Energy dependence measurements of remmeters and albedo neutron dosimeters at neutron energies of thermal and between 2 keV and 5. 67 MeV
暂无分享,去创建一个
The instruments tested included a 9-inch sphere Portable Neutron Rem Counter Model PNR-4 manufactured by Eberline Instrument Corporation, Santa Fe, N.M.; an Andersson-Braum type Remmeter; and Hankins-type albedo neutron dosimeters composed of cadmium-enclosed TLDs. The standard neutron source at the NBS reactor was used. The observed and calculated response curves are presented and discussed. (WHK)