blood flow measurement by means of CO2 analysis of expiratory gases , 2005, Intensive Care Medicine.

[6]  C. Whitten,et al.  Ventilation, Thermal Noise, and Errors in Cardiac Output Measurements after Cardiopulmonary Bypass , 1993, Anesthesiology.

[7]  A. Gedeon Noninvasive pulmonary blood flow for optimal PEEP. , 1985, Clinical physiology.

[8]  D. G. Haryadi,et al.  Partial CO2 Rebreathing Indirect Fick Technique for Non-Invasive Measurement of Cardiac Output , 2004, Journal of Clinical Monitoring and Computing.

[9]  J. Jones,et al.  The PIO2 vs. SpO2 diagram: a non-invasive measure of pulmonary oxygen exchange. , 1995, European journal of anaesthesiology.

[10]  G. Hedenstierna,et al.  A new method for noninvasive bedside determination of pulmonary blood flow , 1980, Medical and Biological Engineering and Computing.

[11]  I. Gottlieb,et al.  Noninvasive cardiac output determined with a new method based on gas exchange measurements and carbon dioxide rebreathing: A study in animals/pigs , 1992, Journal of Clinical Monitoring.

[12]  R. Roy,et al.  Noninvasive measurement of cardiac output using partial CO2 rebreathing. , 1988, IEEE transactions on bio-medical engineering.

[13]  Sapsford Dj,et al.  The PIO2 vs. SpO2 diagram: a non-invasive measure of pulmonary oxygen exchange. , 1995 .

[14]  L. Goldman,et al.  The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. , 1996, JAMA.

[15]  L. Goldman,et al.  The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. , 1996, JAMA.

[16]  T. Raffin,et al.  Reliability of the thermodilution method in the determination of cardiac output in clinical practice. , 2015, The American review of respiratory disease.