Three-dimensional tracking of cardiac catheters using an inverse geometry x-ray fluoroscopy system.

PURPOSE Scanning beam digital x-ray (SBDX) is an inverse geometry fluoroscopic system with high dose efficiency and the ability to perform continuous real-time tomosynthesis at multiple planes. This study describes a tomosynthesis-based method for 3D tracking of high-contrast objects and present the first experimental investigation of cardiac catheter tracking using a prototype SBDX system. METHODS The 3D tracking algorithm utilizes the stack of regularly spaced tomosynthetic planes that are generated by SBDX after each frame period (15 frames/s). Gradient-filtered versions of the image planes are generated, the filtered images are segmented into object regions, and then a 3D coordinate is calculated for each object region. Two phantom studies of tracking performance were conducted. In the first study, an ablation catheter in a chest phantom was imaged as it was pulled along a 3D trajectory defined by a catheter sheath (10, 25, and 50 mm/s pullback speeds). SBDX tip tracking coordinates were compared to the 3D trajectory of the sheath as determined from a CT scan of the phantom after the registration of the SBDX and CT coordinate systems. In the second study, frame-to-frame tracking precision was measured for six different catheter configurations as a function of image noise level (662-7625 photons/mm2 mean detected x-ray fluence at isocenter). RESULTS During catheter pullbacks, the 3D distance between the tracked catheter tip and the sheath centerline was 1.0 +/- 0.8 mm (mean +/- one standard deviation). The electrode to centerline distances were comparable to the diameter of the catheter tip (2.3 mm), the confining sheath (4 mm outside diameter), and the estimated SBDX-to-CT registration error (+/- 0.7 mm). The tip position was localized for all 332 image frames analyzed and 83% of tracked positions were inside the 3D sheath volume derived from CT. The pullback speeds derived from the catheter trajectories were within 5% of the programed pullback speeds. The tracking precision of ablation and diagnostic catheter tips ranged from +/- 0.2 mm at the highest image fluence to +/- 0.9 mm at the lowest fluence. Tracking precision depended on image fluence, the size of the tracked catheter electrode, and the contrast of the electrode. CONCLUSIONS High speed multiplanar tomosynthesis with an inverse geometry x-ray fluoroscopy system enables 3D tracking of multiple high-contrast objects at the rate of fluoroscopic imaging. The SBDX system is capable of tracking electrodes in standard cardiac catheters with approximately 1 mm accuracy and precision.

[1]  J Clémenty,et al.  Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. , 1998, The New England journal of medicine.

[2]  R Haberl,et al.  A dynamic approach to identifying desired physiological phases for cardiac imaging using multislice spiral CT. , 2003, Medical physics.

[3]  Edward G. Solomon,et al.  Scanning-beam digital x-ray (SBDX) system for cardiac angiography , 1999, Medical Imaging.

[4]  L. Gepstein,et al.  A novel method for nonfluoroscopic catheter-based electroanatomical mapping of the heart. In vitro and in vivo accuracy results. , 1997, Circulation.

[5]  Michael A Speidel,et al.  Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography. , 2006, Medical physics.

[6]  F. Mettler,et al.  Skin injuries from fluoroscopically guided procedures: part 1, characteristics of radiation injury. , 2001, AJR. American journal of roentgenology.

[7]  Michael A Speidel,et al.  Comparison of entrance exposure and signal-to-noise ratio between an SBDX prototype and a wide-beam cardiac angiographic system. , 2006, Medical physics.

[8]  Michael A. Speidel,et al.  Frame-by-frame 3D catheter tracking methods for an inverse geometry cardiac interventional system , 2008, SPIE Medical Imaging.

[9]  Gabriele Vicedomini,et al.  Circumferential Radiofrequency Ablation of Pulmonary Vein Ostia A New Anatomic Approach for Curing Atrial Fibrillation , 2000 .

[10]  Sabine Ernst,et al.  Complete Isolation of Left Atrium Surrounding the Pulmonary Veins: New Insights From the Double-Lasso Technique in Paroxysmal Atrial Fibrillation , 2004, Circulation.

[11]  Kaleem Siddiqi,et al.  Flux driven fly throughs , 2003, 2003 IEEE Computer Society Conference on Computer Vision and Pattern Recognition, 2003. Proceedings..

[12]  Michael A. Speidel,et al.  Performance assessment of the scanning beam digital x-ray (SBDX) system , 1997, Medical Imaging.

[13]  H. Halperin,et al.  Integrated Electroanatomic Mapping With Three-Dimensional Computed Tomographic Images for Real-Time Guided Ablations , 2006, Circulation.

[14]  D. Taussky,et al.  Clinical implementation of a digital tomosynthesis-based seed reconstruction algorithm for intraoperative postimplant dose evaluation in low dose rate prostate brachytherapy. , 2009, Medical physics.

[15]  Masayuki Zuguchi,et al.  Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. , 2006, AJR. American journal of roentgenology.

[16]  J. Ector,et al.  Obesity is a major determinant of radiation dose in patients undergoing pulmonary vein isolation for atrial fibrillation. , 2007, Journal of the American College of Cardiology.

[17]  A. Lowell,et al.  Real-time X-ray tomosynthesis imaging using an ATCA general-purpose data acquisition and analysis platform , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[18]  Mark E. Josephson,et al.  Nonfluoroscopic, in vivo navigation and mapping technology , 1996, Nature Medicine.