Evaluation of a new VMAT QA device, or the “X” and “O” array geometries

We introduce a logical process of three distinct phases to begin the evaluation of a new 3D dosimetry array. The array under investigation is a hollow cylinder phantom with diode detectors fixed in a helical shell forming an “O” axial detector cross section (ArcCHECK), with comparisons drawn to a previously studied 3D array with diodes fixed in two crossing planes forming an “X” axial cross section (Delta 4 ). Phase I testing of the ArcCHECK establishes: robust relative calibration (response equalization) of the individual detectors, minor field size dependency of response not present in a 2D predecessor, and uncorrected angular response dependence in the axial plane. Phase II testing reveals vast differences between the two devices when studying fixed‐width full circle arcs. These differences are primarily due to arc discretization by the TPS that produces low passing rates for the peripheral detectors of the ArcCHECK, but high passing rates for the Delta 4 . Similar, although less pronounced, effects are seen for the test VMAT plans modeled after the AAPM TG119 report. The very different 3D detector locations of the two devices, along with the knock‐on effect of different percent normalization strategies, prove that the analysis results from the devices are distinct and noninterchangeable; they are truly measuring different things. The value of what each device measures, namely their correlation with – or ability to predict – clinically relevant errors in calculation and/or delivery of dose is the subject of future Phase III work. PACS number: 87.55Qr

[1]  Stefano Gianolini,et al.  Introducing gel dosimetry in a clinical environment: customization of polymer gel composition and magnetic resonance imaging parameters used for 3D dose verifications in radiosurgery and intensity modulated radiotherapy. , 2007, Medical physics.

[2]  P. Jursinic,et al.  A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery. , 2003, Medical physics.

[3]  Geoffrey G. Zhang,et al.  Initial dosimetric evaluation of SmartArc – a novel VMAT treatment planning module implemented in a multi‐vendor delivery chain , 2010, Journal of applied clinical medical physics.

[4]  Vladimir Feygelman,et al.  Evaluation of a 3D diode array dosimeter for helical tomotherapy delivery QA. , 2010, Medical Dosimetry.

[5]  Timothy C Zhu,et al.  Energy dependence of commercially available diode detectors for in-vivo dosimetry. , 2007, Medical physics.

[6]  Z Yin,et al.  Response corrections for solid-state detectors in megavoltage photon dosimetry. , 2004, Physics in medicine and biology.

[7]  Dietrich Harder,et al.  Spatial resolution of 2D ionization chamber arrays for IMRT dose verification: single-detector size and sampling step width , 2007, Physics in medicine and biology.

[8]  Clive Baldock,et al.  An experimental investigation into the radiation field offset of a dynamic multileaf collimator , 2006, Physics in medicine and biology.

[9]  D. Low,et al.  A technique for the quantitative evaluation of dose distributions. , 1998, Medical physics.

[10]  R. Kollhoff,et al.  Two-dimensional ionization chamber arrays for IMRT plan verification. , 2006, Medical physics.

[11]  W. Simon,et al.  Total scatter factors and tissue maximum ratios for small radiosurgery fields: comparison of diode detectors, a parallel-plate ion chamber, and radiographic film. , 2000, Medical physics.

[12]  M. Karlsson,et al.  Semi-conductor detectors in output factor measurements. , 1997, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[13]  Bo Lu,et al.  Calibration of a novel four-dimensional diode array. , 2009, Medical physics.

[14]  James A. Purdy,et al.  Intensity-modulated radiotherapy: current status and issues of interest. , 2001, International journal of radiation oncology, biology, physics.

[15]  J. Dempsey,et al.  Evaluation of polymer gels and MRI as a 3-D dosimeter for intensity-modulated radiation therapy. , 1999, Medical physics.

[16]  Lars E Olsson,et al.  MAGIC-type polymer gel for three-dimensional dosimetry: intensity-modulated radiation therapy verification. , 2003, Medical physics.

[17]  Lei Dong,et al.  Retrospective analysis of 2D patient-specific IMRT verifications. , 2005, Medical physics.

[18]  A. Boyer,et al.  Tissue Inhomogeneity Corrections for Megavoltage Photon Beams , 2004 .

[19]  W. Tomé,et al.  IMRT delivery verification using a spiral phantom. , 2003, Medical physics.

[20]  M. Oldham,et al.  Evaluation of a 2D diode array for IMRT quality assurance. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[21]  Comparison of two commercial detector arrays for IMRT quality assurance. , 2007, Journal of applied clinical medical physics.

[22]  D. Albers,et al.  Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine , 2007, Physics in medicine and biology.

[23]  Stine Korreman,et al.  Dosimetric verification of RapidArc treatment delivery , 2009, Acta oncologica.

[24]  M. Kaus,et al.  Development and evaluation of an efficient approach to volumetric arc therapy planning. , 2009, Medical physics.

[25]  J. Seuntjens,et al.  A new formalism for reference dosimetry of small and nonstandard fields. , 2008, Medical physics.

[26]  James L Bedford,et al.  Evaluation of the Delta4 phantom for IMRT and VMAT verification , 2009, Physics in medicine and biology.

[27]  J. Mechalakos,et al.  IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. , 2009, Medical physics.

[28]  A. L. Angelini,et al.  Characterization of a 2D ion chamber array for the verification of radiotherapy treatments , 2005, Physics in medicine and biology.

[29]  Timothy D. Solberg,et al.  Patient specific quality assurance for the delivery of intensity modulated radiotherapy , 2003, Journal of applied clinical medical physics.

[30]  Vladimir Feygelman,et al.  Evaluation of a biplanar diode array dosimeter for quality assurance of step‐and‐shoot IMRT , 2009, Journal of applied clinical medical physics.

[31]  James L. Bedford,et al.  Verification of inverse planning and treatment delivery for segmental IMRT , 2004, Journal of applied clinical medical physics.

[32]  Cedric X. Yu,et al.  Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. , 2003, Medical physics.

[33]  Simon J Doran,et al.  The history and principles of chemical dosimetry for 3-D radiation fields: gels, polymers and plastics. , 2009, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[34]  Daniel A Low,et al.  A fourier analysis of the dose grid resolution required for accurate IMRT fluence map optimization. , 2005, Medical physics.

[35]  David A Jaffray,et al.  Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy. , 2009, Medical physics.

[36]  S Webb,et al.  Some considerations concerning volume-modulated arc therapy: a stepping stone towards a general theory. , 2009, Physics in medicine and biology.

[37]  P. Metcalfe,et al.  Verification of a rounded leaf-end MLC model used in a radiotherapy treatment planning system , 2006, Physics in medicine and biology.

[38]  J. Dempsey,et al.  TG-69: radiographic film for megavoltage beam dosimetry. , 2007, Medical physics.

[39]  T. Zhu,et al.  Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM. , 2008, Medical physics.

[40]  M. Oldham,et al.  A comprehensive evaluation of the PRESAGE/optical-CT 3D dosimetry system. , 2008, Medical physics.

[41]  Gorgen Nilsson,et al.  Characterization and clinical evaluation of a novel IMRT quality assurance system , 2009, Journal of applied clinical medical physics.