Improvement in Proximal Aortic Endograft Fixation: An Experimental Study Using Different Stent-Grafts in Human Cadaveric Aortas

Purpose: To compare the proximal fixation characteristics of different types of stent-grafts (SG) and the adjunctive effect of a proximally placed Palmaz stent. Methods: Human cadaveric aortas were obtained at autopsy and cut into 30 aortic segments. Appropriately oversized (10% to 20%) commercial stent-grafts (Zenith, Ancure, Excluder, Talent) and 4 configurations of homemade Palmaz-based stent-grafts (polytetrafluoroethylene or polyester coverings each with no uncovered proximal edge or a 20-mm bare section of stent) were implanted 20 mm into an aortic segment and balloon dilated. Each segment was placed in an experimental apparatus to measure the dislodgment force. Bare Palmaz stents were used to reinforce the proximal fixation of the commercial stent-grafts, and the measurements were repeated. Manual anastomoses were made and their dislodgment force tested for comparison. Results: The median dislodgment force ranged from 6.5 N for the Excluder to 26.5 N for the Zenith (8.0 N for the Talent, 11.8 N for the Ancure, and 8.1 to 10.7 N for the various Palmaz stent-graft designs). There was no significant difference between the Zenith and the Ancure groups or between the Excluder and the Talent groups. However, the Zenith and Ancure devices required significantly higher (p = 0.0004) force (∼25%) to displace them than the other stent-grafts tested. With the Palmaz stent added to the proximal attachment site, the median dislodgment force for the different commercial endografts was significantly improved for all devices (p < 0.03): 34% for the Zenith device, 69% for the Ancure, 73% for the Talent, and 80% for the Excluder endoprosthesis. Conclusions: The Zenith stent-graft had the best resistance to dislodgment. An additional Palmaz stent placed at the proximal attachment site greatly improves endograft fixation regardless of the type of stent-graft. For complex aneurysm necks or for intraoperative type I endoleak management, an adjunctive Palmaz stent could be used.

[1]  R. Green,et al.  Fate of the proximal aortic cuff: implications for endovascular aneurysm repair. , 1997, Journal of vascular surgery.

[2]  B. Hopkinson,et al.  Perioperative renal function following endovascular repair of abdominal aortic aneurysm with suprarenal and infrarenal stents , 1999, The British journal of surgery.

[3]  T. Resch,et al.  Intraoperative Proximal Endoleaks during AAA Stent-Graft Repair: Evaluation of Risk Factors and Treatment with Palmaz Stents , 2001 .

[4]  J. H. van Bockel,et al.  Endoleak after stent‐graft treatment of abdominal aortic aneurysm: a meta‐analysis of clinical studies , 1999, The British journal of surgery.

[5]  B. Hopkinson,et al.  Adjunctive Procedures for the Treatment of Proximal Type I Endoleak: The Role of Peri-Aortic Ligatures and Palmaz Stenting , 2003, Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists.

[6]  T. Resch,et al.  The impact of stent design on proximal stent-graft fixation in the abdominal aorta: an experimental study. , 2000, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[7]  P. Cluzel,et al.  Renal function after aortic stent-grafting including coverage of renal arterial ostia. , 2002, American journal of kidney diseases : the official journal of the National Kidney Foundation.

[8]  Greenhalgh Indications in Vascular and Endovascular Surgery , 1998 .

[9]  E. Chaikof,et al.  Continued expansion of aortic necks after endovascular repair of abdominal aortic aneurysms. EVT Investigators. EndoVascular Technologies, Inc. , 1998, Journal of vascular surgery.

[10]  D. Gould,et al.  Longitudinal aneurysm shrinkage following endovascular aortic aneurysm repair: a source of intermediate and late complications. , 1999 .

[11]  P. Gaines,et al.  The effects of crossing porcine renal artery ostia with various endovascular stents. , 1999, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[12]  J. Hamming,et al.  Complications in the endovascular repair of abdominal aortic aneurysms: a risk factor analysis. , 1999, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[13]  M. Lindh,et al.  Renal arteries covered by aortic stents: clinical experience from endovascular grafting of aortic aneurysms. , 1997, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[14]  R. Green,et al.  An update of the Zenith endovascular graft for abdominal aortic aneurysms: initial implantation and mid-term follow-up data. , 2001, Journal of vascular surgery.

[15]  M. Lindh,et al.  Endovascular AAA exclusion: will stents with hooks and barbs prevent stent-graft migration? , 1998, Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery.

[16]  M. Picot,et al.  Long-Term Renal Function following Endovascular Aneurysm Repair with Infrarenal and Suprarenal Aortic Stent-Grafts , 2003, Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists.

[17]  J. H. van Bockel,et al.  Stent attachment site-related endoleakage after stent graft treatment: An in vitro study of the effects of graft size, stent type, and atherosclerotic wall changes. , 1999, Journal of vascular surgery.

[18]  M. Horrocks,et al.  Experimental assessment of proximal stent-graft (InterVascular) fixation in human cadaveric infrarenal aortas. , 1999, European Journal of Vascular and Endovascular Surgery.

[19]  T. Resch,et al.  Endovascular management of the juxtarenal aortic aneurysm: can uncovered stents safely cross the renal arteries? , 1999, Seminars in vascular surgery.

[20]  L. Hollier,et al.  Impact of transrenal aortic endograft placement on endovascular graft repair of abdominal aortic aneurysms. , 1998, Journal of vascular surgery.

[21]  G. White,et al.  A Prospective Study of Changes in Morphology and Dimensions of Abdominal Aortic Aneurysms following Endoluminal Repair: A Preliminary Report , 1995, Journal of endovascular surgery : the official journal of the International Society for Endovascular Surgery.

[22]  S M Andrews,et al.  In vitro evaluation of endovascular stents to assess suitability for endovascular graft fixation. , 1995, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[23]  T. Resch,et al.  Distal migration of stent-grafts after endovascular repair of abdominal aortic aneurysms. , 1999, Journal of vascular and interventional radiology : JVIR.

[24]  B. Lindblad,et al.  Endovascular healing is inadequate for fixation of Dacron stent-grafts in human aortoiliac vessels. , 2000, European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery.

[25]  G. White,et al.  Endoleak as a complication of endoluminal grafting of abdominal aortic aneurysms: classification, incidence, diagnosis, and management. , 1997 .

[26]  F. Veith,et al.  Significance of endotension, endoleak, and aneurysm pulsatility after endovascular repair. , 2003, Journal of vascular surgery.