Glenoid loosening after total shoulder arthroplasty: An in vitro CT‐scan study

Glenoid fixation failure has only been grossly characterized. This lack of information hinders attempts to improve fixation because of a lack of methodologies for detecting and monitoring fixation failure. Our goal was twofold: to collect detailed data of glenoid fixation fracture, and to investigate computed tomography (CT)‐scanning as a tool for investigations of fixation failure. Six cadaver scapulas and six bone‐substitute specimens were cyclically loaded and CT‐scanned at clinical settings after 0, 1,000, 5,000, 10,000, 30,000, 50,000 and 70,000 load cycles. The fixation status was evaluated by inspection of the scans. After 70,000 cycles, the specimens were sectioned, and the fixation inspected by microscopy. The results of the microscopy analysis were compared to the CT‐scan analysis. Fracture of the glenoid fixation initiated at the edge of the glenoid rim and propagated towards and around the keel of the implant. The entire process from initiation to complete fracture took place at the polyethylene implant–cement interface, while the cement, the adjacent bone, and the cement–bone interface remained intact. Thus, strengthening the polyethylene–cement interface should improve glenoid fixation. Microscopy results validated the CT methodology, suggesting that the CT technique is reliable. © 2009 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 27:1589–1595, 2009

[1]  F A Matsen,et al.  Glenoid loosening in total shoulder arthroplasty. Association with rotator cuff deficiency. , 1988, The Journal of arthroplasty.

[2]  R. Cofield,et al.  Total shoulder arthroplasty with the Neer prosthesis: long-term results. , 1997, Journal of shoulder and elbow surgery.

[3]  P J Prendergast,et al.  Stress analysis of glenoid component designs for shoulder arthroplasty , 1997, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[4]  G. Walch,et al.  Primary Gleno-humeral Osteoarthritis: Clinical and Radiographic Classification , 1999 .

[5]  B F Morrey,et al.  Stress analyses of glenoid components in total shoulder arthroplasty. , 1999, Journal of shoulder and elbow surgery.

[6]  D R Pichora,et al.  Mechanical testing of shoulder prostheses and recommendations for glenoid design. , 2000, Journal of shoulder and elbow surgery.

[7]  U P Wyss,et al.  Loosening performance of cemented glenoid prosthesis design pairs. , 2001, Clinical biomechanics.

[8]  R. Darmana,et al.  Finite element analysis of the mechanical behavior of a scapula implanted with a glenoid prosthesis. , 2001, Clinical biomechanics.

[9]  P J Prendergast,et al.  Contact stresses in the glenoid component in total shoulder arthroplasty , 2003, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[10]  C Gerber,et al.  Influence of peg design and cement mantle thickness on pull-out strength of glenoid component pegs. , 2003, The Journal of bone and joint surgery. British volume.

[11]  F. V. D. van der Helm,et al.  Stress analysis of cemented glenoid prostheses in total shoulder arthroplasty. , 2004, Journal of biomechanics.

[12]  C. Gerber,et al.  Radiographic and computed tomography analysis of cemented pegged polyethylene glenoid components in total shoulder replacement. , 2005, The Journal of bone and joint surgery. American volume.

[13]  T. Wright,et al.  Glenoid loosening in response to dynamic multi-axis eccentric loading: a comparison between keeled and pegged designs with an equivalent radial mismatch. , 2006, Bulletin (Hospital for Joint Diseases (New York, N.Y.)).

[14]  Kamal I. Bohsali,et al.  Complications of total shoulder arthroplasty. , 2006, The Journal of bone and joint surgery. American volume.

[15]  P Mansat,et al.  Evaluation of the glenoid implant survival using a biomechanical finite element analysis: influence of the implant design, bone properties, and loading location. , 2007, Journal of shoulder and elbow surgery.

[16]  Mark Taylor,et al.  Wear in the prosthetic shoulder: association with design parameters. , 2007, Journal of biomechanical engineering.