Cerebral near infrared spectroscopy: emitter-detector separation must be increased.

We have compared the effect of increasing optode separation (range 0.7-5.5 cm) on the sensitivity of near infrared spectroscopy (NIRS) to discrete reductions in scalp and cerebral oxygenation in 10 healthy men (mean age 32, range 26-39 yr) using multichannel NIRS. During cerebral oligaemia (a mean reduction in middle cerebral artery flow velocity of 47%) induced by a mean reduction in end-tidal PCO2 of 2.4 kPa, the decrease in oxyhaemoglobin detected by NIRS became significantly greater with increasing optode separation (P < 0.0001). In response to scalp hyperaemia induced by inflation and release of a pneumatic scalp tourniquet, increases in oxyhaemoglobin became significantly smaller with increasing optode separation (P < 0.0002). These results are consistent with theoretical models of the behaviour of NIR light in the adult head and support the concept of using multi-detector NIRS to separate intra- and extracranial NIR signal changes. However, the emitter-detector separation used by currently available cerebral oximeters is not large enough to provide optimal spatial resolution.

[1]  R. Nelson,et al.  Changes in tissue oxyhaemoglobin concentration measured using multichannel near infrared spectroscopy during internal carotid angiography , 1997, Journal of neurology, neurosurgery, and psychiatry.

[2]  M. Cope,et al.  Measurement of Cerebral Blood Flow in Dogs with Near Infrared Spectroscopy in the Reflectance Mode Is Invalid , 1997, Journal of cerebral blood flow and metabolism : official journal of the International Society of Cerebral Blood Flow and Metabolism.

[3]  W. Trubel,et al.  The contribution of extracranial blood oxygenation on near‐infrared spectroscopy during carotid thrombendarterectomy , 1997, Anaesthesia.

[4]  G. Litscher,et al.  Transcranial cerebral oximetry , 1997 .

[5]  P. Smielewski,et al.  Use of near infrared spectroscopy for the clinical monitoring of adult brain. , 1996, Journal of biomedical optics.

[6]  P. Reilly,et al.  Cerebral oxygenation monitoring by near-infrared spectroscopy is not clinically useful in patients with severe closed-head injury: a comparison with jugular venous bulb oximetry. , 1996, Critical care medicine.

[7]  G. Litscher,et al.  Cerebral Oximetry in Dead Subjects , 1996, Journal of neurosurgical anesthesiology.

[8]  J D Pickard,et al.  Near-infrared spectroscopy use in patients with head injury. , 1995, Journal of neurosurgery.

[9]  J D Pickard,et al.  Can cerebrovascular reactivity be measured with near-infrared spectroscopy? , 1995, Stroke.

[10]  R. Nelson,et al.  Extracerebral absorption of near infrared light influences the detection of increased cerebral oxygenation monitored by near infrared spectroscopy. , 1995, Journal of neurology, neurosurgery, and psychiatry.

[11]  N. Kane,et al.  Near-infrared spectroscopy in adults: effects of extracranial ischaemia and intracranial hypoxia on estimation of cerebral oxygenation. , 1994, British journal of anaesthesia.

[12]  N. Kane,et al.  Possible limitations of cerebral oximetric monitoring in adults , 1994, Anaesthesia.

[13]  D. Delpy,et al.  Use of the water absorption spectrum to quantify tissue chromophore concentration changes in near-infrared spectroscopy. , 1994, Physics in medicine and biology.

[14]  D. Royston,et al.  A comparison of two systems for assessing cerebral venous oxyhaemoglobin saturation during cardiopulmonary bypass in humans , 1993, Anaesthesia.

[15]  S. M. Bailey,et al.  Near infrared spectroscopy in adults , 1993, Anaesthesia.

[16]  S. Arridge,et al.  Experimentally measured optical pathlengths for the adult head, calf and forearm and the head of the newborn infant as a function of inter optode spacing. , 1992, Advances in experimental medicine and biology.

[17]  Britton Chance,et al.  Experimental study of migration depth for the photons measured at sample surface , 1991, Photonics West - Lasers and Applications in Science and Engineering.

[18]  P. W. Mccormick,et al.  Regional cerebrovascular oxygen saturation measured by optical spectroscopy in humans. , 1991, Stroke.

[19]  L. E. Ostrander,et al.  Assessment Of Layered Properties In Biological Tissue From Dual-wavelength Optical Reflectance Measurements , 1990, [1990] Proceedings of the Twelfth Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[20]  Joseph R. Lakowicz,et al.  Frequency-domain measurements of photon migration in tissues , 1990 .

[21]  S R Arridge,et al.  The effect of optode positioning on optical pathlength in near infrared spectroscopy of brain. , 1990, Advances in experimental medicine and biology.

[22]  R Nossal,et al.  Influence of path length on remote optical sensing of properties of biological tissue. , 1989, Applied optics.

[23]  D. Delpy,et al.  A CCD spectrophotometer to quantitate the concentration of chromophores in living tissue utilising the absorption peak of water at 975 nm. , 1989, Advances in experimental medicine and biology.

[24]  S. Arridge,et al.  Estimation of optical pathlength through tissue from direct time of flight measurement , 1988 .

[25]  Ralph Nossal,et al.  Photon migration in layered media. , 1988, Applied optics.

[26]  G. Weiss,et al.  Model for photon migration in turbid biological media. , 1987, Journal of the Optical Society of America. A, Optics and image science.

[27]  J. Wood Cerebral blood flow : physiologic and clinical aspects , 1987 .

[28]  Christopher P. Crotty,et al.  Biochemistry and Physiology of the Skin , 1984 .

[29]  L. O. Svaasand,et al.  OPTICAL PROPERTIES OF HUMAN BRAIN , 1983, Photochemistry and photobiology.

[30]  F. Jöbsis Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. , 1977, Science.

[31]  S. Kety,et al.  THE EFFECTS OF ALTERED ARTERIAL TENSIONS OF CARBON DIOXIDE AND OXYGEN ON CEREBRAL BLOOD FLOW AND CEREBRAL OXYGEN CONSUMPTION OF NORMAL YOUNG MEN. , 1948, The Journal of clinical investigation.