Myocardial tissue Doppler velocity imaging in children: comparative study between two ultrasound systems.

BACKGROUND The investigators compared pulsed-wave tissue Doppler (PWTD) and color tissue Doppler (CTD) ventricular tissue velocities obtained using Vivid 7 (GE) and iE33 (Philips) ultrasound systems within a 30-minute time frame and examined interobserver and intraobserver variability. METHODS Longitudinal PWTD and CTD images were prospectively acquired in 49 children from the base of the right and left ventricular free wall and septum and stored for offline analysis. RESULTS Intraobserver and interobserver coefficients of variation for PWTD velocities ranged from 4% to 6% and 5% to 10% for GE and from 1% to 10% and 2% to 14% for Philips, respectively. Intraobserver and interobserver coefficients of variation for CTD velocities ranged from 4% to 15% and 6% to 24% for GE and from 6% to 19% and 7% to 25% for Philips, respectively. Mean biases between the two systems for PWTD velocities, expressed as percentage differences, were around 0%, with the narrowest limits of agreement for systolic velocities at the septal annulus and the widest limits of agreement for early diastolic velocities at the lateral tricuspid valve annulus, respectively. Peak CTD-derived velocities were significantly and consistently lower for Philips compared with GE. CONCLUSIONS Agreement between GE-derived and Philips-derived PWTD velocities was reasonable, whereas CTD-derived velocities differed substantially.

[1]  A. Manouras,et al.  Comparison between colour-coded and spectral tissue Doppler measurements of systolic and diastolic myocardial velocities: effect of temporal filtering and offline gain setting. , 2009, European journal of echocardiography : the journal of the Working Group on Echocardiography of the European Society of Cardiology.

[2]  K. Harada,et al.  A comparison of tissue doppler imaging and velocities of transmitral flow in children with elevated left ventricular preload , 2001, Cardiology in the Young.

[3]  Michael Vogel,et al.  Noninvasive Assessment of Left Ventricular Force-Frequency Relationships Using Tissue Doppler–Derived Isovolumic Acceleration: Validation in an Animal Model , 2003, Circulation.

[4]  Per Ask,et al.  Accuracy of spectral Doppler flow and tissue velocity measurements in ultrasound systems. , 2004, Ultrasound in medicine & biology.

[5]  G. Doucende,et al.  Tissue Doppler imaging reproducibility during exercise. , 2008, International journal of sports medicine.

[6]  B Wranne,et al.  A Comparison of Regional Myocardial Velocity Information Derived by Pulsed and Color Doppler Techniques: An In Vitro and In Vivo Study , 2000, Echocardiography.

[7]  C. McMahon,et al.  Impact of cardiac growth on Doppler tissue imaging velocities: a study in healthy children. , 2004, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[8]  M. Friedberg,et al.  Feasibility of a new tissue Doppler based method for comprehensive evaluation of left-ventricular intra-ventricular mechanical dyssynchrony in children with dilated cardiomyopathy. , 2008, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[9]  A. Fraser,et al.  Reproducibility of pulsed wave tissue Doppler echocardiography. , 1999, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[10]  A. Manouras,et al.  Are measurements of systolic myocardial velocities and displacement with colour and spectral Tissue Doppler compatible? , 2009, Cardiovascular ultrasound.

[11]  J. Cleland,et al.  Longitudinal ventricular function: normal values of atrioventricular annular and myocardial velocities measured with quantitative two-dimensional color Doppler tissue imaging. , 2003, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[12]  J. Kingdom,et al.  Assessment of fetal atrioventricular time intervals by tissue Doppler and pulse Doppler echocardiography: normal values and correlation with fetal electrocardiography , 2006, Heart.

[13]  A. Cohen-Solal,et al.  Colour tissue Doppler underestimates myocardial velocity as compared to spectral tissue Doppler: poor reliability between both methods. , 2007, European journal of echocardiography : the journal of the Working Group on Echocardiography of the European Society of Cardiology.

[14]  M. Friedberg,et al.  Relations between systolic and diastolic function in children with dilated and hypertrophic cardiomyopathy as assessed by tissue Doppler imaging. , 2009, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[15]  D. Roberson,et al.  Tissue Doppler imaging measurement of left ventricular systolic function in children: mitral annular displacement index is superior to peak velocity. , 2009, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[16]  M. Belohlavek,et al.  Accuracy, reproducibility, and comparability of Doppler tissue imaging by two high-end ultrasound systems. , 2006, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[17]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[18]  F. T. ten Cate,et al.  Assessment of mitral annular velocities by speckle tracking echocardiography versus tissue Doppler imaging: validation, feasibility, and reproducibility. , 2009, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[19]  Zhimin Chen,et al.  Annular and septal Doppler tissue imaging in children: normal z-score tables and effects of age, heart rate, and body surface area. , 2007, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[20]  K Isaaz,et al.  Quantitation of the motion of the cardiac base in normal subjects by Doppler echocardiography. , 1993, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.

[21]  Sample volume positioning in colour-coded Doppler myocardial imaging: effect of systolic and diastolic tracking. , 2009, Ultrasound in medicine & biology.

[22]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[23]  W. Zoghbi,et al.  Color tissue Doppler myocardial velocities consistently underestimate spectral tissue Doppler velocities: impact on calculation peak transmitral pulsed Doppler velocity/early diastolic tissue Doppler velocity (E/Ea). , 2006, Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography.