Wireless programmable electrochemical drug delivery micropump with fully integrated electrochemical dosing sensors

We present a fully integrated implantable electrolysis-based micropump with incorporated EI dosing sensors. Wireless powering and data telemetry (through amplitude and frequency modulation) were utilized to achieve variable flow control and a bi-directional data link with the sensors. Wireless infusion rate control (0.14–1.04 μL/min) and dose sensing (bolus resolution of 0.55–2 μL) were each calibrated separately with the final circuit architecture and then simultaneous wireless flow control and dose sensing were demonstrated. Recombination detection using the dosing system, as well as, effects of coil separation distance and misalignment in wireless power and data transfer were studied. A custom-made normally closed spring-loaded ball check valve was designed and incorporated at the reservoir outlet to prevent backflow of fluids as a result of the reverse pressure gradient caused by recombination of electrolysis gases. Successful delivery, infusion rate control, and dose sensing were achieved in simulated brain tissue.

[1]  E. Meng,et al.  High-Efficiency MEMS Electrochemical Actuators and Electrochemical Impedance Spectroscopy Characterization , 2012, Journal of Microelectromechanical Systems.

[2]  Fei Zhang,et al.  In vitro and in vivo studies on wireless powering of medical sensors and implantable devices , 2009, 2009 IEEE/NIH Life Science Systems and Applications Workshop.

[3]  J. Spalding,et al.  Toward a chronotherapy of neoplasia: Tolerance of treatment depends upon host rhythms , 1973, Experientia.

[4]  Rui Hu,et al.  Approaches and Challenges of Engineering Implantable Microelectromechanical Systems (MEMS) Drug Delivery Systems for in Vitro and in Vivo Applications , 2012, Micromachines.

[5]  Patricia Wood,et al.  Circadian Timing in Cancer Treatment: The Biological Foundation for an Integrative Approach , 2003, Integrative cancer therapies.

[6]  N. Kaisare,et al.  An analysis of drifts and nonlinearities in electrochemical impedance spectra , 2011 .

[7]  J. Urquhart,et al.  Controlled drug delivery: therapeutic and pharmacological aspects , 2000 .

[8]  J. Fiering,et al.  Local drug delivery with a self-contained, programmable, microfluidic system , 2009, Biomedical microdevices.

[9]  Hrushesky Wj Cancer chronotherapy: a drug delivery challenge. , 1990 .

[10]  Ellis Meng,et al.  A Wireless Implantable Micropump for Chronic Drug Infusion Against Cancer. , 2016, Sensors and actuators. A, Physical.

[11]  Robert Puers,et al.  Omnidirectional Inductive Powering for Biomedical Implants , 2008 .

[12]  Ellis Meng,et al.  On-demand wireless infusion rate control in an implantable micropump for patient-tailored treatment of chronic conditions , 2014, 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[13]  E. Meng,et al.  A wireless implantable drug infusion system with integrated dosing sensors , 2015, 2015 Transducers - 2015 18th International Conference on Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS).

[14]  Ellis Meng,et al.  Design, fabrication, and characterization of an electrochemically-based dose tracking system for closed-loop drug delivery , 2012, 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[15]  E. Meng,et al.  Rapid non-lithography based fabrication process and characterization of Parylene C bellows for applications in MEMS electrochemical actuators , 2011, 2011 16th International Solid-State Sensors, Actuators and Microsystems Conference.

[16]  B. Youan,et al.  Chronopharmaceutics: gimmick or clinically relevant approach to drug delivery? , 2004, Journal of controlled release : official journal of the Controlled Release Society.

[17]  Franco Maloberti,et al.  Servo-controlled remote powering and low-power data communication of implantable bio-systems for freely moving animals , 2014, 2014 IEEE Biomedical Circuits and Systems Conference (BioCAS) Proceedings.

[18]  M. Ferrari,et al.  Drug Delivery Systems , 2006 .

[19]  G Hartsgrove,et al.  Simulated biological materials for electromagnetic radiation absorption studies. , 1987, Bioelectromagnetics.

[20]  Ellis Meng,et al.  A MEMS electrochemical bellows actuator for fluid metering applications , 2013, Biomedical microdevices.

[21]  E. Meng,et al.  MEMS-enabled implantable drug infusion pumps for laboratory animal research, preclinical, and clinical applications. , 2012, Advanced drug delivery reviews.

[22]  B. Bruguerolle,et al.  Rhythmic pattern in pain and their chronotherapy. , 2007, Advanced drug delivery reviews.

[23]  Robert Puers,et al.  Inductive Powering: Basic Theory and Application to Biomedical Systems , 2009 .

[24]  Heidi Gensler A Wireless Implantable MEMS Micropump System for Site-specific Anti-cancer Drug Delivery , 2013 .

[25]  J. Urquhart,et al.  Rate-controlled delivery systems in drug and hormone research. , 1984, Annual review of pharmacology and toxicology.

[26]  Austin J. Brockmeier,et al.  An Association Framework to Analyze Dependence Structure in Time Series , 2012, 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[27]  Mary E. Klingensmith,et al.  The Washington Manual of Surgery , 2008 .