The jPET-D4: detector calibration and acquisition system of the 4-layer DOI-PET scanner

We are developing a high-performance brain PET scanner, JPET-D4. This scanner is designed to achieve not only high spatial resolution but also high sensitivity using four-layered depth-of-interaction (DOI) detectors. The scanner has five detector rings with the ring diameter of 390 mm and each detector ring consists of 24 DOI detectors. In order to identify the crystal of interaction and its energy from PMT output, we have previously proposed the DOI detector calibration method. This method creates position and energy look-up-tables (LUTs) from the uniform irradiation. The position LUT converts location of 2D position histograms to crystal addresses. The energy LUT corrects light output of each crystal. Five detector rings have been installed and calibrated. The average energy resolution for 120 DOI detectors is optimized to 17% plusmn 1.4%. The scatter fraction for this system is 38% with energy window of 400-600 keV. The sensitivity for point source is 112 kcps/MBq with a 250 keV LLD

[1]  H. Murayama,et al.  Energy spectra and their calibration of four-layer DOI detector for brain PET scanner: jPET-D4 , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[2]  H. Murayama,et al.  Three-dimensional array of scintillation crystals with proper reflector arrangement for a depth of interaction detector , 2003, IEEE Transactions on Nuclear Science.

[3]  Klaus Wienhard,et al.  The ECAT HRRT: performance and first clinical application of the new high resolution research tomograph , 2000 .

[4]  H. Haneishi,et al.  Preliminary performance evaluation of the prototype system for a brain DOI-PET scanner: jPET-D4 , 2004, IEEE Symposium Conference Record Nuclear Science 2004..

[5]  H. Murayama,et al.  A performance study of the next generation PET using a new histogramming method for the DOI detector , 2003, 2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515).

[6]  H. Murayama,et al.  Calibration procedure for a DOI detector of high resolution PET through a Gaussian mixture model , 2004, IEEE Transactions on Nuclear Science.

[7]  H. Kawai,et al.  Performance of a PET detector with a 256ch flat panel PS-PMT , 2002 .

[8]  M. Daube-Witherspoon,et al.  Performance of a brain PET camera based on anger-logic gadolinium oxyorthosilicate detectors. , 2003, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[9]  T. Kosugi,et al.  A new high resolution PET scanner dedicated to brain research , 2001 .