Left Ventricular and Myocardial Perfusion Responses to Volume Unloading and Afterload Reduction in a Computer Simulation

Ventricular assist devices (VADs) have been used successfully as a bridge to transplant in heart failure patients by unloading ventricular volume and restoring the circulation. In a few cases, patients have been successfully weaned from these devices after myocardial recovery. To promote myocardial recovery and alleviate the demand for donor organs, we are developing an artificial vasculature device (AVD) that is designed to allow the heart to fill to its normal volume but eject against a lower afterload. Using this approach, the heart ejects its stroke volume (SV) into an AVD anastomosed to the aortic arch, which has been programmed to produce any desired afterload condition defined by an input impedance profile. During diastole, the AVD returns this SV to the aorta, providing counterpulsation. Dynamic computer models of each of the assist devices (AVD, continuous, and pulsatile flow pumps) were developed and coupled to a model of the cardiovascular system. Computer simulations of these assist techniques were conducted to predict physiologic responses. Hemodynamic parameters, ventricular pressure-volume loops, and vascular impedance characteristics were calculated with AVD, continuous VAD, and asynchronous pulsatile VAD support for a range of clinical cardiac conditions (normal, failing, and recovering left ventricle). These simulation results indicate that the AVD may provide better coronary perfusion, as well as lower vascular resistance and elastance seen by the native heart during ejection compared with continuous and pulsatile VAD. Our working hypothesis is that by controlling afterload using the AVD approach, ventricular cannulation can be eliminated, myocardial perfusion improved, myocardial compliance and resistance restored, and effective weaning protocols developed that promote myocardial recovery.

[1]  Daniel L. Ewert,et al.  HEART: an automated beat-to-beat cardiovascular analysis package using Matlab® , 2004, Comput. Biol. Medicine.

[2]  G. Giridharan,et al.  Control strategy for maintaining physiological perfusion with rotary blood pumps. , 2003, Artificial organs.

[3]  A. Abolhoda,et al.  Left Ventricular Outflow Tract Obstruction Associated with Chronic Ventricular Assist Device Support , 2002, ASAIO journal.

[4]  Mikhail Skliar,et al.  Nonlinear controller for ventricular assist devices. , 2002, Artificial organs.

[5]  M. Yacoub,et al.  Minimally invasive explantation of implantable left ventricular assist devices. , 2002, The Journal of thoracic and cardiovascular surgery.

[6]  Mikhail Skliar,et al.  Modeling and Control of a Brushless DC Axial Flow Ventricular Assist Device , 2002, ASAIO journal.

[7]  Guruprasad A. Giridharan,et al.  Modeling of the human circulatory system with an axial flow ventricular assist device , 2001, Proceedings of the 2001 American Control Conference. (Cat. No.01CH37148).

[8]  D. Burkhoff,et al.  Time course of reverse remodeling of the left ventricle during support with a left ventricular assist device. , 2001, The Journal of thoracic and cardiovascular surgery.

[9]  G. Noon,et al.  Regression of fibrosis and hypertrophy in failing myocardium following mechanical circulatory support. , 2001, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[10]  P. McCarthy,et al.  Left ventricular assist device bridge to recovery: a review of the current status. , 2001, The Annals of thoracic surgery.

[11]  R. Hetzer,et al.  Bridging-to-recovery. , 2001, The Annals of thoracic surgery.

[12]  R. Lingle,et al.  Left ventricular assist device as a bridge to patient and myocardial recovery. , 2000, Progress in cardiovascular diseases.

[13]  M C Oz,et al.  Left ventricular assist device-induced reverse ventricular remodeling. , 2000, Progress in cardiovascular diseases.

[14]  C. Klersy,et al.  Results with the Novacor assist system and evaluation of long-term assistance. , 2000, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[15]  S Murabayashi,et al.  Development of rotary blood pump technology: past, present, and future. , 2000, Artificial organs.

[16]  M E Debakey The odyssey of the artificial heart. , 2000, Artificial organs.

[17]  V. Poirier LVADs--a new era in patient care. , 2000, The Journal of cardiovascular management : the official journal of the American College of Cardiovascular Administrators.

[18]  G. Noon,et al.  Decreased expression of tumor necrosis factor-alpha in failing human myocardium after mechanical circulatory support : A potential mechanism for cardiac recovery. , 1999, Circulation.

[19]  D. Harrison,et al.  Increased superoxide in heart failure: a biochemical baroreflex gone awry. , 1999, Circulation.

[20]  D B Olsen Sixth International Symposium for Rotary Blood Pumps. , 1999, Artificial organs.

[21]  T. Mussivand Mechanical Circulatory Devices for the Treatment of Heart Failure , 1985, Journal of cardiac surgery.

[22]  Guruprasad A. Giridharan,et al.  Controller Design for Ventricular Assist Devices , 1999 .

[23]  M Flannery,et al.  Low incidence of myocardial recovery after left ventricular assist device implantation in patients with chronic heart failure. , 1998, Circulation.

[24]  S. Houser,et al.  Myocyte recovery after mechanical circulatory support in humans with end-stage heart failure. , 1998, Circulation.

[25]  L. Tripp,et al.  Phrenic Nerve Injury Following Cardiac Surgery: A Review , 1998, Journal of cardiac surgery.

[26]  J. Wallwork,et al.  Left ventricular assist devices: current status and future applications. , 1997, International journal of cardiology.

[27]  R Hetzer,et al.  Weaning from mechanical cardiac support in patients with idiopathic dilated cardiomyopathy. , 1997, Circulation.

[28]  M C Oz,et al.  Reversal of chronic ventricular dilation in patients with end-stage cardiomyopathy by prolonged mechanical unloading. , 1995, Circulation.

[29]  S D Moulopoulos,et al.  Counterpulsation: historical background, technical improvements, hemodynamic and metabolic effects. , 1994, Cardiology.