TG-69: radiographic film for megavoltage beam dosimetry.

TG-69 is a task group report of the AAPM on the use of radiographic film for dosimetry. Radiographic films have been used for radiation dosimetry since the discovery of x-rays and have become an integral part of dose verification for both routine quality assurance and for complex treatments such as soft wedges (dynamic and virtual), intensity modulated radiation therapy (IMRT), image guided radiation therapy (IGRT), and small field dosimetry like stereotactic radiosurgery. Film is convenient to use, spatially accurate, and provides a permanent record of the integrated two dimensional dose distributions. However, there are several challenges to obtaining high quality dosimetric results with film, namely, the dependence of optical density on photon energy, field size, depth, film batch sensitivity differences, film orientation, processing conditions, and scanner performance. Prior to the clinical implementation of a film dosimetry program, the film, processor, and scanner need to be tested to characterize them with respect to these variables. Also, the physicist must understand the basic characteristics of all components of film dosimetry systems. The primary mission of this task group report is to provide guidelines for film selection, irradiation, processing, scanning, and interpretation to allow the physicist to accurately and precisely measure dose with film. Additionally, we present the basic principles and characteristics of film, processors, and scanners. Procedural recommendations are made for each of the steps required for film dosimetry and guidance is given regarding expected levels of accuracy. Finally, some clinical applications of film dosimetry are discussed.

[1]  Nevill Mott,et al.  The Theory of the Photolysis of Silver Bromide and the Photographic Latent Image , 1938 .

[2]  C. E. SHANNON,et al.  A mathematical theory of communication , 1948, MOCO.

[3]  J. G. Trump,et al.  The film method of tissue dose studies with 2.0 mev. roentgen rays. , 1954, The American journal of roentgenology, radium therapy, and nuclear medicine.

[4]  M. Ehrlich Reciprocity Law for X-Rays.* Part II: Failure in the Reversal Region , 1956 .

[5]  A. Dutreix,et al.  FILM DOSIMETRY OF HIGH‐ENERGY ELECTRONS , 1969, Annals of the New York Academy of Sciences.

[6]  The Photographic Action of Ionizing Radiations , 1970 .

[7]  P K Kijewski,et al.  Wedge-shaped dose distributions by computer-controlled collimator motion. , 1978, Medical physics.

[8]  [The value and limitations of dosimetry by the use of film in cobalt 60 beams (author's transl)]. , 1979, Journal de radiologie.

[9]  D W Anderson,et al.  Comparison of film and ion chamber systems for depth-dose measurements for a 25 MV beam. , 1979, Physics in medicine and biology.

[10]  S C Sharma,et al.  Film dosimetry of megavoltage photon beams: a practical method of isodensity-to-isodose curve conversion. , 1981, Medical physics.

[11]  F. Khan The physics of radiation therapy , 1985 .

[12]  R L Siddon,et al.  Effective wedge angles with a universal wedge. , 1985, Physics in medicine and biology.

[13]  D D Leavitt,et al.  Dynamic wedge field techniques through computer-controlled collimator motion and dose delivery. , 1990, Medical physics.

[14]  M L Giger,et al.  Measurement of the presampling modulation transfer function of film digitizers using a curve fitting technique. , 1990, Medical physics.

[15]  R. Nath,et al.  Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetry. , 1991, Medical physics.

[16]  Three-dimensional photon treatment planning. Report of the Collaborative Working Group on the evaluation of treatment planning for external photon beam radiotherapy. , 1991, International journal of radiation oncology, biology, physics.

[17]  G K Svensson,et al.  Quality assurance in stereotactic radiosurgery using a standard linear accelerator. , 1991, International journal of radiation oncology, biology, physics.

[18]  F. H. Attix Introduction to Radiological Physics and Radiation Dosimetry , 1991 .

[19]  Kunio Doi,et al.  A simple method for determining the modulation transfer function in digital radiography , 1992, IEEE Trans. Medical Imaging.

