A ten-year follow-up of the Reflection cementless acetabular component.

We reviewed the long-term results at ten to 12 years of 118 total hip replacements in 109 patients using a second-generation hemispherical cementless acetabular component (Reflection) designed to address the problem of backside wear. Five patients (five hips) died and six patients (seven hips) were lost to follow-up. The remaining 98 patients (106 hips) had a mean age of 62.9 years (34.0 to 86.2) A rate of revision for aseptic loosening of 0.9%, and predictable results were found with respect to radiological evidence of fixation, lack of pain, walking ability, range of movement and function. One component was revised for aseptic loosening, and of the 101 hips (95.2%) that did not have a revision, minor osteolytic lesions of the pelvis were seen in six (5.9%). Kaplan-Meier survival analysis for the total cohort of 118 hips revealed a 96.4% survival at both ten (95% confidence interval 90 to 98) and 12 years (95% confidence interval 86 to 98).

[1]  S. MacDonald,et al.  The value of anteroposterior pelvic radiographs for evaluating pelvic osteolysis. , 2007, Clinical orthopaedics and related research.

[2]  C. Engh,et al.  Second-generation Porous-coated Cementless Total Hip Arthroplasties Have High Survival , 2006, Clinical orthopaedics and related research.

[3]  E. Salvati,et al.  Backside Wear Is Low in Retrieved Modern, Modular, and Nonmodular Acetabular Liners , 2005, Clinical orthopaedics and related research.

[4]  H. Rubash,et al.  The acetabular insert-metal backing interface: an additional source of polyethylene wear debris. , 2005, The Journal of arthroplasty.

[5]  E. Salvati,et al.  Wear and periprosthetic osteolysis in a match-paired study of modular and nonmodular uncemented acetabular cups. , 2004, The Journal of arthroplasty.

[6]  M. Peterson,et al.  Clinical and radiographic results associated with a modern, cementless modular cup design in total hip arthroplasty. , 2004, The Journal of bone and joint surgery. American volume.

[7]  D. Pedersen,et al.  Cementless acetabular fixation at fifteen years. A comparison with the same surgeon's results following acetabular fixation with cement. , 2004, The Journal of bone and joint surgery. American volume.

[8]  W. Harris Results of Uncemented Cups: A Critical Appraisal at 15 Years , 2003, Clinical orthopaedics and related research.

[9]  L. Dorr,et al.  Cementless Hemispheric Porous-Coated Sockets Implanted with Press-Fit Technique without Screws: Average Ten-Year Follow-up , 2002, The Journal of bone and joint surgery. American volume.

[10]  P. Lachiewicz,et al.  Survival and Polyethylene Wear of Porous-Coated Acetabular Components in Patients Less than Fifty Years Old: Results at Nine to Fourteen Years , 2002, The Journal of bone and joint surgery. American volume.

[11]  C. Engh,et al.  Pattern of osteolysis around two different cementless metal-backed cups: retrospective, radiographic analysis at minimum 10-year follow-up. , 2001, The Journal of arthroplasty.

[12]  C. Engh,et al.  Characterization of Long-Term Femoral-Head-Penetration Rates: Association with and Prediction of Osteolysis* , 2000, The Journal of bone and joint surgery. American volume.

[13]  C. Ranawat,et al.  Clinical and radiographic evaluation of the Harris-Galante Cup , 2000 .

[14]  J J Callaghan,et al.  Wear in total hip and knee replacements. , 2000, The Journal of bone and joint surgery. American volume.

[15]  T. Fehring,et al.  Motion at the modular acetabular shell and liner interface. A comparative study. , 1999, Clinical orthopaedics and related research.

[16]  J. Kabo,et al.  Ultra-high-molecular weight polyethylene wear: an in vitro comparison of acetabular metal types and polished surfaces. , 1999, The Journal of arthroplasty.

[17]  C. V. White,et al.  Simulation of initial frontside and backside wear rates in a modular acetabular component with multiple screw holes. , 1999, Journal of biomechanics.

[18]  W. Harris,et al.  Primary hybrid total hip replacement, performed with insertion of the acetabular component without cement and a precoat femoral component with cement. An average ten-year follow-up study. , 1999, The Journal of bone and joint surgery. American volume.

[19]  C. V. White,et al.  Backside nonconformity and locking restraints affect liner/shell load transfer mechanisms and relative motion in modular acetabular components for total hip replacement. , 1998, Journal of biomechanics.

[20]  J. Galante,et al.  Primary Total Hip Arthroplasty with a Porous-Coated Acetabular Component. Seven-to-Ten-Year Results* , 1997, The Journal of bone and joint surgery. American volume.

[21]  L. Whiteside,et al.  Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups. , 1997, The Journal of arthroplasty.

[22]  J. Kabo,et al.  Wear of the polyethylene liner-metallic shell interface in modular acetabular components. An in vitro analysis. , 1996, The Journal of arthroplasty.

[23]  R. Poss,et al.  The impact of modularity in total hip arthroplasty. , 1995, Clinical orthopaedics and related research.

[24]  C. Colwell,et al.  Polyethylene wear debris in modular acetabular prostheses. , 1995, Clinical orthopaedics and related research.

[25]  C. Rimnac,et al.  Polyethylene and metal debris generated by non-articulating surfaces of modular acetabular components. , 1994, The Journal of bone and joint surgery. British volume.

[26]  J. Bono,et al.  Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML PLUS hip implants with an ACS polyethylene liner. , 1994, The Journal of arthroplasty.

[27]  M. Ritter,et al.  Polyethylene wear in uncemented acetabular components. , 1994, The Journal of bone and joint surgery. British volume.

[28]  Michael Tanzer,et al.  Concerns with modularity in total hip arthroplasty. , 1994, Clinical orthopaedics and related research.

[29]  A. Greenwald,et al.  A comparison of the disassociation strength of modular acetabular components. , 1993, Clinical orthopaedics and related research.

[30]  D W Murray,et al.  Survival analysis of joint replacements. , 1993, The Journal of bone and joint surgery. British volume.

[31]  C. Engh,et al.  Evaluation of cementless acetabular component migration. An experimental study. , 1989, The Journal of arthroplasty.

[32]  J. Charnley,et al.  Radiological demarcation of cemented sockets in total hip replacement. , 1976, Clinical orthopaedics and related research.