Functional Performance of Different Venous Limb Options in Simulated Neonatal/Pediatric Cardiopulmonary Bypass Circuits

Objective Hemodilution is a concern in cardiopulmonary bypass (CPB). Using a smaller dual tubing rather than a single larger inner diameter (ID) tubing in the venous limb to decrease prime volume has been a standard practice. The purpose of this study is to evaluate these tubing options. Methods Four different CPB circuits primed with blood (hematocrit 30%) were investigated. Two setups were used with two circuits for each one. In Setup I, a neonatal oxygenator was connected to dual 3/16" ID venous limbs (Circuit A) or to a single 1/4" ID venous limb (Circuit B); and in Setup II, a pediatric oxygenator was connected to dual 1/4" ID venous limbs (Circuit C) or a single 3/8" ID venous limb (Circuit D). Trials were conducted at arterial flow rates of 500 ml/min up to 1500 ml/min (Setup I) and up to 3000 ml/min (Setup II), at 36°C and 28°C. Results Circuit B exhibited a higher venous flow rate than Circuit A, and Circuit D exhibited a higher venous flow rate than Circuit C, at both temperatures. Flow resistance was significantly higher in Circuits A and C than in Circuits B (P<0.001) and D (P<0.001), respectively. Conclusion A single 1/4" venous limb is better than dual 3/16" venous limbs at all flow rates, up to 1500 ml/min. Moreover, a single 3/8" venous limb is better than dual 1/4" venous limbs, up to 3000 ml/min. Our findings strongly suggest a revision of perfusion practice to include single venous limb circuits for CPB.

[1]  A. Ündar Translational Research on Evaluation of Pediatric Cardiopulmonary Bypass Oxygenators. , 2018, Artificial organs.

[2]  R. Neirotti,et al.  Translational Research: Comparing Oxygenators From Different International Markets. , 2018, Artificial organs.

[3]  A. Ündar,et al.  In Vitro Evaluation of Pediatric Hollow‐Fiber Membrane Oxygenators on Hemodynamic Performance and Gaseous Microemboli Handling: An International Multicenter/Multidisciplinary Approach , 2017, Artificial organs.

[4]  A. Ündar,et al.  Impact of Translational Research on Optimization of Neonatal Cardiopulmonary Bypass Circuits and Techniques—The Penn State Health Approach , 2017, Artificial organs.

[5]  A. Undar,et al.  In-Vitro Evaluation of Two Types of Neonatal Oxygenators in Handling Gaseous Microemboli and Maintaining Optimal Hemodynamic Stability During Cardiopulmonary Bypass , 2016, Brazilian journal of cardiovascular surgery.

[6]  R. Neirotti Cardiopulmonary Bypass: a Forgotten Area of Searching for New Knowledge in Brazil and the Importance of Translational Research , 2016, Brazilian journal of cardiovascular surgery.

[7]  M. van Meurs,et al.  Impaired microcirculatory perfusion in a rat model of cardiopulmonary bypass: the role of hemodilution. , 2016, American journal of physiology. Heart and circulatory physiology.

[8]  J. Cooper,et al.  Hemodilution and priming solutions , 2015 .

[9]  G. Matte Perfusion for Congenital Heart Surgery: Notes on Cardiopulmonary Bypass for a Complex Patient Population , 2015 .

[10]  A. Ündar,et al.  Evaluation of different diameter arterial tubing and arterial cannulae in simulated neonatal/pediatric cardiopulmonary bypass circuits. , 2015, Artificial organs.

[11]  D. Likosky,et al.  Embolic activity during in vivo cardiopulmonary bypass. , 2014, The journal of extra-corporeal technology.

[12]  J. Clark,et al.  The impact of cerebral embolization during infant cardiac surgery on neurodevelopmental outcomes at intermediate follow-up , 2014, Perfusion.

[13]  C. Knott-Craig,et al.  Bloodless surgery in a pediatric Jehovah's Witness. , 2013, The journal of extra-corporeal technology.

[14]  A. Undar,et al.  Vacuum-assisted venous drainage and gaseous microemboli in cardiopulmonary bypass. , 2008, The journal of extra-corporeal technology.

[15]  A. Ündar,et al.  Delivery of Gaseous Microemboli With Vacuum-Assisted Venous Drainage During Pulsatile and Nonpulsatile Perfusion in a Simulated Neonatal Cardiopulmonary Bypass Model , 2008, ASAIO journal.

[16]  R. Koehler,et al.  Blood viscosity modulates tissue perfusion: sometimes and somewhere. , 2007, Transfusion alternatives in transfusion medicine : TATM.

[17]  R. Hetzer,et al.  Transfusion-free cardiopulmonary bypass in Jehovah's Witness patients weighing less than 5 kg. , 2005, The journal of extra-corporeal technology.

[18]  L. Foubert,et al.  Low extracorporeal priming volumes for infants: a benefit? , 1996, Perfusion.

[19]  T. Treasure,et al.  The Impact of Microemboli During Cardiopulmonary Bypass on Neuropsychological Functioning , 1994, Stroke.

[20]  J. Murkin Pathophysiology of cardiopulmonary bypass , 1989, Canadian journal of anaesthesia = Journal canadien d'anesthesie.

[21]  R A DEVLOO,et al.  ANESTHESIA FOR CARDIAC SURGERY. , 1965, The Surgical clinics of North America.