Clinical comparison of automatic, noninvasive measurements of blood pressure in the forearm and upper arm.

BACKGROUND When the upper arm (area from shoulder to elbow) is inaccessible and/or a standard-sized blood pressure cuff does not fit, some healthcare workers use the forearm to measure blood pressure. OBJECTIVE To compare automatic noninvasive measurements of blood pressure in the upper arm and forearm. METHODS A descriptive, correlational comparison study was conducted in the emergency department of a 1071-bed teaching hospital. Subjects were 204 English-speaking patients 6 to 91 years old in medically stable condition who had entered the department on foot or by wheelchair and who had no exclusions to using their left upper extremity. A Welch Allyn Vital Signs 420 series monitor was used to measure blood pressure in the left upper arm and forearm with the subject seated and the upper arm or forearm at heart level. RESULTS Pearson r correlation coefficients between measurements in the upper arm and forearm were 0.88 for systolic blood pressure and 0.76 for diastolic blood pressure (P < .001 for both). Mean systolic pressures, but not mean diastolic pressures, in the upper arm and forearm differed significantly (t = 2.07, P = .04). A Bland-Altman analysis indicated that the distances between the mean values and the limits of agreement for the 2 sites ranged from 15 mm Hg (mean arterial pressure) to 18.4 mm Hg (systolic pressure). CONCLUSIONS Despite strict attention to correct cuff size and placement of the upper arm or forearm at heart level, measurements of blood pressure obtained noninvasively in the arm and forearm of seated patients in stable condition are not interchangeable.

[1]  G Parati,et al.  Blood pressure measuring devices: recommendations of the European Society of Hypertension , 2001, BMJ : British Medical Journal.

[2]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[3]  Ulrich Tholl,et al.  Measuring blood pressure: pitfalls and recommendations. , 2004, Nephrology, dialysis, transplantation : official publication of the European Dialysis and Transplant Association - European Renal Association.

[4]  B. Morgenstern,et al.  Human blood pressure determination by sphygmomanometry. , 1993, Circulation.

[5]  Julie Waterhouse,et al.  Clinical comparison of automatic, noninvasive measurements of blood pressure in the forearm and upper arm with the patient supine or with the head of the bed raised 45 degrees: a follow-up study. , 2006, American journal of critical care : an official publication, American Association of Critical-Care Nurses.

[6]  A S Berson,et al.  National Standard for Measurement of Resting and Ambulatory Blood Pressures With Automated Sphygmomanometers , 1993, Hypertension.

[7]  T. Thien,et al.  Body position and blood pressure measurement in patients with diabetes mellitus , 2002, Journal of internal medicine.

[8]  J. Graves,et al.  Prevalence of blood pressure cuff sizes in a referral practice of 430 consecutive adult hypertensives , 2001, Blood pressure monitoring.

[9]  C. Teague,et al.  Evaluation of an instrument for noninvasive blood pressure monitoring in the forearm. , 1996, Biomedical Instrumentation & Technology.

[10]  W. Flemons,et al.  Measuring agreement between diagnostic devices. , 2003, Chest.

[11]  E. Friedmann,et al.  A review of nursing research on blood pressure. , 2002, Journal of nursing scholarship : an official publication of Sigma Theta Tau International Honor Society of Nursing.

[12]  J. Cozad The Practice of Nursing Research: Conduct, Critique and Utilization , 1989 .

[13]  Lawrence J Appel,et al.  Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. , 2005, Hypertension.

[14]  N S Latman,et al.  Evaluation of instruments for noninvasive blood pressure monitoring of the wrist. , 1997, Biomedical instrumentation & technology.

[15]  M. Hill,et al.  How to take a precise blood pressure. , 1991, The American journal of nursing.

[16]  B. Tachovsky,et al.  Indirect auscultatory blood pressure measurement at two sites in the arm. , 1985, Research in nursing & health.

[17]  A. Térent,et al.  Epidemiological perspective of body position and arm level in blood pressure measurement. , 1994, Blood pressure.

[18]  K. Dobbin,et al.  Noninvasive blood pressure monitoring. , 1998, Critical care nurse.

[19]  N. Burns,et al.  The Practice of Nursing Research: Conduct, Critique, & Utilization , 1987 .

[20]  Comparison of forearm and upper arm blood pressures. , 1999, Prehospital emergency care : official journal of the National Association of EMS Physicians and the National Association of State EMS Directors.

[21]  R. Netea,et al.  Influence of body and arm position on blood pressure readings: an overview. , 2003, Journal of Hypertension.

[22]  W. Elliott,et al.  Recommendations for Blood Pressure Measurement in Humans and Experimental Animals: Part 1. Blood Pressure Measurement in Humans. A Statement for Professionals From the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure ResearchPickering , 2006 .

[23]  S. Allison,et al.  Power analysis. , 2004, The Journal of orthopaedic and sports physical therapy.