Ultra-high spatial resolution multi-energy CT using photon counting detector technology

Two ultra-high-resolution (UHR) imaging modes, each with two energy thresholds, were implemented on a research, whole-body photon-counting-detector (PCD) CT scanner, referred to as sharp and UHR, respectively. The UHR mode has a pixel size of 0.25 mm at iso-center for both energy thresholds, with a collimation of 32 × 0.25 mm. The sharp mode has a 0.25 mm pixel for the low-energy threshold and 0.5 mm for the high-energy threshold, with a collimation of 48 × 0.25 mm. Kidney stones with mixed mineral composition and lung nodules with different shapes were scanned using both modes, and with the standard imaging mode, referred to as macro mode (0.5 mm pixel and 32 × 0.5 mm collimation). Evaluation and comparison of the three modes focused on the ability to accurately delineate anatomic structures using the high-spatial resolution capability and the ability to quantify stone composition using the multi-energy capability. The low-energy threshold images of the sharp and UHR modes showed better shape and texture information due to the achieved higher spatial resolution, although noise was also higher. No noticeable benefit was shown in multi-energy analysis using UHR compared to standard resolution (macro mode) when standard doses were used. This was due to excessive noise in the higher resolution images. However, UHR scans at higher dose showed improvement in multi-energy analysis over macro mode with regular dose. To fully take advantage of the higher spatial resolution in multi-energy analysis, either increased radiation dose, or application of noise reduction techniques, is needed.

[1]  J. Schlomka,et al.  Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography , 2008, Physics in medicine and biology.

[2]  Karl Stierstorfer,et al.  Photon counting CT at elevated X-ray tube currents: contrast stability, image noise and multi-energy performance , 2014, Medical Imaging.

[3]  J. Lingeman,et al.  Micro-computed tomography for analysis of urinary calculi , 2010, Urological Research.

[4]  Amir Pourmorteza,et al.  Abdominal Imaging with Contrast-enhanced Photon-counting CT: First Human Experience. , 2016, Radiology.

[5]  K. Taguchi,et al.  Vision 20/20: Single photon counting x-ray detectors in medical imaging. , 2013, Medical physics.

[6]  P. Shikhaliev Energy-resolved computed tomography: first experimental results , 2008, Physics in medicine and biology.

[7]  Edgar Kraft,et al.  A research prototype system for quantum-counting clinical CT , 2010, Medical Imaging.

[8]  Baiyu Chen,et al.  Evaluation of conventional imaging performance in a research whole-body CT system with a photon-counting detector array , 2016, Physics in medicine and biology.

[9]  Polad M Shikhaliev,et al.  Photon counting multienergy x-ray imaging: effect of the characteristic x rays on detector performance. , 2009, Medical physics.

[10]  P. Shikhaliev Computed tomography with energy-resolved detection: a feasibility study , 2008, Physics in medicine and biology.

[11]  D. Maeding,et al.  Preliminary results obtained from a novel CdZnTe pad detector and readout ASIC developed for an automatic baggage inspection system , 2000, 2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149).

[12]  Katharina Hahn,et al.  Dose-efficient ultrahigh-resolution scan mode using a photon counting detector computed tomography system , 2016, Journal of medical imaging.

[13]  P. Shikhaliev,et al.  Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application , 2011, Physics in medicine and biology.

[14]  E. Nygard,et al.  Photon counting energy dispersive detector arrays for x-ray imaging , 2007, 2007 IEEE Nuclear Science Symposium Conference Record.

[15]  Polad M Shikhaliev Photon counting spectral CT: improved material decomposition with K-edge-filtered x-rays. , 2012, Physics in medicine and biology.