A Fuzzy Logic Approach for a Wearable Cardiovascular and Aortic Monitoring System

A new methodology for fault detection on wearable medical devices is proposed. The basic strategy relies on correctly classifying the captured physiological signals, in order to identify whether the actual cause is a wearer health abnormality or a system functional flaw. Data fusion techniques, namely fuzzy logic, are employed to process the physiological signals, like the electrocardiogram (ECG) and blood pressure (BP), to increase the trust levels of the captured data after rejecting or correcting distorted vital signals from each sensor, and to provide additional information on the patient’s condition by classifying the set of signals into normal or abnormal condition (e.g. arrhythmia, chest angina, and stroke). Once an abnormal situation is detected in one or several sensors the monitoring system runs a set of tests in a fast and energy efficient way to check if the wearer shows a degradation of his health condition or the system is reporting erroneous values.

[1]  Brian L. Wardle,et al.  Implantable Flexible Pressure Measurement System Based on Inductive Coupling , 2015, IEEE Transactions on Biomedical Engineering.

[2]  Jeffrey M. Hausdorff,et al.  Physionet: Components of a New Research Resource for Complex Physiologic Signals". Circu-lation Vol , 2000 .

[3]  Georges Delhomme,et al.  Wearable Electronic Systems: Applications to Medical Diagnostics/Monitoring , 2011 .

[4]  Sungmee Park,et al.  Enhancing the quality of life through wearable technology , 2003, IEEE Engineering in Medicine and Biology Magazine.

[5]  Rita Paradiso,et al.  A wearable health care system based on knitted integrated sensors , 2005, IEEE Transactions on Information Technology in Biomedicine.

[6]  R. Paradiso,et al.  Wearable health care system for vital signs monitoring , 2003, 4th International IEEE EMBS Special Topic Conference on Information Technology Applications in Biomedicine, 2003..

[7]  Charles Sodini,et al.  A long-term wearable electrocardiogram measurement system , 2013, 2013 IEEE International Conference on Body Sensor Networks.

[8]  Naif Alajlan,et al.  Implementation of a Mobile Health System for Monitoring ECG signals , 2014 .

[9]  N. Kannathal,et al.  Data Fusion of Multimodal Cardiovascular Signals , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[10]  Willis J. Tompkins,et al.  A Real-Time QRS Detection Algorithm , 1985, IEEE Transactions on Biomedical Engineering.

[11]  Isabel G. Trindade,et al.  Characterization of the electrode-skin impedance of textile electrodes , 2014, Design of Circuits and Integrated Systems.