[20]  K. Lam,et al.  Generation and use of measurement-based 3-D dose distributions for 3-D dose calculation verification. , 1992, Medical physics.

[21]  Photon beam dosimetry at a blocked beam edge using diffusion approximation. , 1992, Physics in medicine and biology.

[22]  John C. Russ,et al.  The Image Processing Handbook , 2016, Microscopy and Microanalysis.

[23]  M. Evans,et al.  A simple technique for film dosimetry. , 1992, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[24]  J. Cygler,et al.  Commissioning and quality assurance of treatment planning computers. , 1993, International journal of radiation oncology, biology, physics.

[25]  D. Leavitt,et al.  Evaluation of a diode detector array for measurement of dynamic wedge dose distributions. , 1993, Medical physics.

[26]  John C. Russ,et al.  The image processing handbook (3. ed.) , 1995 .

[27]  Jun Wang,et al.  Film digitization aliasing artifacts caused by grid line patterns , 1994, IEEE Trans. Medical Imaging.

[28]  M H Idzes,et al.  Film dosimetry for radiotherapy treatment planning verification of a 6 MV tangential breast irradiation. , 1994, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[29]  J. Palta,et al.  Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40. , 1994, Medical physics.

[30]  P. Shragge,et al.  Calibration of film for accurate megavoltage photon dosimetry. , 1994, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[31]  Film dosimetry in water in a 23 MV therapeutic photon beam. , 1995 .

[32]  L. V. van Battum,et al.  Film dosimetry in water in a 23 MV therapeutic photon beam. , 1995, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[33]  D A Low,et al.  Dosimetry and clinical implementation of dynamic wedge. , 1995, International Journal of Radiation Oncology, Biology, Physics.

[34]  A. Welsh,et al.  An investigation into the comparison between different dosimetric methods of measuring profiles and depth doses for dynamic wedges on a Varian 600C linear accelerator. , 1995, Physics in medicine and biology.

[35]  P. Childs,et al.  Beam data measurements for dynamic wedges on Varian 600C (6 MV) and 2100C (6 and 10 MV) linear accelerators. , 1995, Physics in medicine and biology.

[36]  D A Jaffray,et al.  Tests for evaluating laser film digitizers. , 1995, Medical physics.

[37]  I. Das,et al.  Dosimetry of high energy photon and electron beams with CEA films. , 1996, Medical physics.

[38]  P W Hoban,et al.  A comparison of dosimetry techniques in stereotactic radiosurgery. , 1996, Physics in medicine and biology.

[39]  A W Beavis,et al.  Implementation of the Varian EDW into a commercial RTP system. , 1996, Physics in medicine and biology.

[40]  T LoSasso,et al.  Testing of dynamic multileaf collimation. , 1996, Medical physics.

[41]  Portal imaging with film-cassette combinations: what film should we use? , 1996 .

[42]  T C Zhu,et al.  Characterizing output for dynamic wedges. , 1996, Medical physics.

[43]  I J Yeo,et al.  A filtration method for improving film dosimetry in photon radiation therapy. , 1997, Medical physics.

[44]  H M Kooy,et al.  Comparison of miniature multileaf collimation (MMLC) with circular collimation for stereotactic treatment. , 1997, International journal of radiation oncology, biology, physics.

[45]  L. Reinstein,et al.  Comparison of dose response of radiochromic film measured with He-Ne laser, broadband, and filtered light densitometers. , 1997, Medical physics.

[46]  J F Williamson,et al.  Quantitative evaluation of radiochromic film response for two-dimensional dosimetry. , 1997, Medical physics.

[47]  D. Verellen,et al.  Initial experience with intensity-modulated conformal radiation therapy for treatment of the head and neck region. , 1997, International journal of radiation oncology, biology, physics.

[48]  A. Haus,et al.  Processor quality control in laser imaging systems. , 1997, Medical physics.

[49]  S. Burch,et al.  A new approach to film dosimetry for high energy photon beams: lateral scatter filtering. , 1997, Medical physics.

