Ballistocardiogram as Proximal Timing Reference for Pulse Transit Time Measurement: Potential for Cuffless Blood Pressure Monitoring

Goal: We tested the hypothesis that the ballistocardiogram (BCG) waveform could yield a viable proximal timing reference for measuring pulse transit time (PTT). Methods: From 15 healthy volunteers, we measured PTT as the time interval between BCG and a noninvasively measured finger blood pressure (BP) waveform. To evaluate the efficacy of the BCG-based PTT in estimating BP, we likewise measured pulse arrival time (PAT) using the electrocardiogram (ECG) as proximal timing reference and compared their correlations to BP. Results: BCG-based PTT was correlated with BP reasonably well: the mean correlation coefficient (r ) was 0.62 for diastolic (DP), 0.65 for mean (MP), and 0.66 for systolic (SP) pressures when the intersecting tangent method was used as distal timing reference. Comparing four distal timing references (intersecting tangent, maximum second derivative, diastolic minimum, and systolic maximum), PTT exhibited the best correlation with BP when the systolic maximum method was used (mean r value was 0.66 for DP, 0.67 for MP, and 0.70 for SP). PTT was more strongly correlated with DP than PAT regardless of the distal timing reference: mean r value was 0.62 versus 0.51 (p = 0.07) for intersecting tangent, 0.54 versus 0.49 (p = 0.17) for maximum second derivative, 0.58 versus 0.52 (p = 0.37) for diastolic minimum, and 0.66 versus 0.60 (p = 0.10) for systolic maximum methods. The difference between PTT and PAT in estimating DP was significant (p = 0.01) when the r values associated with all the distal timing references were compared altogether. However, PAT appeared to outperform PTT in estimating SP ( p = 0.31 when the r values associated with all the distal timing references were compared altogether). Conclusion: We conclude that BCG is an adequate proximal timing reference in deriving PTT, and that BCG-based PTT may be superior to ECG-based PAT in estimating DP. Significance: PTT with BCG as proximal timing reference has potential to enable convenient and ubiquitous cuffless BP monitoring.

[1]  Survi Kyal,et al.  Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice , 2015, IEEE Transactions on Biomedical Engineering.

[2]  H MANDELBAUM,et al.  Studies Utilizing the Portable Electromagnetic Ballistocardiograph: IV. The Clinical Significance of Serial Ballistocardiograms Following Acute Myocardial Infarction , 1953, Circulation.

[3]  B. Morgenstern,et al.  Human blood pressure determination by sphygmomanometry. , 1993, Circulation.

[4]  Qiao Zhang,et al.  Noninvasive cuffless blood pressure estimation using pulse transit time and Hilbert-Huang transform , 2013, Comput. Electr. Eng..

[5]  C. Douniama,et al.  Blood pressure tracking capabilities of pulse transit times in different arterial segments: A clinical evaluation , 2009, 2009 36th Annual Computers in Cardiology Conference (CinC).

[6]  L Giovangrandi,et al.  Non-invasive cardiac output trending during exercise recovery on a bathroom-scale-based ballistocardiograph , 2009, Physiological measurement.

[7]  Alberto Avolio,et al.  Determination of Aortic Pulse Wave Velocity From Waveform Decomposition of the Central Aortic Pressure Pulse , 2008, Hypertension.

[8]  J W Gordon,et al.  Certain Molar Movements of the Human Body produced by the Circulation of the Blood. , 1877, Journal of anatomy and physiology.

[9]  C MackDavid,et al.  Development and preliminary validation of heart rate and breathing rate detection using a passive, ballistocardiography-based sleep monitoring system , 2009 .

[10]  Jeffrey Wood,et al.  Using Individualized Pulse Transit Time Calibration to Monitor Blood Pressure During Exercise , 2013, ICIMTH.

[11]  Heather T. Ma,et al.  A Blood Pressure Monitoring Method for Stroke Management , 2014, BioMed research international.

