Patient dosimetry for hybrid MRI-radiotherapy systems.

A novel geometry has been proposed for a hybrid magnetic resonance imaging (MRI)-linac system in which a 6 MV linac is mounted on the open end of a biplanar, low field (0.2 T) MRI magnet on a single gantry that is free to rotate around the patient. This geometry creates a scenario in which the magnetic field vector remains fixed with respect to the incident photon beam, but moves with respect to the patient as the gantry rotates. Other proposed geometries are characterized by a radiation source rotating about a fixed cylindrical magnet where the magnetic field vector remains fixed with respect to the patient. In this investigation we simulate the inherent dose distribution patterns within the two MRI-radiation source geometries using PENELOPE and EGSnrc Monte Carlo radiation transport codes with algorithms implemented to account for the magnetic field deflection of charged particles. Simulations are performed in phantoms and for clinically realistic situations. The novel geometry results in a net Lorentz force that remains fixed with respect to the patient (in the cranial-caudal direction) and results in a cumulative influence on dose distribution for a multiple beam treatment scenario. For a case where patient anatomy is reasonably homogeneous (brain plan), differences in dose compared to a conventional (no magnetic field) case are minimal for the novel geometry. In the case of a lung plan where the inhomogeneous patient anatomy allows for the magnetic field to have significant influence on charged particle transport, larger differences occur in a predictable manner. For a system using a fixed cylindrical geometry and higher magnetic field (1.5 T), differences from the case without a magnetic field are significantly greater.

[1]  S. Naqvi,et al.  Control of photon beam dose profiles by localized transverse magnetic fields. , 2000, Physics in medicine and biology.

[2]  J. Baró,et al.  An algorithm for Monte Carlo simulation of coupled electron-photon transport , 1997 .

[3]  J. Baró,et al.  PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter , 1995 .

[4]  A F Bielajew,et al.  The effect of strong longitudinal magnetic fields on dose deposition from electron and photon beams. , 1993, Medical physics.

[5]  B W Raaymakers,et al.  Experimental verification of magnetic field dose effects for the MRI-accelerator , 2007, Physics in medicine and biology.

[6]  A N T J Kotte,et al.  Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose deposition in a transverse magnetic field. , 2004, Physics in medicine and biology.

[7]  B W Raaymakers,et al.  Integrating a MRI scanner with a 6 MV radiotherapy accelerator: impact of the surface orientation on the entrance and exit dose due to the transverse magnetic field , 2007, Physics in medicine and biology.

[8]  D W O Rogers,et al.  Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM code. , 2002, Medical physics.

[9]  B W Raaymakers,et al.  Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electrons , 2005, Physics in medicine and biology.

[10]  M. J. Berger ESTAR, PSTAR, and ASTAR: Computer programs for calculating stopping-power and range tables for electrons, protons, and helium ions , 1992 .

[11]  X. Li,et al.  Conformal photon-beam therapy with transverse magnetic fields: a Monte Carlo study. , 2001 .

[12]  Sasa Mutic,et al.  A Device for Realtime 3D Image-Guided IMRT , 2005 .

[13]  B. G. Fallone,et al.  Sci-Thur PM Therapy-03: Automated Monte Carlo Simulation and Validation of an External Beam Radiotherapy Treatment Plan , 2006 .

[14]  D. Jette Magnetic fields with photon beams: Monte Carlo calculations for a model magnetic field. , 2000, Medical physics.

[15]  Tomas Kron,et al.  Magnetic resonance imaging for adaptive cobalt tomotherapy: A proposal , 2006, Journal of medical physics.

[16]  I. Kawrakow Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version. , 2000, Medical physics.

[17]  J J W Lagendijk,et al.  Dose optimization for the MRI-accelerator: IMRT in the presence of a magnetic field , 2007, Physics in medicine and biology.

[18]  Satyapal Rathee,et al.  TU‐C‐M100F‐01: Development of a Linac‐MRI System for Real‐Time ART , 2007 .

[19]  Jean M Moran,et al.  Magnetic confinement of electron and photon radiotherapy dose: a Monte Carlo simulation with a nonuniform longitudinal magnetic field. , 2005, Medical physics.

[20]  I. Kawrakow Accurate condensed history Monte Carlo simulation of electron transport. II. Application to ion chamber response simulations. , 2000, Medical physics.