[50]  C. De Wagter,et al.  Characteristics of a commercially available film digitizer and their significance for film dosimetry. , 1998, Physics in medicine and biology.

[51]  D. Low,et al.  Phantoms for IMRT dose distribution measurement and treatment verification. , 1998, International journal of radiation oncology, biology, physics.

[52]  J A Purdy,et al.  Multiple machine implementation of enhanced dynamic wedge. , 1998, International journal of radiation oncology, biology, physics.

[53]  C. Ling,et al.  Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. , 1998, Medical physics.

[54]  J R Sykes,et al.  An experimental investigation of the tongue and groove effect for the Philips multileaf collimator. , 1998, Physics in medicine and biology.

[55]  J. Tsai,et al.  Dosimetric verification of the dynamic intensity-modulated radiation therapy of 92 patients. , 1998, International journal of radiation oncology, biology, physics.

[56]  G. Starkschall,et al.  American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: quality assurance for clinical radiotherapy treatment planning. , 1998, Medical physics.

[57]  Use of CEA TVS film for measuring high energy photon beam dose distributions. , 1998, Medical physics.

[58]  J F Williamson,et al.  Quantitative optical densitometry with scanning-laser film digitizers. , 1999, Medical physics.

[59]  N Suchowerska,et al.  Perturbation of radiotherapy beams by radiographic film: measurements and Monte Carlo simulations. , 1999, Physics in medicine and biology.

[60]  B. Clark,et al.  The use of radiographic film for linear accelerator stereotactic radiosurgical dosimetry. , 1999, Medical physics.

[61]  How much does film sensitivity increase at depth for larger field sizes? , 1999, Medical physics.

[62]  A new approach to film dosimetry for high-energy photon beams using organic plastic scintillators. , 1999, Physics in medicine and biology.

[63]  B. Clark,et al.  A practical technique for verification of three-dimensional conformal dose distributions in stereotactic radiosurgery. , 2000, Medical physics.

[64]  I. Das,et al.  Choice of Radiation Detector in Dosimetry of Stereotactic Radiosurgery-Radiotherapy , 2000 .

[65]  T LoSasso,et al.  Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode. , 2001, Medical physics.

[66]  CCD film digitizers in clinical practice: evaluation of the main properties. , 2001, Medical informatics and the Internet in medicine.

[67]  J. Bourland American Association of Physicists in Medicine , 2001 .

[68]  N. Suchowerska,et al.  Directional dependence in film dosimetry: radiographic and radiochromic film. , 2001, Physics in medicine and biology.

[69]  Mpp Advances in Film Processing Systems Technology and Quality Control in Medical Imaging , 2001 .

[70]  J C Rosenwald,et al.  Variation of sensitometric curves of radiographic films in high energy photon beams. , 2001, Medical physics.

[71]  Lei Dong,et al.  Rapid radiographic film calibration for IMRT verification using automated MLC fields. , 2002, Medical physics.

[72]  The value of radiographic film for the characterization of intensity-modulated beams. , 2002, Physics in medicine and biology.

[73]  Étude d’un film récemment commercialisé pour la dosimétrie des faisceaux de photons de haute énergie , 2002 .

[74]  Nesrin Dogan,et al.  Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy. , 2002, Physics in medicine and biology.

[75]  L E Reinstein,et al.  Comparison of three high-resolution digitizers for radiochromic film dosimetry. , 2002, Medical physics.

[76]  J. Rownd,et al.  Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static multileaf collimator. , 2002, Medical physics.

[77]  [Study of a new commercial film for high energy photon dosimetry]. , 2002, Cancer radiotherapie : journal de la Societe francaise de radiotherapie oncologique.

[78]  J. Dempsey,et al.  Toward automated quality assurance for intensity- modulated radiation therapy. , 2002, International journal of radiation oncology, biology, physics.

[79]  A. Olch Dosimetric performance of an enhanced dose range radiographic film for intensity-modulated radiation therapy quality assurance. , 2002, Medical physics.

