An evaluation of accuracy and repeatability of a novel gait analysis device

IntroductionRestoration of gait is a particularly important patient-based functional outcome following lower extremity trauma. A new portable device which measures gait parameters, the IDEEA, may be particularly useful in evaluating post-traumatic gait parameters in the office setting, but the accuracy and repeatability of this device relative to standard gait laboratory footswitches are unknown.Material and methodsTwelve healthy subjects were tested simultaneously using the IDEEA device and standard gait laboratory footswitches, at slow, medium, and fast speeds. Parameters evaluated were single-limb support time (SLS), double-limb support time (DLS), swing phase duration (SPD), cycle duration, and cadence.ResultsThe repeatability between right and left measurements tended to be better for the IDEEA, and was at least as good as the foot switches. The absolute differences of the timed gait parameters between the two devices were all in the range of 0.03 s, which is within the data sampling resolution of the gait laboratory foot switches (0.04 s). Furthermore, assuming a 1-s gait cycle, these differences account for only 3% of the gait cycle, which is also well within the clinical parameters for evaluating and differentiating between treatments.ConclusionThis device is accurate and repeatable, and may facilitate the evaluation of gait function in post-traumatic patients in settings outside of the traditional gait laboratory.

[1]  D A Boone,et al.  Step activity monitor: long-term, continuous recording of ambulatory function. , 1999, Journal of rehabilitation research and development.

[2]  F. Coutts,et al.  Gait analysis in the therapeutic environment. , 1999, Manual therapy.

[3]  Kuan Zhang,et al.  Measurement of human daily physical activity. , 2003, Obesity research.

[4]  F. Kummer,et al.  Postoperative Weight-Bearing after a Fracture of the Femoral Neck or an Intertrochanteric Fracture* , 1998, The Journal of bone and joint surgery. American volume.

[5]  T P Schmalzried,et al.  Step activity monitor: Increased accuracy in quantifying ambulatory activity , 1999, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[6]  Jeffrey M. Hausdorff,et al.  Footswitch system for measurement of the temporal parameters of gait. , 1995, Journal of biomechanics.

[7]  O. Löfman,et al.  Health-related quality of life after osteoporotic fractures , 2004, Osteoporosis International.

[8]  B. E. Maki,et al.  Gait Changes in Older Adults: Predictors of Falls or Indicators of Fear? , 1997, Journal of the American Geriatrics Society.

[9]  J. Otis,et al.  The relationship of stride characteristics to pain before and after total knee arthroplasty. , 1989, Clinical orthopaedics and related research.

[10]  M. J. Muêller,et al.  Monitoring in-shoe plantar pressures, temperature, and humidity: reliability and validity of measures from a portable device. , 2001, Archives of physical medicine and rehabilitation.

[11]  L Klenerman,et al.  Gait analysis in the assessment of functional performance before and after total hip replacement. , 1981, Journal of biomedical engineering.

[12]  D. Sutherland,et al.  A practical guide to gait analysis. , 2002, The Journal of the American Academy of Orthopaedic Surgeons.

[13]  Salim Ghoussayni,et al.  Assessment and validation of a simple automated method for the detection of gait events and intervals. , 2004, Gait & posture.

[14]  T. Mittlmeier,et al.  Analysis of Morphology and Gait Function After Intraarticular Calcaneal Fracture , 1993, Journal of orthopaedic trauma.

[15]  Clinical gait analyzer. , 1974, Bulletin of prosthetics research.

[16]  S. Silverman,et al.  Deterioration in Quality of Life Following Hip Fracture: A Prospective Study , 2000, Osteoporosis International.

[17]  Sheldon R Simon,et al.  Quantification of human motion: gait analysis-benefits and limitations to its application to clinical problems. , 2004, Journal of biomechanics.

[18]  Joseph M. Czerniecki Gait analysis in the amputee: Has it helped the amputee or contributed to the development of improved prosthetic components? , 1995 .

[19]  Kuan Zhang,et al.  Improving energy expenditure estimation for physical activity. , 2004, Medicine and science in sports and exercise.

[20]  R. Craik,et al.  Can elderly patients who have had a hip fracture perform moderate- to high-intensity exercise at home? , 2005, Physical therapy.