[12]  B. Alpert,et al.  Oscillometric blood pressure: a review for clinicians. , 2014, Journal of the American Society of Hypertension : JASH.

[13]  D Shapiro,et al.  Pulse transit time and blood pressure: an intensive analysis. , 1983, Psychophysiology.

[14]  J. Jalonen,et al.  Static-charge-sensitive bed ballistocardiography in cardiovascular monitoring. , 1996, Clinical physiology.

[15]  Guanqun Zhang,et al.  Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressure. , 2011, Journal of applied physiology.

[16]  T. Togawa,et al.  Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration , 2000, Medical and Biological Engineering and Computing.

[17]  X. Aubert,et al.  Cuffless Estimation of Systolic Blood Pressure for Short Effort Bicycle Tests: The Prominent Role of the Pre-Ejection Period , 2006, 2006 International Conference of the IEEE Engineering in Medicine and Biology Society.

[18]  Gilwon Yoon,et al.  Non-constrained Blood Pressure Monitoring Using ECG and PPG for Personal Healthcare , 2009, Journal of Medical Systems.

[19]  John G. Wood,et al.  MCDONALDʼS BLOOD FLOW IN ARTERIES: THEORETICAL, EXPERIMENTAL AND CLINICAL PRINCIPLES, 4TH EDITION , 1998 .

[20]  Kouhyar Tavakolian,et al.  Ballistocardiography and Seismocardiography: A Review of Recent Advances , 2015, IEEE Journal of Biomedical and Health Informatics.

[21]  A Mathewson,et al.  An examination of calibration intervals required for accurately tracking blood pressure using pulse transit time algorithms , 2013, Journal of Human Hypertension.

[22]  Norbert Noury,et al.  A review of methods for non-invasive and continuous blood pressure monitoring: Pulse transit time method is promising? , 2014 .

[23]  A. Patzak,et al.  Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method , 2011, European Journal of Applied Physiology.

[24]  Richard M. Wiard,et al.  Robust ballistocardiogram acquisition for home monitoring , 2009, Physiological measurement.

[25]  M. A. Younessi Heravi,et al.  Continuous and Cuffless Blood Pressure Monitoring Based on ECG and SpO2 Signals ByUsing Microsoft Visual C Sharp , 2014, Journal of biomedical physics & engineering.

[26]  Roman Cmejla,et al.  Parameters for mean blood pressure estimation based on electrocardiography and photoplethysmography , 2011, 2011 International Conference on Applied Electronics.

[27]  Ramakrishna Mukkamala,et al.  Improved pulse transit time estimation by system identification analysis of proximal and distal arterial waveforms. , 2011, American journal of physiology. Heart and circulatory physiology.

[28]  O. Chételat,et al.  Parametric estimation of pulse arrival time: a robust approach to pulse wave velocity , 2009, Physiological measurement.

[29]  T. Schaeffter,et al.  A Technical Assessment of Pulse Wave Velocity Algorithms Applied to Non-invasive Arterial Waveforms , 2013, Annals of Biomedical Engineering.

[30]  R. Talley,et al.  Evaluation of the pre-ejection period as an estimate of myocardial contractility in dogs. , 1971, The American journal of cardiology.

[31]  Y.T. Zhang,et al.  A Novel Calibration Method for Noninvasive Blood Pressure Measurement Using Pulse Transit Time , 2007, 2007 4th IEEE/EMBS International Summer School and Symposium on Medical Devices and Biosensors.

[32]  Roozbeh Jafari,et al.  BioWatch — A wrist watch based signal acquisition system for physiological signals including blood pressure , 2014, 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[33]  J. D. Bourland,et al.  Measurement of pulse-wave velocity using a beat-sampling technique , 2006, Annals of Biomedical Engineering.

[34]  Alan Bernjak,et al.  Pulse transit times to the capillary bed evaluated by laser Doppler flowmetry , 2009, Physiological measurement.

