The extension of the NIST BRDF scale from 1100 nm to 2500 nm

Measurements of bi-directional reflectance factor for diffuse reflectance from 1100 nm to 2500 nm using extended-range indium gallium arsenide (exInGaAs) detectors in the NIST Spectral Tri-function Automated Reference Reflectometer (STARR) facility are described. The determination of bi-directional reflectance factor with low uncertainties requires the exInGaAs radiometer to be characterized for low-noise performance, linearity and spatial uniformity. The instrument characterizations will be used to establish a total uncertainty budget for the reflectance factor. To independently check the bi-directional reflectance factors, measurements also were made in a separate facility in which the reflectance factor is determined using calibrated spectral irradiance and radiance standards. The total combined uncertainties for the diffuse reflectances range from < 1 % at 1100 nm to 2.5 % at 2500 nm. At NIST, these measurement capabilities will evolve into a calibration service for diffuse spectral reflectance in this wavelength region.