Quasidynamic calibration of stroboscopic scanning white light interferometer with a transfer standard

Abstract. A stroboscopic scanning white light interferometer (SSWLI) can characterize both static features and motion in micro(nano)electromechanical system devices. SSWLI measurement results should be linked to the meter definition to be comparable and unambiguous. This traceability is achieved by careful error characterization and calibration of the interferometer. The main challenge in vertical scale calibration is to have a reference device with reproducible out-of-plane movement. A piezo-scanned flexure guided stage with capacitive sensor feedback was attached to a mirror and an Invar steel holder with a reference plane—forming a transfer standard that was calibrated by laser interferometry with 2.3 nm uncertainty. The moving mirror vertical position was then measured with the SSWLI, relative to the reference plane, between successive mirror position steppings. A light-emitting diode pulsed at 100 Hz with 0.5% duty cycle synchronized to the CCD camera and a halogen light source were used. Inside the scanned 14 μm range, the measured SSWLI scale amplification coefficient error was 0.12% with 4.5 nm repeatability of the steps. For SWLI measurements using a halogen lamp, the corresponding results were 0.05% and 6.7 nm. The presented methodology should permit accurate traceable calibration of the vertical scale of any SWLI.

[1]  Carlo Ferri,et al.  Variability in measurements of micro lengths with a white light interferometer , 2008, Qual. Reliab. Eng. Int..

[2]  T J Quinn,et al.  Practical realization of the definition of the metre, including recommended radiations of other optical frequency standards (2001) , 2003 .

[3]  Emmanuel Bruno Jean Paul Brousseau,et al.  Calibration of a white light interferometer for the measurement of micro-scale dimensions , 2010 .

[4]  E Hæggström,et al.  Stroboscopic supercontinuum white-light interferometer for MEMS characterization. , 2012, Optics letters.

[5]  Jari Turunen,et al.  Elementary-field representations in partially coherent optics , 2011 .

[6]  Edward Hæggström,et al.  Effect of LED spectral shift on vertical resolution in stroboscopic white light interferometry , 2008, Photonics Europe.

[7]  Mikko Merimaa,et al.  Traceability of Laser Frequency Calibrations at MIKES , 2007, IEEE Transactions on Instrumentation and Measurement.

[8]  R Felder,et al.  Practical realization of the definition of the metre, including recommended radiations of other optical frequency standards (2003) , 2005 .

[9]  Antti Lassila,et al.  Design and performance of an advanced metrology building for MIKES , 2011 .

[10]  Kieran G. Larkin,et al.  Efficient nonlinear algorithm for envelope detection in white light interferometry , 1996 .

[11]  Erkki Ikonen,et al.  Interferometers for calibration of length standards , 1995 .

[12]  Peter de Groot,et al.  Stroboscopic white-light interference microscopy. , 2006, Applied optics.

[13]  J C Wyant,et al.  Offset of coherent envelope position due to phase change on reflection. , 2001, Applied optics.

[14]  J C Wyant,et al.  Fringe modulation skewing effect in white-light vertical scanning interferometry. , 2000, Applied optics.

[15]  N. Gershenfeld Physics of the Future: How Science Will Shape Human Destiny and Our Daily Lives by the Year 2100 , 2011 .

[16]  Claudiu L Giusca,et al.  Calibration of the scales of areal surface topography measuring instruments: part 2. Amplification, linearity and squareness , 2012 .

[17]  Hanne Ludvigsen,et al.  Scanning white-light interferometry with a supercontinuum source. , 2009, Optics letters.

[18]  Edward Hæggström,et al.  Stroboscopic scanning white light interferometry at 2.7 MHz with 1.6 µm coherence length using a non-phosphor LED source. , 2013, Optics express.

[19]  Kent Robertson Van Horn,et al.  Design and application , 1967 .

[20]  S. W. Kim,et al.  Compensation of phase change on reflection in white-light interferometry for step height measurement. , 2001, Optics letters.

[21]  Y. N. Ning,et al.  Three-wavelength combination source for white-light interferometry , 1993, IEEE Photonics Technology Letters.

[22]  A. Nolvi,et al.  High-speed stroboscopic imaging with frequency-doubled supercontinuum , 2013, 2013 Conference on Lasers & Electro-Optics Europe & International Quantum Electronics Conference CLEO EUROPE/IQEC.

[23]  A Lassila,et al.  Calibration of a commercial AFM: traceability for a coordinate system , 2007 .

[24]  Mikko Merimaa,et al.  A method for linearization of a laser interferometer down to the picometre level with a capacitive sensor , 2011 .