[35]  M. Masè,et al.  Feasibility of cuff-free measurement of systolic and diastolic arterial blood pressure. , 2011, Journal of electrocardiology.

[36]  P. Obrist,et al.  Pulse transit time: relationship to blood pressure and myocardial performance. , 1979, Psychophysiology.

[37]  Carmen C. Y. Poon,et al.  An Evaluation of the Cuffless Blood Pressure Estimation Based on Pulse Transit Time Technique: a Half Year Study on Normotensive Subjects , 2009, Cardiovascular engineering.

[38]  R. Payne,et al.  Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure. , 2006, Journal of applied physiology.

[39]  I. Starr,et al.  STUDIES ON THE ESTIMATION OF CARDIAC OUPTUT IN MAN, AND OF ABNORMALITIES IN CARDIAC FUNCTION, FROM THE HEART'S RECOIL AND THE BLOOD'S IMPACTS; THE BALLISTOCARDIOGRAM , 1939 .

[40]  Kiichi Sagawa,et al.  Effect of nitroglycerin on aortic impedance, diameter, and pulse-wave velocity. , 1988 .

[41]  G. V. Marie,et al.  The relationship between arterial blood pressure and pulse transit time during dynamic and static exercise. , 1984, Psychophysiology.

[42]  C. Ahlstrom,et al.  Noninvasive investigation of blood pressure changes using the pulse wave transit time: a novel approach in the monitoring of hemodialysis patients , 2005, Journal of Artificial Organs.

[43]  N Westerhof,et al.  Measurement of pulse wave velocity using a beat-sampling technique. , 1989, Annals of biomedical engineering.

[44]  Nicholas Bari Olivier,et al.  Improved Pulse Wave Velocity Estimation Using an Arterial Tube-Load Model , 2014, IEEE Transactions on Biomedical Engineering.

[45]  S. J. Wilson,et al.  Clinical applications of pulse transit time in paediatric critical care , 2009, Journal of medical engineering & technology.

[46]  G. Parati,et al.  Wearable Seismocardiography , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[47]  Carmen C. Y. Poon,et al.  Cuff-less and Noninvasive Measurements of Arterial Blood Pressure by Pulse Transit Time , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[48]  Federico S. Cattivelli,et al.  Noninvasive Cuffless Estimation of Blood Pressure from Pulse Arrival Time and Heart Rate with Adaptive Calibration , 2009, 2009 Sixth International Workshop on Wearable and Implantable Body Sensor Networks.

[49]  Gregory T. A. Kovacs,et al.  Rapid Assessment of Cardiac Contractility on a Home Bathroom Scale , 2011, IEEE Transactions on Information Technology in Biomedicine.

[50]  L. Geddes,et al.  Measurements of Young's Modulus of Elasticity of the Canine Aorta with Ultrasound , 1979 .

[51]  A. Varri,et al.  A new method for measuring the ballistocardiogram using EMFi sensors in a normal chair , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[52]  Jin-Oh Hahn,et al.  Estimation of pulse transit time using two diametric blood pressure waveform measurements. , 2010, Medical engineering & physics.

[53]  A. Weissler,et al.  Effects of adrenergic receptor activation and blockade on the systolic preejection period, heart rate, and arterial pressure in man. , 1967, The Journal of clinical investigation.

[54]  O. Zaderykhin,et al.  Adaptive algorithm for continuous monitoring of blood pressure using a pulse transit time , 2013, 2013 IEEE XXXIII International Scientific Conference Electronics and Nanotechnology (ELNANO).

[55]  Gi-Ryon Kim,et al.  Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time , 2010 .

[56]  James T. Patrie,et al.  Development and Preliminary Validation of Heart Rate and Breathing Rate Detection Using a Passive, Ballistocardiography-Based Sleep Monitoring System , 2009, IEEE Transactions on Information Technology in Biomedicine.