Estimation of cerebral oxy- and deoxy-haemoglobin concentration changes in a layered adult head model using near-infrared spectroscopy and multivariate statistical analysis

The non-invasive measurement of cerebral oxy- (DeltaHbO(br)2) and deoxy-haemoglobin (DeltaHHb(br)) changes using near-infrared spectroscopy instruments is often affected by the absorption in the extracerebral layer. We have exploited the multivariate calibration (partial least squares, PLS) method to minimize the errors for a range of blood volume, oxygen saturation and extracerebral layer thicknesses. The changes in the mean time of flight of photons (Delta tau) and attenuation (DeltaA) on the surface of a 3D adult head model were simulated using a finite-element method based on the diffusion equation. The PLS was then performed to identify the optimal number of detectors, their positions and weightings, to optimize the estimation of DeltaHbO(br)2 and DeltaHHb(br). We define the 'nominal accuracy' as the accuracy of estimating DeltaHbO(br)2 and DeltaHHb(br) over a nominal range of extracerebral layer thicknesses and 'robustness' as the accuracy beyond the nominal range. The results showed that for one or two detectors, Delta tau performed better than DeltaA while using them together gave the best performance. When more detectors were used, the performances of using Delta tau, DeltaA or both together became comparable, showing that a larger number of detectors can compensate for the performance of a simple DeltaA measurement despite this measurement having a relatively lower sensitivity to intracerebral absorption changes.

[1]  D. Wertheim,et al.  NIRS in adults--effects of increasing optode separation. , 1994, Advances in experimental medicine and biology.

[2]  M. Copet,et al.  A Monte Carlo investigation of optical pathlength in inhomogeneous tissue and its application to near-infrared spectroscopy , 1993 .

[3]  Sergio Fantini,et al.  Semi-infinite-geometry boundary problem for light migration in highly scattering media: a frequency-domain study in the diffusion approximation , 1994 .

[4]  Shlomo Havlin,et al.  Photon Migration in a Two-layer Turbid Medium a Diffusion Analysis , 1992 .

[5]  M. Forina,et al.  Multivariate calibration. , 2007, Journal of chromatography. A.

[6]  S. Arridge,et al.  Spectral Dependence of Temporal Point Spread Functions in Human Tissues , 2022 .

[7]  C. Cooper,et al.  Absolute quantification of deoxyhaemoglobin concentration in tissue near infrared spectroscopy. , 1994, Physics in medicine and biology.

[8]  Eiji Okada,et al.  Theoretical and experimental investigation of the effect of sulci on light propagation in brain tissue , 1995, European Conference on Biomedical Optics.

[9]  P. Hopton,et al.  Measurement of cerebral blood volume using near-infrared spectroscopy and indocyanine green elimination. , 1999, Journal of applied physiology.

[10]  D T Delpy,et al.  Measurement of hemoglobin flow and blood flow by near-infrared spectroscopy. , 1993, Journal of applied physiology.

[11]  D. Delpy,et al.  Quantification in tissue near–infrared spectroscopy , 1997 .

[12]  M. Patterson,et al.  Noninvasive determination of the optical properties of two-layered turbid media , 1998 .

[13]  D. Delpy,et al.  Quantitation of cerebral blood volume in human infants by near-infrared spectroscopy. , 1990, Journal of applied physiology.

[14]  A. Kienle,et al.  In vivo determination of the optical properties of muscle with time-resolved reflectance using a layered model. , 1999, Physics in medicine and biology.

[15]  Richard S. J. Frackowiak,et al.  Cerebral blood flow, blood volume and oxygen utilization. Normal values and effect of age. , 1990, Brain : a journal of neurology.

[16]  D. Delpy,et al.  Near-infrared light propagation in an adult head model. II. Effect of superficial tissue thickness on the sensitivity of the near-infrared spectroscopy signal. , 2003, Applied optics.

[17]  B. Wilson,et al.  A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo. , 1992, Medical physics.

[18]  D T Delpy,et al.  In vivo measurements of the wavelength dependence of tissue-scattering coefficients between 760 and 900 nm measured with time-resolved spectroscopy. , 1997, Applied optics.

[19]  S R Arridge,et al.  An investigation of light transport through scattering bodies with non-scattering regions. , 1996, Physics in medicine and biology.

[20]  David T Delpy,et al.  A New Method for the Measurement of Cerebral Blood Volume and Total Circulating Blood Volume Using Near Infrared Spatially Resolved Spectroscopy and Indocyanine Green: Application and Validation in Neonates , 2004, Pediatric Research.

[21]  M. Schweiger,et al.  A finite element approach for modeling photon transport in tissue. , 1993, Medical physics.

[22]  S. A. Hill Non-invasive measurement of cerebral blood flow. , 1997, British journal of anaesthesia.

[23]  Yukio Kobayashi,et al.  Tissue oxygenation monitor using NIR spatially resolved spectroscopy , 1999, Photonics West - Biomedical Optics.

[24]  D. Delpy,et al.  Near-infrared light propagation in an adult head model. I. Modeling of low-level scattering in the cerebrospinal fluid layer. , 2003, Applied optics.

[25]  M. Patterson,et al.  Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[26]  Michael S Patterson,et al.  Haemoglobin oxygenation of a two-layer tissue-simulating phantom from time-resolved reflectance: effect of top layer thickness. , 2002, Physics in medicine and biology.

[27]  D. Delpy,et al.  Measurement of the optical properties of the adult human head with spatially resolved spectroscopy and changes of posture. , 2003, Advances in experimental medicine and biology.

[28]  A. Villringer,et al.  Determining changes in NIR absorption using a layered model of the human head , 2001, Physics in medicine and biology.

[29]  Heidrun Wabnitz,et al.  Bed-side assessment of cerebral perfusion in stroke patients based on optical monitoring of a dye bolus by time-resolved diffuse reflectance , 2005, NeuroImage.

[30]  A. Villringer,et al.  Noninvasive monitoring of cerebral blood flow by a dye bolus method: separation of brain from skin and skull signals. , 2002, Journal of biomedical optics.

[31]  A. Villringer,et al.  Time-resolved multidistance near-infrared spectroscopy of the adult head: intracerebral and extracerebral absorption changes from moments of distribution of times of flight of photons. , 2004, Applied optics.

[32]  D. Delpy,et al.  Quantification of adult cerebral hemodynamics by near-infrared spectroscopy. , 1994, Journal of applied physiology.

[33]  Sergio Fantini,et al.  Optical measurements of absorption changes in two-layered diffusive media. , 2004, Physics in medicine and biology.

[34]  S. D. Jong SIMPLS: an alternative approach to partial least squares regression , 1993 .

[35]  V. Ntziachristos,et al.  Non-invasive and quantitative near-infrared haemoglobin spectrometry in the piglet brain during hypoxic stress, using a frequency-domain multidistance instrument† , 2001, Physics in medicine and biology.