Use of TrueBeam developer mode for imaging QA

The purpose of this study was to automate regular Imaging QA procedures to become more efficient and accurate. Daily and monthly imaging QA for SRS and SBRT protocols were fully automated on a Varian linac. A three‐step paradigm where the data are automatically acquired, processed, and analyzed was defined. XML scripts were written and used in developer mode in a TrueBeam linac to automatically acquire data. MATLAB R013B was used to develop an interface that could allow the data to be processed and analyzed. Hardware was developed that allowed the localization of several phantoms simultaneously on the couch. 14 KV CBCTs from the Emma phantom were obtained using a TrueBeam onboard imager as example of data acquisition and analysis. The images were acquired during two months. Artifacts were artificially introduced in the images during the reconstruction process using iTool reconstructor. Support vector machine algorithms to automatically identify each artifact were written using the Machine Learning MATLAB R2011 Toolbox. A daily imaging QA test could be performed by an experienced medical physicist in 14.3±2.4 min. The same test, if automated using our paradigm, could be performed in 4.2±0.7 min. In the same manner, a monthly imaging QA could be performed by a physicist in 70.7±8.0 min and, if fully automated, in 21.8±0.6 min. Additionally, quantitative data analysis could be automatically performed by Machine Learning Algorithms that could remove the subjectivity of data interpretation in the QA process. For instance, support vector machine algorithms could correctly identify beam hardening, rings and scatter artifacts. Traditional metrics, as well as metrics that describe texture, are needed for the classification. Modern linear accelerators are equipped with advanced 2D and 3D imaging capabilities that are used for patient alignment, substantially improving IGRT treatment accuracy. However, this extra complexity exponentially increases the number of QA tests needed. Using the new paradigm described above, not only the bare minimum — but also best practice — QA programs could be implemented with the same manpower. PACS number: 87, 87.10.‐e

[1]  Iori Sumida,et al.  Evaluation of imaging performance of megavoltage cone-beam CT over an extended period , 2014, Journal of radiation research.

[2]  Fang-Fang Yin,et al.  Task Group 142 report: quality assurance of medical accelerators. , 2009, Medical physics.

[3]  G. Tourassi Journey toward computer-aided diagnosis: role of image texture analysis. , 1999, Radiology.

[4]  Jan-Jakob Sonke,et al.  Quality assurance for image-guided radiation therapy utilizing CT-based technologies: a report of the AAPM TG-179. , 2012, Medical physics.

[5]  Moyed Miften,et al.  Commissioning and clinical implementation of a mega-voltage cone beam CT system for treatment localization. , 2007, Medical physics.

[6]  D Lazos,et al.  SU-E-J-21: An Intercomparison of Imaging Performance of Two Linac-Mounted Imaging Systems Used in Radiation Therapy: TrueBeam and Trilogy. , 2012, Medical physics.

[7]  Sanjiv S Samant,et al.  Use of a line-pair resolution phantom for comprehensive quality assurance of electronic portal imaging devices based on fundamental imaging metrics. , 2009, Medical physics.

[8]  Minsong Cao,et al.  A quality assurance phantom for electronic portal imaging devices , 2011, Journal of applied clinical medical physics.

[9]  Jonas D. Fontenot,et al.  AAPM Medical Physics Practice Guideline 2.a: Commissioning and quality assurance of X‐ray‐based image‐guided radiotherapy systems , 2014, Journal of applied clinical medical physics.

[10]  K. Winston,et al.  A system for stereotactic radiosurgery with a linear accelerator. , 1986, International journal of radiation oncology, biology, physics.

[11]  R L Morin,et al.  A practical method to measure the MTF of CT scanners. , 1982, Medical physics.

[12]  Song Gao,et al.  SU‐C‐213CD‐04: Evaluation of IsoCal Imaging Isocenter Calibration System for Varian OBI Machines , 2012 .

[13]  Leen-Kiat Soh,et al.  Texture analysis of SAR sea ice imagery using gray level co-occurrence matrices , 1999, IEEE Trans. Geosci. Remote. Sens..

[14]  David A Clausi An analysis of co-occurrence texture statistics as a function of grey level quantization , 2002 .