Basis function cross‐correlations for Robust k‐space sample density compensation, with application to the design of radiofrequency excitations

The problem of k‐space sample density compensation is restated as the normalization of the independent information that can be expressed by the ensemble of Fourier basis functions corresponding to the trajectory. Specifically, multiple samples (complex exponential functions) may be contributing to each independent information element (independent basis function). Normalization can be accomplished by solving a linear system based on the cross‐correlation matrix of the underlying Fourier basis functions. The solution to this system is straightforward and can be obtained without resorting to discretization since the cross‐correlations of Fourier basis functions are analytically known. Furthermore, no restrictions are placed on the k‐space trajectory and its point‐spread function. Additionally, the linear system can be used to elucidate key trade‐offs involved in k‐space trajectory design. The approach can be used to compensate samples acquired for image reconstruction or designed for low flip angle radiofrequency (RF) excitation. Here it is demonstrated for the latter application, using reversed spiral trajectories. In this case the linear system approach enables one to easily incorporate additional constraints such as smoothness to the solution. For typical RF excitation durations (<20 ms) it is shown that density compensation can even be achieved without numerical iteration. Magn Reson Med 57:338–352, 2007. © 2007 Wiley‐Liss, Inc.

[1]  P Börnert,et al.  On the performance and accuracy of 2D navigator pulses. , 1999, Magnetic resonance imaging.

[2]  G H Glover,et al.  Simple analytic spiral K‐space algorithm , 1999, Magnetic resonance in medicine.

[3]  Douglas C Noll,et al.  Iterative RF pulse design for multidimensional, small‐tip‐angle selective excitation , 2005, Magnetic resonance in medicine.

[4]  Peter Börnert,et al.  Spatial excitation using variable‐density spiral trajectories , 2003, Journal of magnetic resonance imaging : JMRI.

[5]  Andres Guesalaga,et al.  Fast three‐dimensional k‐space trajectory design using missile guidance ideas , 2004, Magnetic resonance in medicine.

[6]  Frank J Rybicki,et al.  Strategies for inner volume 3D fast spin echo magnetic resonance imaging using nonselective refocusing radio frequency pulses. , 2005, Medical physics.

[7]  Nathaniel E. Helwig,et al.  An Introduction to Linear Algebra , 2006 .

[8]  Peter Jezzard,et al.  Selective arterial spin labeling (SASL): Perfusion territory mapping of selected feeding arteries tagged using two‐dimensional radiofrequency pulses , 2003, Magnetic resonance in medicine.

[9]  A. Macovski,et al.  Selection of a convolution function for Fourier inversion using gridding [computerised tomography application]. , 1991, IEEE transactions on medical imaging.

[10]  Dwight G Nishimura,et al.  Fast 3D imaging using variable‐density spiral trajectories with applications to limb perfusion , 2003, Magnetic resonance in medicine.

[11]  J. Pauly,et al.  Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm [NMR imaging]. , 1991, IEEE transactions on medical imaging.

[12]  Volker Rasche,et al.  Resampling of data between arbitrary grids using convolution interpolation , 1999, IEEE Transactions on Medical Imaging.

[13]  Jeffrey A. Fessler,et al.  Nonuniform fast Fourier transforms using min-max interpolation , 2003, IEEE Trans. Signal Process..

[14]  A Macovski,et al.  Multifrequency interpolation for fast off‐resonance correction , 1997, Magnetic resonance in medicine.

[15]  D. Hoult The solution of the bloch equations in the presence of a varying B1 field—An approach to selective pulse analysis , 1979 .

[16]  B K Rutt,et al.  Effects of polar sampling in k‐space , 1996, Magnetic resonance in medicine.

[17]  L P Panych,et al.  Selection of high‐definition 2D virtual profiles with multiple RF pulse excitations along interleaved echo‐planar k‐space trajectories , 1999, Magnetic resonance in medicine.

[18]  Christopher J. Hardy,et al.  Two‐dimensional spatially selective spin inversion and spin‐echo refocusing with a single nuclear magnetic resonance pulse , 1987 .

[19]  Christopher J. Hardy,et al.  Correcting for nonuniform k-space sampling in two-dimensional NMR selective excitation , 1990 .

[20]  Jens Frahm,et al.  Two‐dimensional spatially‐selective RF excitation pulses in echo‐planar imaging , 2002, Magnetic resonance in medicine.

[21]  R. Hoge,et al.  Density compensation functions for spiral MRI , 1997, Magnetic resonance in medicine.

[22]  J. D. O'Sullivan,et al.  A Fast Sinc Function Gridding Algorithm for Fourier Inversion in Computer Tomography , 1985, IEEE Transactions on Medical Imaging.

[23]  D N Firmin,et al.  Locally focused 3D coronary imaging using volume‐selective RF excitation , 1999, Magnetic resonance in medicine.

[24]  J. Pipe,et al.  Sampling density compensation in MRI: Rationale and an iterative numerical solution , 1999, Magnetic resonance in medicine.

[25]  Dwight G Nishimura,et al.  Time‐optimal multidimensional gradient waveform design for rapid imaging , 2004, Magnetic resonance in medicine.

[26]  Jeffrey L. Stuart Linear Algebra, 3rd ed./Linear Algebra and its Applications, 3rd ed./Linear Algebra: A Geometric Approach/Introduction to Linear Algebra, 3rd ed , 2005 .

[27]  Yousef Saad,et al.  Iterative methods for sparse linear systems , 2003 .

[28]  Douglas C Noll,et al.  Small tip angle three‐dimensional tailored radiofrequency slab‐select pulse for reduced B1 inhomogeneity at 3 T , 2005, Magnetic resonance in medicine.

[29]  Albert Macovski,et al.  A linear class of large-tip-angle selective excitation pulses , 1989 .

[30]  Henry Stark,et al.  Spiral sampling: theory and an application to magnetic resonance imaging , 1988 .

[31]  René M. Botnar,et al.  3D coronary vessel wall imaging utilizing a local inversion technique with spiral image acquisition , 2001, Magnetic resonance in medicine.

[32]  Douglas C Noll,et al.  Multishot 3D slice‐select tailored RF pulses for MRI , 2002, Magnetic resonance in medicine.

[33]  A. Takahashi,et al.  Compensation of multi‐dimensional selective excitation pulses using measured k‐space trajectories , 1995, Magnetic resonance in medicine.

[34]  Jiarui Lin,et al.  Reconstruction of MR images from data acquired on an arbitrary k‐space trajectory using the same‐image weight , 2002, Magnetic resonance in medicine.