Temporary persistence of conduction block after prolonged kilohertz frequency alternating current on rat sciatic nerve

OBJECTIVE Application of Kilohertz Frequency Alternating Current (KHFAC) waveforms can result in nerve conduction block that is induced in less than a second. Conduction recovers within seconds when KHFAC is applied for about 5 - 10 minutes. This study investigated the effect of repeated and prolonged application of KHFAC on rat Sciatic nerve with bipolar platinum electrodes. Approach: Varying durations of KHFAC at signal amplitudes for conduction block with intervals of no stimulus were studied. Nerve conduction was monitored by recording peak Gastrocnemius muscle force utilizing stimulation electrodes proximal (PS) and distal (DS) to a blocking electrode. The PS signal traveled through the block zone on the nerve, while the DS went directly to the motor end-plate junction. The PS/DS force ratio provided a measure of conduction patency of the nerve in the block zone. Main Results: Conduction recovery times were found to be significantly affected by the cumulative duration of KHFAC application. Peak stimulated muscle force returned to pre-block levels immediately after cessation of KHFAC delivery when it was applied for less than about 15 minutes. They fell significantly but recovered to near pre-block levels for cumulative stimulus between 50 +/- 20 minutes, for the tested On / Off times and frequencies. Conduction recovered in two phases, an initial fast one (60 -80% recovery), followed by a slower phase. No permanent conduction block was seen at the end of the observation period during any experiment. Significance: This Carry-over Block Effect (COBE) may be exploited to provide continuous conduction block in peripheral nerves without continuous application of KHFAC. .

[1]  K. Kilgore,et al.  Design, fabrication and evaluation of a conforming circumpolar peripheral nerve cuff electrode for acute experimental use , 2011, Journal of Neuroscience Methods.

[2]  Caixin Sun,et al.  Effects of Steep Pulsed Electric Fields (SPEF) on Mitochondrial Transmembrane Potential of Human Liver Cancer Cell , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[3]  Niloy Bhadra,et al.  High‐frequency electrical conduction block of mammalian peripheral motor nerve , 2005, Muscle & nerve.

[4]  J. Oliveira Mitochondrial Membrane Potential and Dynamics , 2012 .

[5]  K. Kilgore,et al.  Nerve conduction block utilising high-frequency alternating current , 2004, Medical and Biological Engineering and Computing.

[6]  K. Kilgore,et al.  Electrical conduction block in large nerves: High‐frequency current delivery in the nonhuman primate , 2011, Muscle & nerve.

[7]  Konstantinos Alataris,et al.  Effect of High‐Frequency Alternating Current on Spinal Afferent Nociceptive Transmission , 2013, Neuromodulation : journal of the International Neuromodulation Society.

[8]  K. Kilgore,et al.  Reversible Nerve Conduction Block Using Kilohertz Frequency Alternating Current , 2014, Neuromodulation : journal of the International Neuromodulation Society.

[9]  C. N. Honda,et al.  Effects of high-frequency alternating current on axonal conduction through the vagus nerve , 2011, Journal of neural engineering.

[10]  Tianshuang Qiu,et al.  Post stimulus effects of high frequency biphasic electrical current on a fibre's conductibility in isolated frog nerves , 2013, Journal of neural engineering.

[11]  Linlin Zhu,et al.  Effects of high frequency electrical stimulation on nerve's conduction of action potentials , 2011, 2011 4th International Conference on Biomedical Engineering and Informatics (BMEI).

[12]  Jicheng Wang,et al.  Post-stimulation block of frog sciatic nerve by high-frequency (kHz) biphasic stimulation , 2017, Medical & Biological Engineering & Computing.

[13]  Narendra Bhadra,et al.  Bladder voiding by combined high frequency electrical pudendal nerve block and sacral root stimulation , 2008, Neurourology and urodynamics.

[14]  Niloy Bhadra,et al.  Direct current contamination of kilohertz frequency alternating current waveforms , 2014, Journal of Neuroscience Methods.