A study of the distribution of dose across the hands of interventional radiologists and cardiologists.

The magnitude and distribution of doses across the hands of interventional radiologists and cardiologists have been studied. The aims were to determine the region of highest dose, investigate variations in dose distribution, and propose an effective method for dose monitoring. Doses have been measured using sets of up to 18 thermoluminescent dosemeters (TLDs) for 183 single procedures. Important factors influencing the dose to the hand are the type of procedure, particularly the access route, the X-ray equipment used, and the experience of the operator. Radiologists performing percutaneous procedures received the highest doses, because of the proximity of their hands to the X-ray tube. The majority of procedures involve a combination of twisting and prodding actions, and the relative proportions of each determine the parts of the fingers which receive a higher dose. For most interventional radiology and cardiology procedures the bases of the ring and little fingers receive the highest dose. However, during percutaneous procedures the tips of the middle and ring fingers could receive doses which were 20-30% higher than this. For radiologists and cardiologists with a mixed workload, monitoring using TLD rings located at the base of the little or the ring fingers on either hand should provide a reasonable estimate of dose to the most exposed area. Monitoring is recommended for operators who may receive over 50 mSv to their hands per year, and should be considered for operators carrying out therapeutic procedures involving patient dose-area products over 500 Gy cm2 per month.

[1]  J. Law,et al.  Finger doses received by raediologists during Chiba needle percutaneous cholangiography. , 1980, The British journal of radiology.

[2]  W. Walker,et al.  Extremity doses during interventional radiology. , 1990, Clinical radiology.

[3]  J R Williams,et al.  The interdependence of staff and patient doses in interventional radiology. , 1997, The British journal of radiology.

[4]  T. Vehmas Radiation exposure during standard and complex interventional procedures. , 1997, The British journal of radiology.

[5]  C. Martin,et al.  Application of ALARP to extremity doses for hospital workers , 2003, Journal of radiological protection : official journal of the Society for Radiological Protection.

[6]  J. Damilakis,et al.  Radiation exposure to the hands of operators during angiographic procedures. , 1995, European journal of radiology.

[7]  T. Vehmas,et al.  Measuring radiation exposure during percutaneous drainages: can shoulder dosemeters be used to estimate finger doses? , 1992, The British journal of radiology.

[8]  A Spence,et al.  Ionising radiations regulations 1999. , 2000, The Veterinary record.

[9]  J. Law Doses to head and arms of radiologists during fluoroscopy. , 1985, The British journal of radiology.

[10]  C J Martin,et al.  Balancing patient dose and image quality. , 1999, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[11]  T. Vehmas What factors influence radiologists' finger doses during percutaneous drainages under fluoroscopic guidance? , 1993, Health physics.

[12]  An assessment of the radiation dose received by staff using fluoroscopic equipment. , 1982, The British journal of radiology.

[13]  H. W. Julius,et al.  Exposure of the radiologist to scattered radiation during angiography. , 1975, Radiology.

[14]  A. Lunderquist,et al.  Personnel exposure to radiation at some angiographic procedures. , 1981, Radiology.

[15]  J. Felmlee,et al.  Hand dose measurements in interventional radiology. , 1991, Health physics.

[16]  J H Ellis,et al.  Radiation protection of the hand in interventional radiology: should it fit like a glove? , 1996, Radiology.

[17]  C J Martin,et al.  Radiation doses to the legs of radiologists performing interventional procedures: are they a cause for concern? , 2003, The British journal of radiology.

[18]  F. Mettler,et al.  The radiologist and angiographic procedures. Absorbed radiation dose. , 1987, Investigative radiology.

[19]  F. Marshall,et al.  Monitoring radiation exposure to medical personnel during percutaneous nephrolithotomy. , 1986, Urology.

[20]  S. Mattsson,et al.  Radiation Dose to Patients and Personnel during Fluoroscopy at Percutaneous Renal Stone Extraction , 1989, Acta radiologica.

[21]  Miller Kl,et al.  Variations in dose to the extremities of vascular/interventional radiologists. , 1999 .

[22]  K. Faulkner Personal monitoring during fluoroscopy , 1992 .

[23]  Burgess Ae,et al.  Finger doses in special procedures. , 1984 .