On-line technique for preparing and measuring stable carbon isotope of total dissolved inorganic carbon in water samples (δ 13 CTDIC)

A fast and completely automated procedure is proposed for the preparation and determination of δ 13 C of total inorganic carbon dissolved in water (δ 13 CTDIC). This method is based on the acidification of water samples transforming the whole dissolved inorganic carbon species into CO2. Water samples are directly injected by syringe into 5.9 ml vials with screw caps which have a pierciable rubber septum. An Analytical Precision «Carbonate Prep System» was used both to flush pure helium into the vials and to automatically dispense a fixed amount of H3PO4. Full-equilibrium conditions between produced CO2 and water are reached at a temperature of 70°C (± 0.1°C) in less than 24 h. Carbon isotope ratios ( 13 C/ 12 C) were measured on an AP 2003 continuous flow mass spectrometer, connected on-line with the injection system. The precision and reproducibility of the proposed method was tested both on aqueous standard solutions prepared using Na2CO3 with δ