[80]  Inhwan Jason Yeo,et al.  Film dosimetry for intensity modulated radiation therapy: dosimetric evaluation. , 2002, Medical physics.

[81]  Investigation of Kodak extended dose range (EDR) film for megavoltage photon beam dosimetry. , 2002, Physics in medicine and biology.

[82]  J. Dempsey,et al.  Dosimetry of therapeutic photon beams using an extended dose range film. , 2002, Medical physics.

[83]  B G Fallone,et al.  Dosimetric IMRT verification with a flat-panel EPID. , 2003, Medical physics.

[84]  Peter B Greer,et al.  Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy. , 2003, Medical physics.

[85]  R. Pötter,et al.  Normalized sensitometric curves for the verification of hybrid IMRT treatment plans with multiple energies. , 2003, Medical physics.

[86]  Contribution of Cerenkov radiation in high-energy x-ray and electron beam film dosimetry using water-substitute phantoms. , 2003, Physics in medicine and biology.

[87]  Dosimetric verification of x-ray fields with steep dose gradients using an electronic portal imaging device. , 2003, Physics in medicine and biology.

[88]  C. Ling,et al.  Using the frame averaging of aS500 EPID for IMRT verification , 2003, Journal of applied clinical medical physics.

[89]  The response of Kodak EDR2 film in high-energy electron beams. , 2003, Medical physics.

[90]  I. Chetty,et al.  Dosimetric comparison of extended dose range film with ionization measurements in water and lung equivalent heterogeneous media exposed to megavoltage photons , 2003, Journal of applied clinical medical physics.

[91]  P. Jursinic,et al.  A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery. , 2003, Medical physics.

[92]  Thomas LoSasso,et al.  Predicting energy response of radiographic film in a 6 MV x-ray beam using Monte Carlo calculated fluence spectra and absorbed dose. , 2004, Medical physics.

[93]  I. Yeo,et al.  EDR2 film dosimetry for IMRT verification using low-energy photon filters. , 2004, Medical physics.

[94]  I. Rosen,et al.  Effect of processing time delay on the dose response of Kodak EDR2 film. , 2004, Medical physics.

[95]  R. Mohan,et al.  Monte Carlo computation of dosimetric amorphous silicon electronic portal images. , 2004, Medical physics.

[96]  M. Oldham,et al.  Evaluation of a 2D diode array for IMRT quality assurance. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[97]  J. Dempsey,et al.  Comparison of small photon beams measured using radiochromic and silver-halide films in solid water phantoms. , 2004, Medical physics.

[98]  M. Bucciolini,et al.  Verification of IMRT fields by film dosimetry. , 2003, Medical physics.

[99]  Tom Depuydt,et al.  The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[100]  Calibration of Kodak EDR2 film for patient skin dose assessment in cardiac catheterization procedures. , 2004, Physics in medicine and biology.

[101]  Lei Dong,et al.  Dosimetric accuracy of Kodak EDR2 film for IMRT verifications. , 2005, Medical physics.

[102]  T. Losasso,et al.  Influence of phantom material and phantom size on radiographic film response in therapy photon beams. , 2005, Medical physics.

[103]  C De Wagter,et al.  Practical and dosimetric implications of a new type of packaging for radiographic film. , 2005, Physics in medicine and biology.

[104]  R. Kollhoff,et al.  The Schwarzschild effect of the dosimetry film Kodak EDR 2 , 2005, Physics in medicine and biology.

[106]  Physical mechanism of the Schwarzschild effect in film dosimetry--theoretical model and comparison with experiments. , 2006, Physics in medicine and biology.

[107]  Sang Joon Kim,et al.  A Mathematical Theory of Communication , 2006 .

[108]  SU-FF-T-174: Dose Rate Dependence of Film Dosimetry in Radiation Treatment: Study of Reciprocity Law , 2006 .

[109]  Frank Herbert Attix,et al.  Introduction to Radiological Physics and Radiation Dosimetry: Attix/Introduction , 2007 .