Hip simulator wear testing according to the newly introduced standard ISO 14242

Abstract Ultra-high molecular weight polyethylene (UHMWPE) acetabular cups were tested against alumina-ceramic femoral heads using a new type of hip joint simulator according to ISO/FDIS 14242-1. Bovine serum as well as newborn calf serum were used as test fluids. Total polyethylene wear was determined by weight loss of the cups. In addition, wear depth and its distribution were recorded by means of a coordinate measurement system. Wear particle analysis and inspection of the worn polyethylene surfaces using light and scanning electron microscopy (SEM) were performed to analyse damage and identify the acting wear mechanisms. The total wear rate was determined to be 22.07 ± 1.75 mg/106 cycles for the bovine serum group and 26.57 ± 3.55 mg/106 cycles for the calf serum group. Unexpectedly, the formation of two wear vectors corresponding to recent clinical findings was detected. Retrieved polyethylene wear debris was comparable in size and shape with clinical findings. The test method described by ISO/FDIS 14242-1 produced reliable and reproducible wear data using UHMWPE acetabular cups articulating against alumina-ceramic heads. In the authors' opinion, the lubricant composition should be described in more detail, since the protein and additive content seem to have a high impact on the wear results. It needs to be emphasized that the findings of this study cannot be regarded as a general validation of hip wear tests according to ISO/FDIS 14242-1 but are limited to the material combinations investigated herein. Further testing of other clinically relevant materials and interlaboratory ring tests must follow.

[1]  I C Clarke,et al.  Charnley wear model for validation of hip simulators' ball diameter versus polytetrafluoroethylene and polyethylene wear , 1997, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[2]  I. Clarke,et al.  Clinical and hip simulator comparisons of ceramic-on-polyethylene and metal-on-polyethylene wear. , 2000, Clinical orthopaedics and related research.

[3]  H. Mckellop,et al.  Effect of protein lubrication on the wear properties of materials for prosthetic joints. , 1999, Journal of biomedical materials research.

[4]  V O Saikko,et al.  Wear of polyethylene acetabular cups against alumina femoral heads. 5 prostheses compared in a hip simulator for 35 million walking cycles. , 1993, Acta orthopaedica Scandinavica.

[5]  H. Mckellop,et al.  Frictional heating of bearing materials tested in a hip joint wear simulator , 1997, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[6]  D Dowson,et al.  A multi-station hip joint simulator study of the performance of 22 mm diameter zirconia-ultra-high molecular weight polyethylene total replacement hip joints , 1999, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[7]  L. Zichner,et al.  In-vivo-Verschleiß der Gleitpaarungen Keramik-Polyethylen gegen Metall-Polyethylen , 1997, Der Orthopäde.

[8]  J Fisher,et al.  Wear of ultra-high molecular weight polyethylene acetabular cups in a physiological hip joint simulator in the anatomical position using bovine serum as a lubricant , 1997, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[9]  J O Galante,et al.  Composition and morphology of wear debris in failed uncemented total hip replacement. , 1994, The Journal of bone and joint surgery. British volume.

[10]  T. Schmalzried,et al.  The multifactorial nature of polyethylene wear in vivo. , 1998, The Journal of bone and joint surgery. American volume.

[11]  L. Zichner,et al.  [In-vivo wear of the slide combinations ceramics-polyethylene as opposed to metal-polyethylene]. , 1997, Der Orthopade.

[12]  T P Andriacchi,et al.  A striated pattern of wear in ultrahigh-molecular-weight polyethylene components of Miller-Galante total knee arthroplasty. , 1998, The Journal of arthroplasty.

[13]  G. Selvik,et al.  Socket wear assessment. A comparison of three different radiographic methods. , 1993, The Journal of arthroplasty.

[14]  Scott,et al.  WEAR PARTICLE ANALYSES OF CONVENTIONAL AND CROSSLINKED UHMWPE TESTED IN AN ANATOMIC HIP SIMULATOR , 2001 .

[15]  A Sarmiento,et al.  The origin of submicron polyethylene wear debris in total hip arthroplasty. , 1995, Clinical orthopaedics and related research.

[16]  W H Harris,et al.  Analysis of the kinematics of different hip simulators used to study wear of candidate materials for the articulation of total hip arthroplasties , 1998, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[17]  A. Wang,et al.  Comparison of the size and morphology of UHMWPE wear debris produced by a hip joint simulator under serum and water lubricated conditions. , 1996, Biomaterials.

[18]  M. Ritter,et al.  Radiolucency at the bone-cement interface in total knee replacement. The effects of bone-surface preparation and cement technique. , 1994, The Journal of bone and joint surgery. American volume.

[19]  P. Campbell,et al.  Isolation of predominantly submicron-sized UHMWPE wear particles from periprosthetic tissues. , 1995, Journal of biomedical materials research.

[20]  T D Brown,et al.  The Frank Stinchfield Award. 3-Dimensional sliding/contact computational simulation of total hip wear. , 1996, Clinical orthopaedics and related research.

[21]  T. Bauer,et al.  Three-Dimensional Analysis of Multiple Wear Vectors in Retrieved Acetabular Cups* , 1997, The Journal of bone and joint surgery. American volume.

[22]  J Fisher,et al.  A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates , 1999, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[23]  C. M. Agrawal,et al.  Isolation and characterization of polyethylene wear debris associated with osteolysis following total shoulder arthroplasty. , 1999, The Journal of bone and joint surgery. American volume.

[24]  V. Saikko,et al.  Type of motion and lubricant in wear simulation of polyethylene acetabular cup , 1999, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[25]  P Slätis,et al.  A Five-Station Hip Joint Simulator for Wear Rate Studies , 1992, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[26]  I. Clarke,et al.  Hip-simulator ranking of polyethylene wear: comparisons between ceramic heads of different sizes. , 1996, Acta orthopaedica Scandinavica.

[27]  T. Bauer,et al.  Polyethylene wear vector in vivo: A three‐dimensional analysis using retrieved acetabular components and radiographs , 1999, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[28]  C. Bragdon,et al.  The Importance of Multidirectional Motion on the Wear of Polyethylene , 1996, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.