In this paper, some key techniques on development of on-line MR urine analyzing system based on AOTF (Acousto - Optics Tunable Filter) are introduced. Problems about designing the optical system including collimation of incident light and working distance (the shortest distance for separating incident light and diffracted light) are analyzed and researched. DDS (Direct Digital Synthesizer) controlled by microprocessor is used to realize the wavelength scan. The experiment results show that this MR urine analyzing system based on. AOTF has 10000 - 4000cm-1 wavelength range and O.3ms wavelength transfer rate. Compare with the conventional Fourier Transform NIP. spectrophotometer for analyzing multi-components in urine, this system features low cost, small volume and on-line measurement function. Unscrambler software (multivariate statistical software by CAMO Inc. Norway) is selected as the software for processing the data. This system can realize on line quantitative analysis of protein, urea and creatinine in urine.
[1]
R A Shaw,et al.
Quantitation of protein, creatinine, and urea in urine by near-infrared spectroscopy.
,
1996,
Clinical biochemistry.
[2]
Screening of therapeutical drugs and substances of abuse in human body fluids by near-IR laser Raman spectroscopy
,
1998,
Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Vol.20 Biomedical Engineering Towards the Year 2000 and Beyond (Cat. No.98CH36286).
[3]
Yukihiro Ozaki,et al.
Quantitative analysis of metabolites in urine using a highly precise, compact near-infrared Raman spectrometer
,
1996
.