Evaluation of a Novel Noninvasive Respiration Monitor Providing Continuous Measurement of Minute Ventilation in Ambulatory Subjects in a Variety of Clinical Scenarios

BACKGROUND:Currently there is no technology that noninvasively measures the adequacy of ventilation in nonintubated patients. A novel, noninvasive Respiratory Volume Monitor (RVM) has been developed to continuously measure and display minute ventilation (MV), tidal volume (TV), and respiratory rate (RR) in a variety of clinical settings. We demonstrate the RVM’s accuracy and precision as compared with a standard spirometer under a variety of clinically relevant breathing patterns in nonintubated subjects. METHODS:Thirty-one voluntary subjects completed the primary study. MV, TV, and RR measurements were collected from the RVM and spirometer simultaneously for each participant on day 1 and day 2 and analyzed to determine accuracy, precision, and bias for normal, fast, slow, irregular, and closed-glottis breathing. RESULTS:Data demonstrated that RVM and spirometer measurements of MV and TV are equivalent in a wide range of ambulatory subjects with an average error <10% (95% confidence interval for accuracy <16%, precision <12%, and bias <11%). Repeated measures analysis of variance found no significant difference between spirometry and RVM individual measurements of MV, TV, and RR (P > 0.7), whereas a paired-difference equivalent test demonstrated, with 99% power, that both MV and TV measurements from the 2 devices are equivalent within ±15%. CONCLUSIONS:This study demonstrates RVM’s clinically relevant accuracy and precision in measuring MV, TV, and RR over a 24-hour period and during various breathing patterns.

[1]  J. Sanchís,et al.  Non-invasive ventilation , 2002, Thorax.

[2]  S. Nava,et al.  Time of non-invasive ventilation , 2006, Intensive Care Medicine.

[3]  M. Moskowitz,et al.  Discharge decision-making in a medical intensive care unit. Identifying patients at high risk of unexpected death or unit readmission. , 1988, The American journal of medicine.

[4]  Luc J. Teppema,et al.  Anesthetics and control of breathing , 2011, Respiratory Physiology & Neurobiology.

[5]  C. Pitoyo,et al.  Non-invasive ventilation in acute respiratory failure. , 2014, Acta medica Indonesiana.

[6]  D. Altman,et al.  STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT , 1986, The Lancet.

[7]  I. Mitchell,et al.  Missed opportunities? An observational study of vital sign measurements. , 2008, Critical care and resuscitation : journal of the Australasian Academy of Critical Care Medicine.

[8]  S. Fouzas,et al.  Knowledge on Pulse Oximetry Among Pediatric Health Care Professionals: A Multicenter Survey , 2010, Pediatrics.

[9]  Anthony Martínez,et al.  Minute ventilation recovery time: a predictor of extubation outcome. , 2003, Chest.

[10]  Tapabrata Maiti,et al.  Analysis of Longitudinal Data (2nd ed.) (Book) , 2004 .

[11]  Evangelos Manolis,et al.  Patient severity as an indicator of nursing workload in the intensive care unit. , 2007, Nursing in critical care.

[12]  L. Lynn,et al.  Patterns of unexpected in-hospital deaths: a root cause analysis , 2011, Patient safety in surgery.

[13]  S. Mcgrath,et al.  A Review of Current and Emerging Approaches to Address Failure-to-Rescue , 2011, Anesthesiology.

[14]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[15]  K. Tremper,et al.  Pulse oximetry. , 1989, Anesthesiology.

[16]  P. Bacchetti,et al.  Sample size calculations in clinical research. , 2002, Anesthesiology.

[17]  C. Manthous Rapid-response teams. , 2011, The New England journal of medicine.

[18]  F. Bongard,et al.  Accuracy of capnography in nonintubated surgical patients. , 1992, Chest.

[19]  Y. Kaneko Clinical perspectives on capnography during sedation and general anesthesia in dentistry. , 1995, Anesthesia progress.

[20]  L. O'Brien-Pallas,et al.  Nursing workload associated with adverse events in the postanesthesia care unit. , 1999, Anesthesiology.

[21]  J H Auchincloss,et al.  Changes in tidal volume, frequency, and ventilation induced by their measurement. , 1972, Journal of applied physiology.

[22]  S. Nava,et al.  Non-invasive ventilation in elderly patients with acute hypercapnic respiratory failure: a randomised controlled trial. , 2011, Age and ageing.

[23]  J. Orr,et al.  Anatomic dead space cannot be predicted by body weight. , 2008, Respiratory care.

[24]  P. Diggle Analysis of Longitudinal Data , 1995 .

[25]  M. Good Capnography: Uses, Interpretation, and Pitfalls , 1990 .

[26]  N. MacIntyre,et al.  Respiratory controversies in the critical care setting. Conference summary. , 2007, Respiratory care.

[27]  G. Blike,et al.  Impact of Pulse Oximetry Surveillance on Rescue Events and Intensive Care Unit Transfers: A Before-and-After Concurrence Study , 2010, Anesthesiology.

[28]  E. Radford,et al.  Ventilation standards for use in artificial respiration. , 1955, Journal of applied physiology.

[29]  N. MacIntyre,et al.  Respiratory Controversies in the Critical Care Setting , 2007 .

[30]  M. Elliott Non-invasive ventilation: established and expanding roles. , 2011, Clinical medicine.