Electrical Failure of an ICD Lead due to a Presumed Insulation Defect Only Diagnosed by a Maximum Output Shock

A 55‐year‐old male patient presented after a single shock caused by oversensing of isolated nonphysiologic signals on both the distal HV and pace‐sense channels. No other abnormalities were found. He subsequently returned complaining of device “vibration” and his St. Jude implantable defibrillator (ICD; St. Jude Medical, St. Paul, MN, USA) was found to be in VVI backup mode and could not be interrogated. Direct testing in the electrophysiology lab showed normal lead impedances and thresholds with an inability to reproduce the abnormal signals. Detailed cine fluoroscopy of the leads found no abnormalities. A new ICD was connected and successfully delivered a 20‐joule shock but failed to deliver a maximum output (39‐joule) shock. The new ICD was again found to be in backup mode. A new Endotak Reliance G lead (Boston Scientific, Natick, MA, USA) was implanted and a maximum‐output shock was successful using a new Fortify DR ICD. This case likely represents a Durata lead insulation defect in the form of an inside‐out abrasion under the distal HV coil. Increased awareness of this defect is warranted, particularly since routine interrogation and submaximum‐output shocks may fail to detect the problem.

[1]  S. Saba,et al.  Failure-free survival of the Durata defibrillator lead. , 2013, Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology.

[2]  C. Swerdlow,et al.  Inside-out insulation failure of a defibrillator lead with abrasion-resistant coating. , 2013, Heart rhythm.

[3]  L. Poggio,et al.  Failure of implantable cardioverter-defibrillator leads: a matter of lead size? , 2013, Heart rhythm.

[4]  Christopher R. Ellis,et al.  High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: fluoroscopic grading scale and correlation to extracted leads. , 2012, Heart rhythm.

[5]  R. Hauser,et al.  Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads. , 2012, Heart rhythm.

[6]  D. Leong,et al.  Unrecognized Failure of a Narrow Caliber Defibrillation Lead: The Role of Defibrillation Threshold Testing in Identifying an Unprotected Individual , 2012, Pacing and clinical electrophysiology : PACE.

[7]  R. Hauser,et al.  Riata implantable cardioverter-defibrillator lead failure: analysis of explanted leads with a unique insulation defect. , 2012, Heart rhythm.

[8]  R. Doshi,et al.  Is High-Voltage Lead Integrity Measurement Adequate During Defibrillator Generator Replacement ? , 2012 .

[9]  U. Lakshmanadoss,et al.  Riata lead failure with normal electrical lead parameters and normal fluoroscopic appearance , 2012, Journal of Interventional Cardiac Electrophysiology.

[10]  H. Calkins,et al.  Premature Failure of a Riata Defibrillator Lead Without Impedance Change or Inappropriate Sensing: A Case Report and Review of the Literature , 2011, Journal of cardiovascular electrophysiology.