Artifacts in pulse transit time measurements using standard patient monitoring equipment

Objective Pulse transit time (PTT) refers to the time it takes a pulse wave to travel between two arterial sites. PTT can be estimated, amongst others, using the electrocardiogram (ECG) and photoplethysmogram (PPG). Because we observed a sawtooth artifact in the PTT while using standard patient monitoring equipment for ECG and PPG, we explored the reasons for this artifact. Methods PPG and ECG were simulated at a heartrate of both 100 and 160 beats per minute while using a Masimo PPG post-processing module and a Philips patient monitor setup at the neonatal intensive care unit. Two different post-processing modules were used. PTT was defined as the difference between the R-peak in the ECG and the point of 50% increase in the PPG. Results A sawtooth artifact was seen in all simulations. Both length (59.2 to 72.4 s) and amplitude (30.8 to 36.0 ms) of the sawtooth were dependent on the post-processing module used. Furthermore, the absolute PTT value differed up to 250 ms depending on post-processing module and heart rate. The sawtooth occurred because the PPG wave continuously showed a minimal prolongation during the length of the sawtooth, followed by a sudden shortening. Both artifacts were generated in the post-processing module containing Masimo algorithms. Conclusion Post-processing of the PPG signal in the Masimo module of the Philips patient monitor introduces a sawtooth in PPG and derived PTT. This sawtooth, together with a large module-dependent absolute difference in PTT, renders the thus-derived PTT insufficient for clinical purposes.

[1]  M. Nitzan,et al.  Photoplethysmographic assessment of pulse transit time correlates with echocardiographic measurement of stroke volume in preterm infants with patent ductus arteriosus , 2018, Journal of Perinatology.

[2]  M. Nitzan,et al.  Increase in pulse transit time to the foot after epidural anaesthesia treatment , 2000, Medical and Biological Engineering and Computing.

[3]  S. J. Wilson,et al.  Clinical applications of pulse transit time in paediatric critical care , 2009, Journal of medical engineering & technology.

[4]  Gordon R. Williams,et al.  Variability in time delay between two models of pulse oximeters for deriving the photoplethysmographic signals , 2005, Physiological measurement.

[5]  M. Elgendi On the Analysis of Fingertip Photoplethysmogram Signals , 2012, Current cardiology reviews.

[6]  B. Chakrabarti,et al.  Pulse transit time changes in subjects exhibiting sleep disordered breathing. , 2017, Respiratory medicine.

[7]  B. Galland,et al.  The use of pulse transit time in pediatric sleep studies: A systematic review. , 2018, Sleep medicine reviews.

[8]  P. Kyriacou,et al.  Evaluation of the Linear Relationship Between Pulse Arrival Time and Blood Pressure in ICU Patients: Potential and Limitations , 2018, Front. Physiol..

[9]  M. Karm,et al.  Pulse Transit Time as a Predictor of the Efficacy of a Celiac Plexus Block in Patients With Chronic Intractable Abdominal Pain , 2016, The Clinical journal of pain.

[10]  Anton Vonk-Noordegraaf,et al.  Stroke volume response during exercise measured by acetylene uptake and MRI , 2007, Physiological measurement.

[11]  R. Stolker,et al.  Comparison of bilateral pulse arrival time before and after induced vasodilation by axillary block , 2012, Physiological measurement.

[12]  Margot J Davey,et al.  Pulse transit time as a surrogate measure of changes in systolic arterial pressure in children during sleep , 2014, Journal of sleep research.

[13]  J Y A Foo,et al.  Age-related factors that confound peripheral pulse timing characteristics in Caucasian children , 2005, Journal of Human Hypertension.

[14]  John Allen Photoplethysmography and its application in clinical physiological measurement , 2007, Physiological measurement.

[15]  Clinical Application of Pulse Transit Time and Correlation With Intrapartum Fetal Heart Rate Monitoring: A Preliminary Study of 18 Full-Term Infants , 2008, Reproductive Sciences.

[16]  F. Wallois,et al.  Noninvasive Technique for the Diagnosis of Patent Ductus Arteriosus in Premature Infants by Analyzing Pulse Wave Phases on Photoplethysmography Signals Measured in the Right Hand and the Left Foot , 2014, PloS one.