Rapid prototyping compliant arterial phantoms for in-vitro studies and device testing

BackgroundCompliant vascular phantoms are desirable for in-vitro patient-specific experiments and device testing. TangoPlus FullCure 930® is a commercially available rubber-like material that can be used for PolyJet rapid prototyping. This work aims to gather preliminary data on the distensibility of this material, in order to assess the feasibility of its use in the context of experimental cardiovascular modelling.MethodsThe descending aorta anatomy of a volunteer was modelled in 3D from cardiovascular magnetic resonance (CMR) images and rapid prototyped using TangoPlus. The model was printed with a range of increasing wall thicknesses (0.6, 0.7, 0.8, 1.0 and 1.5 mm), keeping the lumen of the vessel constant. Models were also printed in both vertical and horizontal orientations, thus resulting in a total of ten specimens. Compliance tests were performed by monitoring pressure variations while gradually increasing and decreasing internal volume. Knowledge of distensibility was thus derived and then implemented with CMR data to test two applications. Firstly, a patient-specific compliant model of hypoplastic aorta suitable for connection in a mock circulatory loop for in-vitro tests was manufactured. Secondly, the right ventricular outflow tract (RVOT) of a patient necessitating pulmonary valve replacement was printed in order to physically test device insertion and assess patient’s suitability for percutaneous pulmonary valve intervention.ResultsThe distensibility of the material was identified in a range from 6.5 × 10-3 mmHg-1 for the 0.6 mm case, to 3.0 × 10-3 mmHg-1 for the 1.5 mm case. The models printed in the vertical orientation were always more compliant than their horizontal counterpart. Rapid prototyping of a compliant hypoplastic aorta and of a RVOT anatomical model were both feasible. Device insertion in the RVOT model was successful.ConclusionValues of distensibility, compared with literature data, show that TangoPlus is suitable for manufacturing arterial phantoms, with the added benefit of being compatible with PolyJet printing, thus guaranteeing representative anatomical finishing, and quick and inexpensive fabrication. The appealing possibility of printing models of non-uniform wall thickness, resembling more closely certain physiological scenarios, can also be explored. However, this material appears to be too stiff for modelling the more compliant systemic venous system.

[1]  D. Bergel Cardiovascular fluid dynamics , 1972 .

[2]  H Reul,et al.  CAD-design, stress analysis and in vitro evaluation of three leaflet blood-pump valves. , 1992, Journal of biomedical engineering.

[3]  A. Yoganathan,et al.  Toward designing the optimal total cavopulmonary connection: an in vitro study. , 1999, The Annals of thoracic surgery.

[4]  R. Reneman,et al.  Static versus Dynamic Distensibility of the Carotid Artery in Humans , 2000, Journal of Vascular Research.

[5]  Hristo N Nikolov,et al.  A thin-walled carotid vessel phantom for Doppler ultrasound flow studies. , 2004, Ultrasound in medicine & biology.

[6]  J. Michael Textbook of Medical Physiology , 2005 .

[7]  Mark Fogel,et al.  In vitro flow analysis of a patient-specific intraatrial total cavopulmonary connection. , 2005, The Annals of thoracic surgery.

[8]  Jürgen Hennig,et al.  Rapid vessel prototyping: vascular modeling using 3t magnetic resonance angiography and rapid prototyping technology , 2005, Magnetic Resonance Materials in Physics, Biology and Medicine.

[9]  F. Migliavacca,et al.  Percutaneous pulmonary valve implantation based on rapid prototyping of right ventricular outflow tract and pulmonary trunk from MR data. , 2007, Radiology.

[10]  S. Sherwin,et al.  Pulse wave propagation in a model human arterial network: assessment of 1-D numerical simulations against in vitro measurements. , 2007, Journal of biomechanics.

[11]  Nico Westerhof,et al.  Noninvasively assessed pulmonary artery stiffness predicts mortality in pulmonary arterial hypertension. , 2007, Chest.

[12]  Can an intermittent pneumatic compression system monitor venous filling in the leg? , 2008, Journal of medical engineering & technology.

[13]  L. Boussel,et al.  In vitro non-rigid life-size model of aortic arch aneurysm for endovascular prosthesis assessment. , 2008, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[14]  R. Wicker,et al.  Patient-Specific Compliant Vessel Manufacturing Using Dip-Spin Coating of Rapid Prototyped Molds , 2008 .

[15]  J. Hennig,et al.  3D MR flow analysis in realistic rapid‐prototyping model systems of the thoracic aorta: Comparison with in vivo data and computational fluid dynamics in identical vessel geometries , 2008, Magnetic resonance in medicine.

[16]  T. Z. Teisseyre,et al.  Comparison of particle image velocimetry and phase contrast MRI in a patient-specific extracardiac total cavopulmonary connection. , 2008, Journal of biomechanical engineering.

[17]  Silvia Schievano,et al.  Experimental setup to evaluate the performance of percutaneous pulmonary valved stent in different outflow tract morphologies. , 2009, Artificial organs.

[18]  D. Silva,et al.  Dimensional error of selective laser sintering, three-dimensional printing and PolyJet models in the reproduction of mandibular anatomy. , 2009, Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery.

[19]  M J Pearcy,et al.  A naturally shaped silicone ventricle evaluated in a mock circulation loop: A preliminary study , 2009, Journal of medical engineering & technology.

[20]  Shigeo Sora,et al.  SIMULATION OF AND TRAINING FOR CEREBRAL ANEURYSM CLIPPING WITH 3‐DIMENSIONAL MODELS , 2009, Neurosurgery.

[21]  G. Biglino,et al.  In Vitro Simulation and Validation of the Circulation with Congenital Heart Defects. , 2010, Progress in pediatric cardiology.

[22]  M. Jerosch-Herold,et al.  Maladaptive Aortic Properties in Children After Palliation of Hypoplastic Left Heart Syndrome Assessed by Cardiovascular Magnetic Resonance Imaging , 2010, Circulation.

[23]  Philippe Pibarot,et al.  Assessment of left heart and pulmonary circulation flow dynamics by a new pulsed mock circulatory system , 2010 .

[24]  A. Taylor,et al.  First-in-man implantation of a novel percutaneous valve: a new approach to medical device development. , 2010, EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology.

[25]  Which stent for diabetic patients: the glass half-full or half-empty? , 2010, European heart journal.

[26]  P. Ugolini,et al.  Influence of age and sex on aortic distensibility assessed by MRI in healthy subjects. , 2010, Magnetic resonance imaging.

[27]  S. Sherwin,et al.  Pulse wave propagation in a model human arterial network: Assessment of 1-D visco-elastic simulations against in vitro measurements , 2011, Journal of biomechanics.

[28]  Giovanni Biglino,et al.  Virtual and real bench testing of a new percutaneous valve device: a case study. , 2012, EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology.

[29]  G. Biglino,et al.  In Vitro Study of the Norwood Palliation: A Patient-Specific Mock Circulatory System , 2012, ASAIO journal.

[30]  Giovanni Biglino,et al.  Finite Element Strategies to Satisfy Clinical and Engineering Requirements in the Field of Percutaneous Valves , 2012, Annals of Biomedical Engineering.

[31]  J. Pepper,et al.  A mock circulatory system with physiological distribution of terminal resistance and compliance: application for testing the intra-aortic balloon pump. , 2012, Artificial organs.