Accuracy of tumor motion compensation algorithm from a robotic respiratory tracking system: a simulation study.

The Synchrony Respiratory Tracking System (RTS) is a treatment option of the CyberKnife robotic treatment device to irradiate extra-cranial tumors that move due to respiration. Advantages of RTS are that patients can breath normally and that there is no loss of linac duty cycle such as with gated therapy. Tracking is based on a measured correspondence model (linear or polynomial) between internal tumor motion and external (chest/abdominal) marker motion. The radiation beam follows the tumor movement via the continuously measured external marker motion. To establish the correspondence model at the start of treatment, the 3D internal tumor position is determined at 15 discrete time points by automatic detection of implanted gold fiducials in two orthogonal x-ray images; simultaneously, the positions of the external markers are measured. During the treatment, the relationship between internal and external marker positions is continuously accounted for and is regularly checked and updated. Here we use computer simulations based on continuously and simultaneously recorded internal and external marker positions to investigate the effectiveness of tumor tracking by the RTS. The Cyberknife does not allow continuous acquisition of x-ray images to follow the moving internal markers (typical imaging frequency is once per minute). Therefore, for the simulations, we have used data for eight lung cancer patients treated with respiratory gating. All of these patients had simultaneous and continuous recordings of both internal tumor motion and external abdominal motion. The available continuous relationship between internal and external markers for these patients allowed investigation of the consequences of the lower acquisition frequency of the RTS. With the use of the RTS, simulated treatment errors due to breathing motion were reduced largely and consistently over treatment time for all studied patients. A considerable part of the maximum reduction in treatment error could already be reached with a simple linear model. In case of hysteresis, a polynomial model added some extra reduction. More frequent updating of the correspondence model resulted in slightly smaller errors only for the few recordings with a time trend that was fast, relative to the current x-ray update frequency. In general, the simulations suggest that the applied combined use of internal and external markers allow the robot to accurately follow tumor motion even in the case of irregularities in breathing patterns.

[1]  Steven D Chang,et al.  Robotics and Radiosurgery – The Cyberknife , 2002, Stereotactic and Functional Neurosurgery.

[2]  J. McClelland,et al.  MRI-based measurements of respiratory motion variability and assessment of imaging strategies for radiotherapy planning , 2006, Physics in medicine and biology.

[3]  M J Murphy,et al.  The Cyberknife: a frameless robotic system for radiosurgery. , 1997, Stereotactic and functional neurosurgery.

[4]  S Minohara,et al.  Respiratory gated irradiation system for heavy-ion radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[5]  H. Kubo,et al.  Respiration gated radiotherapy treatment: a technical study. , 1996, Physics in medicine and biology.

[6]  Steven D Chang,et al.  An Analysis of the Accuracy of the CyberKnife: A Robotic Frameless Stereotactic Radiosurgical System , 2003, Neurosurgery.

[7]  Koichi Yamazaki,et al.  Insertion and fixation of fiducial markers for setup and tracking of lung tumors in radiotherapy. , 2005, International journal of radiation oncology, biology, physics.

[8]  M. V. van Herk,et al.  Physical aspects of a real-time tumor-tracking system for gated radiotherapy. , 2000, International journal of radiation oncology, biology, physics.

[9]  G Kim Prisk,et al.  Techniques for measurement of thoracoabdominal asynchrony , 2002, Pediatric pulmonology.

[10]  Kristy K Brock,et al.  Accuracy of daily image guidance for hypofractionated liver radiotherapy with active breathing control. , 2005, International journal of radiation oncology, biology, physics.

[11]  Martin J Murphy,et al.  Issues in respiratory motion compensation during external-beam radiotherapy. , 2002, International journal of radiation oncology, biology, physics.

[12]  William Main,et al.  An Analysis of the Accuracy of the CyberKnife: A Robotic Frameless Stereotactic Radiosurgical System , 2003, Neurosurgery.

[13]  Lili Wang,et al.  Introduction of audio gating to further reduce organ motion in breathing synchronized radiotherapy. , 2002, Medical physics.

[14]  J. Wong,et al.  The use of active breathing control (ABC) to reduce margin for breathing motion. , 1999, International journal of radiation oncology, biology, physics.

[15]  M. V. van Herk,et al.  Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. , 2002, International journal of radiation oncology, biology, physics.

[16]  Uwe Oelfke,et al.  Online correction for respiratory motion: evaluation of two different imaging geometries , 2005, Physics in medicine and biology.

[17]  Shinichi Shimizu,et al.  Intrafractional tumor motion: lung and liver. , 2004, Seminars in radiation oncology.

[18]  Koichi Yamazaki,et al.  Feasibility of insertion/implantation of 2.0-mm-diameter gold internal fiducial markers for precise setup and real-time tumor tracking in radiotherapy. , 2003, International journal of radiation oncology, biology, physics.

[19]  Steve B. Jiang,et al.  Residual motion of lung tumours in gated radiotherapy with external respiratory surrogates , 2005, Physics in medicine and biology.

[20]  Geoffrey Hugo,et al.  Changes in the respiratory pattern during radiotherapy for cancer in the lung. , 2006, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[21]  J. Adler,et al.  Robotic Motion Compensation for Respiratory Movement during Radiosurgery , 2000, Computer aided surgery : official journal of the International Society for Computer Aided Surgery.

[22]  Michael Flentje,et al.  Impact of target reproducibility on tumor dose in stereotactic radiotherapy of targets in the lung and liver. , 2003, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[23]  Y. Tsunashima,et al.  Correlation between the respiratory waveform measured using a respiratory sensor and 3D tumor motion in gated radiotherapy. , 2004, International journal of radiation oncology, biology, physics.